Download CM4 User Manual - Honeywell Analytics
Transcript
CM4 Four-Point Continuous Monitor 2 switches off the pump and accesses the pump limits settings. Selecting 3 is for Sample Line Integrity Option, (see Section 10.3). 2. Select 1 to toggle the pump On. A new pressure counts value is displayed. This value is the “normal reading.” You will use this value as a base to set the High and Low Limits. 3. Select 2 to set the High and Low limits. The display will show: LIMITS HI = 500 LO = 400 The cursor flashes under the first number of the high limit. Using the arrow keys, moves the cursor location left or right. Decimal values 0-9 can also be entered. 4. Set the High Limit value. The recommended default value is 52 +/-12 counts above the “normal reading,” allowing the instrument as installed to exhaust the total flow up to a head pressure of +1 PSIG (52 torr). Vacuum levels that are above the set point may indicate high exhaust pressure, blocked exhaust, low capacity pump, or low supply vacuum. The total flow is reduced, affecting transport times and flow rate through the Chemcassette®. This is due to low differential pressures across the orifice of each point. 5. Set the Low Limit value. The default value recommended is 52 +/-12 counts below the “normal reading,” allowing the instrument to indicate a 25% loss of total flow. Vacuum levels below the limit indicates too much vacuum is detected. Possible causes for this condition are blocked filters at the end of a sample line, internal blocking of the orifices, or a blocked sample line. The vacuum level is produced by pump capacity and the limiting orifice used in each point. Adding four 300-foot sample lines increases the total vacuum level, changing the count value approximately -15 counts or 1/2 inch HG (12 torr). CM4 Technical Manual Note: Setting the limits to 0 will disable the fault being issued. Once the high and low limits have been selected, press ENTER to store the values. The message “Pump Limits Updated” is printed along with the Time/Date/Serial Number. Pressing ESCAPE will exit the routine, returning the values to pre-programmed levels and not printing any message. The pump limits are also incorporated into the Service Mode Diagnostics 4.3 printout. The pump limits are also listed on the system configuration printout (2.6 Print Config). Note: If at any time the system is reset to factory defaults (Function 4.4–RESET SYSTEM), this procedure will need to be redone since the low and high limit values will be reset to factory default values. 8.10 Function 4.9 – Reboot A reboot function has been added to allow rebooting without cycling Model CM4 power. Figure 8-5 shows the prompts displayed. Display Function 9 - REBOOT Reboot ENTER CODE 9XXXX Appears only if menu password is enabled via Section 3.6.2 SAVE CONFIG 1=Y 0=NO Offer to backup configuration CONF SAVED INTERNALLY Confirms that a backup was performed. Also a line will be printed: CONFIGURATION SAVED INTERNALLY RESTARTING CM4 8-13 Reboot display Figure 8-5: Reboot Function Prompts