Part numbers F.9.1 Attaching the Proximal Flow Sensor for SST During SST the ventilator prompts to attach the Proximal Flow Sensor. To attach the Proximal Flow Sensor to the patient circuit 1. Verify the Proximal Flow Sensor, pneumatic lines, and connector are not damaged. Open the connector panel door and firmly attach the sensor connector to the receptacle in the BDU’s front connector port labeled Prox. 2. Figure F-4. Attaching Proximal Flow Sensor to Ventilator 1 Proximal Flow Sensor connector insertion port 2 Proximal Flow Sensor connector 3. When prompted, block the breathing circuit wye. 4. When prompted to attach the Proximal Flow Sensor, unblock the circuit wye and insert the smaller end of the sensor into the wye. 5. When prompted, cap or seal the larger end of the sensor (marked with “UP” and an arrow). 6. Follow the prompts to complete SST. Operator’s Manual F-11