Download R&S UPP Audio Analyzer

Transcript
R&S®UPP
Remote Control - Command Description
Distortion Measurements
SENSe<n1>:FUNCtion:REFNment <refnment>
Parameters:
<refnment>
N1 | N2 | N4 | N8
SENSe:VOLTage:EQUalize <state>
Parameters:
<state>
ON | OFF
7.10.2 THD
●
​SENSe<n1>:​FUNCtion​ on page 687
●
​SENSe:​FUNCtion:​DISTortion<n3>​ on page 693
●
​SENSe:​FUNCtion:​MMODe​ on page 688
●
​SENSe:​VOLTage:​FUNDamental​ on page 693
●
​SENSe:​VOLTage:​FUNDamental:​MODE​ on page 693
SENSe:FUNCtion:DISTortion<n3> <state>
select the desired even or odd harmonics for THD measurement mode LSELectdi. The
measurement mode is selected with command .​SENSe:​FUNCtion:​MMODe​
on page 688.
Suffix:
<n3>
Parameters:
<state>
Example:
.
2 | 3| 4 | 5 | 6 | 7 | 8 | 9
designates the harmonics to be used
ON | OFF
SENSe:FUNCtion THD
SENSe:FUNCtion:MMODe LSELectdi
SENSe:FUNCtion:DISTortion2 ON
SENSe:FUNCtion:DISTortion4 ON
SENSe:FUNCtion:DISTortion7 ON
The distortion voltage of harmonics d2, d4 and d9 (root from the
squared sum of the selected harmonics) is measured.
SENSe:VOLTage:FUNDamental <fundamental>
Parameters:
<fundamental>
float
SENSe:VOLTage:FUNDamental:MODE <mode>
Parameters:
<mode>
User Manual 1411.1055.32 ─ 03
AUTO | GENTrack | SOURce | VALue | FIXed
693