Download Manual
Transcript
Operation 3-13 Using the scanner with the buffer The Model 2001 internal buffer can be used to store readings taken while using the scanner. The following paragraphs outline the steps necessary for fundamental data storage of scanner readings. Step 1: Configure channel 1. Press CONFIG-CHAN. The instrument will display the following: CONFIGURE CHANNEL INTERNAL-CHANS 2. EXTERNAL-INPUTS Select INTERNAL-CHANS. The Model 2001/2002 will display the following: SET INTERNAL CHANNELS 1=DCV 3. 4. 2=DCV 3=DCV 4=DCV 5=DCV Use the cursor and range keys to select channels and functions, then press ENTER when you have set all functions. Press EXIT to return to normal display. Step 2: Configure buffer 1. Press CONFIG-STORE. The unit will display the following: CONFIG DATA STORE BURST-MODE DATA GROUP CONTROL CLEAR-ALL 2. COUNT FEED Select COUNT, then press ENTER. The instrument will display the following: SET BUFFER SIZE ENTER-COUNT 3. USE-TRIGGER-MODEL Select ENTER-COUNT and press ENTER to view the following: BUFFER SIZE = 00010 4. 5. Using the cursor and range keys, select the number of readings to store. Usually, this number will be the same as the number of channels you are scanning. For example, if you are using all 10 scanner channels, select 10 readings. Press ENTER to complete programming data storage; then EXIT to return to normal display. Step 3: Configure scan, perform scan, and recall readings 1. From normal display, press CONFIG-SCAN. The instrument will display the following: SCAN OPERATION INTERNAL 2. 3. EXTERNAL Select INTERNAL, then press ENTER. Press SCAN, then follow the prompts. See “Configuring scan parameters”.