Model 6487 Reference Manual
Calibration
16-9
Table 16-5
20µA to 20mA range calibration summary
Model 6487 Range Calibrator Currents*
20µA
0µA
+20.00000µA
-20.00000µA
200µA
0µA
+200.0000µA
-200.0000µA
2mA
0mA
+2.000000mA
-2.000000mA
20mA
0mA
+20.00000mA
-20.00000mA
* Calibrate zero, positive full-scale, and negative full-scale