Attaching the Proximal Flow Sensor for SST
During SST the ventilator prompts to attach the Proximal Flow Sensor.
To attach the Proximal Flow Sensor to the patient circuit
1.
Verify the Proximal Flow Sensor, pneumatic lines, and connector are not damaged.
Open the connector panel door and firmly attach the sensor connector to the
receptacle in the BDU’s front connector port labeled Prox.
2.
Figure F-4. Attaching Proximal Flow Sensor to Ventilator
1
Proximal Flow Sensor connector
insertion port
2
Proximal Flow Sensor connector
3.
When prompted, block the breathing circuit wye.
4.
When prompted to attach the Proximal Flow Sensor, unblock the circuit wye and
insert the smaller end of the sensor into the wye.
5.
When prompted, cap or seal the larger end of the sensor (marked with “UP” and
an arrow).
6.
Follow the prompts to complete SST.
Operator’s Manual
F-11