Download Varian Liquid Chromatography Modules Pre

Transcript
Varian Liquid Chromatography Modules
3.1.1
Electrical Requirements
Table 3. System electrical specifications
Page 18 of 30
System unit
Required supply voltage
Nominal rating
210/218 pump
115-230 V AC, 50–60 Hz
270 VA/550 VA
SD-1 pump
115-230 V AC, 50–60 Hz
350 VA
SD-2 pump
230 V AC, 50–60 Hz
1000 VA
325 UV-Vis detector
120/230 V AC, 50–60 Hz
130 VA
335 Photodiode Array detector 120/230 V AC, 50–60 Hz
130 VA
380/385-LC ELS detector
115-230 V AC, 50–60 Hz
480 VA
356-LC RI detector
115-230 V AC, 50–60 Hz
150 VA
410 autosampler
115-230 V AC, 50–60 Hz
200 VA
420 autosampler
115-230 V AC, 50–60 Hz
250 VA
430 autosampler
115-230 V AC, 50–60 Hz
250 VA
440-LC fraction collector
115-230 V AC, 50–60 Hz
100 VA
500 column valve module
115-230 V AC, 50–60 Hz
250 VA
510 column thermostat
115-230 V AC, 50–60 Hz
400 VA
530 fluidics module
115-230 V AC, 50–60 Hz
150 VA
Personal computer (typical)
115, 120, 220, 240 VAC
300 VA
Printer (typical)
115, 120, 220, 240 VAC
85 VA
Publication no. 8510253100, Issue 1, July 2009