Download 4861706 User Manual SMO5HP-2
Transcript
USER MAINTENANCE (CONTINUED) Temperature Calibration (Continued) Reference Device 8. Once the chamber has stabilized (no fluctuations of 0.2˚C or greater) for a half hour, compare the reference temperature device and oven temperature display readings. a. If the readings are the same, or the difference between the two (2) falls within the acceptable range of your protocol, the display is now accurately showing the chamber temperature. The Temperature Verification procedure is now complete. b. See step 9 if a difference falls outside the acceptable range of your protocol again. Reference Device 9. Repeat steps 2 – 8, up to two more times. a. Three calibration attempts may be required to successfully calibrate ovens that are more than ±2°C out of calibration. 10. If the temperature readings difference still falls outside your protocol after three calibration attempts, contact your distributor or Sheldon Technical Support for assistance. End of procedure 33 | P a g e