SYS + value: permit to see the actual systolic value (in mmHg) and change this value by
pushing on the button. Press on the control permit to have access to the pressure setting
message box.
value +mmHg: display of the Mean pressure value. Mean pressure is calculated using
the formula (Sys+2Dia)/3.
DIA + value: display of the diastolic value.
Flow + value mL/min : display of the perfusate flow
Res + value: display of the renal resistance value. Renal resistance is calculated by mean
pressure (mmHg)/ flow (ml/min).
On the bottom of the screen, different controls and information are also available:
- Stop perfusion: Only displayed when the pulse pump is OFF. This command permits to
end the perfusion and goes back to the Main Menu screen.
-
-
-
: Battery symbol which permits to have a global view of the remaining charge
of the batteries. You can also push on this symbol to display a message box showing the
percentage of charging of each battery and the estimated hours remaining.
Transport button : you can lock and unlock the touchpad screen by pushing on this
button
After 60 seconds, if no one touches the display, the touchpad is automatically locked. To
unlock the touchpad you can touch any area of the screen to display a message box to
confirm your choice.
When the AC line is plugged into a proper socket, a symbol appears on the screen to alert
the user the Main is plugged.