Download Manual - Sin
Transcript
29.6 620RE and 620RE4 tube element loading Always isolate the pump from the mains power supply before opening the guard or performing any positioning, removal or maintenance. 620RE element pumpheads are factory set to accept Watson-Marlow LoadSure tube elements. Pumping performance will be adversely affected if LoadSure elements are not used. Open the guard using a suitable screwdriver or a 5mm Allen key. Disengage the rollers. Locate one of the “D”-shaped flanges into the lower port. (The “D” flange ensures that the element can only be loaded correctly). Wrap the tube element around the disengaged rollers of the rotor. Locate the second “D”-shaped flange into the upper port. Ensure the flat face of each “D” flange sits flush to the flange sealing face of the track. Engage the rollers. Close the guard and push it against the track until the latch engages. Connecting LoadSure elements to supply and discharge tubes LoadSure sanitary elements - which have white connectors - are connected to a tubing system using Tri-clamps and EPDM gaskets. Hold the connector end of the supply or discharge tube against the element connector, with an EPDM gasket between them. Use a Tri-clamp to engage both flanges squarely, close it and tighten. Watson-Marlow 620DuN and 620Du User Manual 89