Download User manual

Transcript
XII. SOFTWARE MANUAL
Figure 17. Laser control window (after changing each value press Enter to make
the change).
1
2
3
4
emperature for laser diode. Accuracy:
1. Setting of temperature
O
0.1 C. Range: 10.0O C to 50.0O C
2. Temperature stabilization speed parameter. Range:
Range 1
to 20. For slower
lower temperature stabilization set greater
value. This is necessary when temperature is
oscillating. Value of 5 should be suitable for most
standard systems.
3. Current limiting for laser diode TEC. Range:
Range 1.1 A
to 4 A, accuracy 0.004A.
4. Current and temperature display at the moment.
5
5. Voltage limit for laser diode. Accuracy:
Accuracy 0.001V.
Range: 1.500V to 3V. NOTE: This value always must
be greater than laser diode operating voltage in case
of
voltage
drop
in
wires.
Figure 18. LD temperature control and voltage limit control.
control
Information is subject to change without notice
Copyright 2012
21