TRACe Subsystem
FSP
TRACe Subsystem
COMMAND
TRACe<1|2>
[:DATA]
:COPY
PARAMETERS
UNIT
TRACE1 | TRACE2 | TRACE3 |
PWCDp | CTABle |
SPURious,
<block>|<numeric_value>
...
TRACE1|TRACE2|TRACE3,
TRACE1|TRACE2|TRACE3
-
<Boolean>
<numeric_value>
--
COMMENT
-
:IQ
:AVERage
[:STATe]
:COUNt
:DATA?
:MEMory
:SET
query only
query only
<numeric_value>, <numeric_value>
NORMal,
<numeric_value>,
<numeric_value>,
IMMediate | EXTernal |
IFPower | RFPower,
POSitive,
<numeric_value>,
<numeric_value>
<numeric_value>
:SRATe
[:STATe]
:SYNChronize
[:STATe]
--, ---,
-HZ
HZ
HZ
<Boolean>
<Boolean>
General Trace Commands
TRACe<1|2>[:DATA]
TRACE1| TRACE2| TRACE3, | PWCDp | CTABle | SPURious, <block> |
<numeric_value>
This command transfers trace data from the control computer to the instrument, the query reads
trace data out of the instrument. The associated measurement window is selected with the numeric
suffix of TRACe<1|2>.
Note:
If the FM demodulator (option FS-K7) is active, only the displayed trace data is read out
and recalled. A portion of the measurement data that can be called by means of a marker,
however, is calculated from the raw measurement data. These results are no longer
available after recalling a trace; the associated queries generate a query error.
Example:
"TRAC TRACE1,"+A$
"TRAC? TRACE1"
Characteristics: *RST value:
SCPI:
Mode:
(A$: data list in the current format)
conforming
all
Return values:
The returned values are scaled in the current level unit. Returned FM-modulated measurement
values (activated option FS-K7) are scaled in Hz.
1164.4556.12
6.1-324
E-1