8. Installation
8.1 Installation conditions: Install the system in the following environment.
• On a flat, level floor that is sturdy enough to support the system’s weight
• Where not subjected to excessive mechanical vibrations
• Where not exposed to direct sunlight
• Which is void of sharp temperature fluctuations
• Which is not dusty
• Which is not humid
• Away from flammables
• Where not exposed to combustible or corrosive gases
• Away from equipment that generates noise
8.2 Installation space
Secure a minimum 40 cm of space on the left, right and rear of the system.
Though space requirements on the side of the system are not strictly
enforced, always ensure enough space behind the system for a person to
pass.
8.3 Grounding
Ensure 100 Ω or less resistance against ground.
* To ensure safe system operation, ground the power cable.
9. Test performance
9.1 Measurement feature
(1) Stress power supply unit
① Stress voltage
② Minimum setting resolution
③ Impressed voltage accuracy
1.0 to 100V DC
0.1 V
± (0.7% of setting + 300 mV)
(2) Measurement equipment specifications
① Current measurement range
0.1fA to 20mA
② Minimum resolution
0.1fA
③ Measurement accuracy
±1.015% (20pA range FS)
(3) System measurement range and accuracy
(Including actual environmental effects)
① Resistance measurement range 1.0 × 105 Ω to 1.0 × 1013 Ω (In 100 V stressing)
1.0 × 103 Ω to 1.0 × 1011 Ω (In 1 V stressing)
② DC measurement range
10pA to 1.0mA
* The resistance measurement range is determined by the DC measurement range and
applied voltage.
③ Resistance measurement accuracy
Accuracy is indicated at the end of the measuring cable.
[Measurement conditions]
Ambient temperature
23˚C±5˚C
Measurement mode
Long
Measurement range
Auto
Averaging (Times)
4
Charging time (sec)
60
* The above specifications do not apply to the first measurement after measurement start.
6