USER MAINTENANCE (CONTINUED)
Temperature Calibration (Continued)
Reference Device
8. Once the chamber has stabilized (no fluctuations of 0.2˚C or greater) for
a half hour, compare the reference temperature device and oven
temperature display readings.
a. If the readings are the same, or the difference between the
two (2) falls within the acceptable range of your protocol, the
display is now accurately showing the chamber temperature.
The Temperature Verification procedure is now
complete.
b. See step 9 if a difference falls outside the acceptable range
of your protocol again.

Reference Device
9. Repeat steps 2 – 8, up to two more times.
a. Three calibration attempts may be required to successfully calibrate
ovens that are more than ±2°C out of calibration.

10. If the temperature readings difference still falls outside your protocol after
three calibration attempts, contact your distributor or Sheldon Technical
Support for assistance.
End of procedure
33 | P a g e