Download SVAN 977 - Inteccon
Transcript
SVAN 977 USER'S MANUAL_______________________________________________ 27 _ Setting the integration period The Integr. Period position (integration period) defines the period during which the signal is being measured (integrated) and stored as the set of Summary Results (SR). The integration period can be set in the special window, which is opened by pressing the <3 3> or <4 4> push- buttons. <4 4> The measurement will stop automatically after this period, or the measurement will start again when the selected Repetition Cycle is greater than one. The definitions of the measurement results in which the integration period is used is given in App. D. To set the integration period the user should define the required hours, minutes and seconds fields. The appropriate field may be selected by pressing the <3 3> or <4 4> push- buttons. Value of hour, minute and second is changing by means of the <5 5> or <6 6> push- buttons. Setting the number of repetition of measurement cycles The Repetition Cycles position defines the number of cycles (with the measurement period defined by Integration Period) to be performed by the instrument. The Repetition Cycles number values are within the limits [1, 1000]. Detector’s type selection The RMS Integration position defines the detector type for the calculations of the Leq, LEPd, LN% and Sel functions. Two options are available: linear (Lin) and exponential (Exp). The formulae used for the Leq calculation are given in Appendix D. LN% Selecting Lin is required to obtain the true RMS value of the measured signal. When this option is selected the value of the Leq, LEPd, LN% and Sel functions do not depend on the detector time constant: Fast, Slow or Impulse (the results are displayed without the indicator of the detectors selected in the profiles). In this case, the indicator Lin. (or L) is displayed in the different modes of the result presentation. Selecting Exp enables the user to fulfil the requirements of other standards for the time averaged Leq measurements. When this option is selected the value of the Leq, LEPd, LN% and Sel function depends on the detector time constant (Slow, Fast, Impulse). The results are displayed with the indicator of the detectors selected in the profiles (path: <Menu> / Measurement / Profiles).