Download PDF
Transcript
RTI The Energy Dependence of the CT dose Profiler The Energy Dependence of the CT Dose Profiler Revision C, April 2009 This report contains information of the energy dependence of the CT Dose Profiler from RTI Electronics. 1.1 Definition of Beam Qualities We made the comparison for beam quality RQR which is defined in the IEC report 61267 Medical diagnostic X-ray equipment – Radiation conditions for use in the determination of characteristics. In addition, we also used beam quality RQT. Beam Quality kV HVL (approximate values) RQR 4 60 2,12 RQR 6 80 3,02 RQR 8 100 4,0 RQR 9 120 5,02 RQR 10 150 6,63 RQT 9 120 8.4 Table 1. The kV and HVL of the beam qualities investigated in this paper. RQR is the family of beam qualities that is used in conventional radiography without any added filtration. RQT is a quality that simulates the filtration added when a measurement is done inside a CT phantom. It gives a corresponding HVL value. We chose the test the energy dependence only at RQT 9. The energy dependence is displayed in terms of energy correction factors, that is if this the factor is multiplied with the measured dose if the detector, the detector will have no energy dependence. 1 The Energy Dependence of the CT Dose Profiler Revision C, April 2009 C