6. INSPECTION AND CHECK PROCEDURES
Setting Procedure
(1)
Press the [INPUT] key to display the "AMP Settings" window as shown below, and then make the
following settings.
Input
: DC
Range
: 50 mV
Filter
: Off
Input : DC
(2)
Range :50mV
Filter : Off
Press the [MEMORY] key to display the "MEMORY Settings" window as shown below, and then make
the following settings.
Capture Destination
: Memory
Sampling Interval : 10µs (HSV amp: 1µs)
Block Size
: 128 x 4 (HSV amp: 16kW x4)
Capture Destination : Memory
Sampling Interval : 10µs
(set 1 us when you have
the HSV amp)
Block Size : 128 x 4
( set 16 kW x4 when
you have the HSV
amp)
GL1000-UM-251-9370
6-15