Download Triggered - Astronics Test Systems
Transcript
Publication No. 980897 Rev. A 3156B User Manual Name Variable Type instrHandle ViSession The Instrument Handle is used to identify the unique session or communication channel between the driver and the instrument. If more than one instrument of the model type is used, this will be used to differentiate between them. Note that the initialize function may be used to associate multiple instrument handles with a single instrument. hopDwellTime ViReal64 Programs dwell time for the fixed dwell-time frequency hop function. The same dwell time will be valid for each frequency hop. Description NOTE: Dwell time is programmed in units of µs (microsecond). For example, to program 1.37ms dwell time, enter the value as follows: 1370 Valid range: 0.5 to 21e6 (µs) Default: 0.5 Return Values Displays the return status of the function call. If the function was successful, it will return a status of VI_SUCCESS, otherwise it will return an error code. Passing the error code into the function "ri3156B_error_message()" will return a string describing the error. Ri3156b_set(query)_hop_marker Description Programs where on the frequency list the 3156B will generate a pulse, designated as Hop marker, or index point. The marker pulse is generated at the SYNC output connector. It can also be place on one of the backplane trigger lines using the ri3156B_ set_TTLTRG_n_output_state function call. Channel Dependency: independent C Syntax ViStatus ri3156b_set_hop_marker (ViSession instrHandle, ViInt16 hopMarkerIndex) ViStatus ri3156b_query_hop_marker (ViSession instrHandle, ViInt16 * hopMarkerIndex) Parameters Name Variable Type instrHandle ViSession The Instrument Handle is used to identify the unique session or communication channel between the driver and the instrument. If more than one instrument of the model type is used, this will be used to differentiate between them. Note that the initialize function may be used to associate multiple instrument handles with a single instrument. hopMarkerIndex ViInt16 Selects the hop step index for the marker output. Description Valid range: 1 to 4096 (without exceeding the physical Astronics Test Systems Programming Reference 5-95