Technical Menu screen 16 : Settings-Pressure/Flow 3
This screen is displayed when you press Initial stroke button on the Technical Menu screen 14:
Settings-Pressure/flow 1.
Please set the stroke/flow the system will operate during the kidney connection and until the
beginning of the perfusion (Perfusion main Screen).
The system displays the value of the stroke (target) and the value of the associated estimate flow
(TBD during the final validation).
Stroke = 0 to 200 : flow =0-1 mL/min
Stroke= 300 to 500 : flow= 0-10 mL/min
Stroke = 600 to 800 : flow = 20-80 mL/min
Stroke= 900 to 1100 : flow = 80-150mL/min
Stroke =1200 to 1400 : flow = 150-230 mL/min
Stroke = 1500 to 1700 : flow = 250-300 mL/min
Stroke= 1800-2000 : flow = 310-350 mL/min
Stroke > 2000 : flow = 350-400mL/min
You can increase or decrease the flow by pushing on the up or down arrow and push OK button
to validate this value and go to the Technical Menu screen 14: Settings-Pressure/flow 1.