Download GL820 midi Data Logger System
Transcript
GL820 Analog Input Circuit and Measurement Ranges Each GL820 analog input channel features electrical isolation using a photo MOS relay switching method to maintain safe and accurate measurements in demanding industrial environments. + 50Ω 500kΩ 0.05µF 500kΩ 0.05µF 50Ω Channel Switching Relay Voltage Measurement Ranges per Channel Range Maximum SPAN Minimum SPAN Minimum Resolution 20mV -22.000 to +22.000mV 0.200mV 0.001mV 50mV -55.00 to +55.00mV 0.50mV 0.01mV 100mV -110.00 to +110.00mV 1.00mV 0.01mV 200mV -220.00 to +220.00mV 2.00mV 0.01mV 500mV -550.0 to +550.0mV 5.0mV 0.1mV 1V -1.1000 to +1.1000V 0.0100V 0.0001V 2V -2.2000 to +2.2000V 0.0200V 0.0001V 5V -5.500 to +5.500V 0.050V 0.001V 10V -11.000 to +11.000V 0.100V 0.001V 20V -22.000 to +22.000V 0.200V 0.001V 50V -55.00 to +55.00V 0.50V 0.01V Process Current Measurement (with model R250 resistor) Range Maximum SPAN Minimum SPAN Minimum Resolution 1-5 V -5.500 to +5.500V 0.050V 0.001V Temperature Measurement Ranges per Channel Range Maximum SPAN Minimum SPAN Measurement Range K -200 to +1370°C J -200 to +1100°C T -200 to +400°C R 0 to +1600°C E -200 to +900°C B -270 to +2000°C S Minimum Resolution +600 to +1920°C 50°C 0.1°C 0 to +1760°C N 0 to +1300°C W 0 to +2315°C PT100 -200 to +850°C JPT100 -200 to +500°C PT1000 -200 to +500°C Optional Humidity Measurement Range www.dataq.com Range Maximum SPAN Minimum SPAN Minimum Resolution 0 to 100% 0 to +110% 1.0% 0.1% 3 330-668-1444