Transcript
Using the Ventilator Setting the Main Parameters Pressure Control Ventilation (PCV) The FLIGHT 60 Ventilator targets and maintains patient airway pressure at the set pressure control level throughout inspiration. Breath termination occurs when either of the following conditions exists: The set Ti elapses. The Peak inspiratory pressure exceeds the Pressure Control setting by 8 cmH2O (mbar). Maximum airway pressure never exceeds the user set High pressure alarm limit setting. The target airway pressure for pressure controlled mandatory breaths in ACMV and SIMV is the display setting above ambient pressure; not above PEEP. In PCV mode, the Waveform, and Ti/Flow ctl control buttons are not utilized and are therefore darkened. However, they remain adjustable. Both time and patient triggered mandatory breaths can be delivered in ACMV and SIMV Pressure Control operation. During SIMV Pressure Control operation, patients can breathe spontaneously between mandatory breaths with or without pressure support. When disconnecting the patient circuit during PCV/PSV ventilation, such as for suctioning, the flow may increase in order to compensate for the low pressure. After reconnecting the patient circuit, the flow automatically readjusts to meet the patient's demand. The PCV mode delivers pressure controlled breaths as the mandatory breaths. When Pressure Control is first initiated or the setting is changed, the first few breaths may cycle off early until the rise profile is optimized. If early cycling off continues, reevaluate the patient circuit configuration and lengthen the tubing as necessary. The minimum target airway pressure is 5 cmH2O/mbar above the set baseline pressure (PEEP). Operator's Manual | 51