NBP: NBP Menu Options
Cuff Size
This option programs the monitor for the appropriate inflation pressure.
The cuff size setting is based on the patient's age. There are three
choices— Adult, Pediatric, and Neonatal. This can be set in Monitor
Defaults.
A label showing your choice— ADT, PED, NEO— is displayed at the bottom
of the parameter window.
If you need to select a different size, follow this procedure:
1. Select CUFF SIZE option from the NBP Menu. A popup menu opens
displaying the three options.
MAIN
MENU
PREV.
MENU
CUFF
SIZE:
ADULT
> ADULT
PEDIAT
NEONATE
113
CUFF SIZE Popup Menu
2. Use the Trim Knob control to make your selection. The change will not be
in effect until the Trim Knob control is pressed and the popup menu closed.
3. Press the Trim Knob control to close the popup menu.
Cuff Inflation
Pressures
The cuff size option determines the inflation pressure for the first
measurement:
Adult, 160 mmHg
Pediatric, 140 mmHg
Neonatal, 125 mmHg
Cuff inflation thereafter is 30 mmHg (adult/pediatric) and 20 mmHg
(neonatal) over last systolic pressure with a maximum inflation pressure of
300 mmHg for adult, 250 mmHg for pediatric, and 150 mmHg for neonatal.
Clear NBP Reading
10-10
If you want to remove the values displayed in the NBP parameter window,
select this option. The values will be replaced with "X". This action also
removes those values from Vital Signs history.