Download R&S

Transcript
LAN Interface
FSP
Configuration of R&S FSP for Using Remote Desktop
The following steps are required to allow an external PC access to the desktop of the R&S FSP.
1. Determine the IP configuration of the network link:
Ø Open the Windows XP start menu using
the Windows key or CTRL-ESC.
Ø Click Settings - Network Connections Local Area Connection.
The Local Area Connection Status dialog
will be opened.
Ø Click the Support tab.
The current TCP/IP configuration will be
displayed.
If the entry "Assigned by DHCP" is
displayed in the Address Type field,
continue with step 2 (Installing a fixed
IP address...)
It would otherwise be sufficient to note the
IP address and to continue with step 3
(Enabling the R&S FSP...)
2. Install a fixed IP address for the TCP/IP protocol as described in the Section "Configuration of
Available Network Protocols (TCP/IP Protocol)".
Note:
Operation with a fixed IP address is strongly recommended.
When a DHCP server is used, a new IP address (which has to be determined first) will be assigned
any time the instrument is restarted. For this reason, a DHCP server would not be suitable for use
for detached operation of the R&S FSP.
3. Enable the R&S FSP for operation with the remote desktop.
1093.4820.12
4.23-26
E-1