Download Revision
Transcript
Displacement Sensor Calibration 3. Evaluate the mid-displacement position for the actuator. Standalone FlexTest SE Controllers: A. Enable the front panel Dial. B. Use the Dial to increase the Setpoint adjustment for a positive command. Select Status > Setpoint C. Adjust the Setpoint to fully extend the actuator and note the displacement signal value. D. Adjust the Setpoint to fully retract the actuator and note the displacement signal value. Automated Controllers: A. Move the Manual Cmd slider to apply a positive command (extend the actuator). B. Adjust the Manual Cmd slider to fully extend the actuator and note the displacement signal value in the Station Signals panel. C. Adjust the Manual Cmd slider to fully retract the actuator and note the displacement signal value in the Station Signals panel. The noted displacement signal values should be within 1% of each other. If these displacement signals are not within 1%, you can evaluate the following procedures to establish the zero reference: • Use the Fine Zero adjustment to shift the sensor conditioner’s zero reference position. Standalone FlexTest SE Controllers: Select Setup > Disp > Calibration > Fine Zero Automated Controllers: Adjust Fine Zero on the Calibration tab. Note • MTS Series 793 Tuning and Calibration Some conditioners have two zero adjustments (coarse and fine). Try to calibrate zero using only the Fine Zero control whenever possible. Using the Coarse Zero control affects the signal before the post amp stage of the conditioner and may require additional Gain adjustments. In some cases, a mechanical adjustment may be necessary to center the LVDT (for instructions, see the actuator product manual). Calibration Procedures 141