To run the On-Demand test:
1. From the Sensors sub-menu, scroll with the arrow keys until
you highlight On-Demand. Press e. If the sensor is
clear, you will see:
On Demand = ###
Clear
Values must be less than 288.
If sensor is blocked, you will see:
On Demand = ###
Blocked
Values must be greater than 480.
If the value is invalid, make a note and call Service.
2. Press b to exit.
The values displayed may fluctuate
on the screen.
7-14 Using Diagnostics
TC6035EZEM Rev. C 10/98