Download FLUKE 5500A Service
Transcript
Calibration and Verification Performance Verification Tests 3 3-36. Thermocouple Measurement Accuracy The Thermocouple Measurement Accuracy test checks the internal temperature reference. To perform this test, measure a lag bath temperature within + 2 °C of the 5500A. Set the 5500A to Internal Reference, J thermocouple type. Make connections with J-type thermocouple wire as shown in Figure 3-1. Table 3-22 shows the test points. Table 3-22. Thermocouple Measurement Accuracy Test Nominal Value (°C) 5500A Reads (°C) Deviation °C Lag bath temperature 90-Day Spec. (°C) 0.1 3-37. Thermocouple Sourcing Accuracy The Thermocouple Sourcing Accuracy test checks the accuracy of the thermocouple measuring circuitry. For this test, measure the dc output at the 5500A front panel TC connector with a dc meter (observe polarity on the TC connector). Select External Reference and the linear output 10 µV/°C as the thermocouple “type.” Use all copper wires for these connections. The Fluke 5500A/Leads test lead kit contains a copper TC plug and wire for this purpose. Table 3-23 shows the test points. Table 3-23. Thermocouple Sourcing Accuracy Test Nominal Value (°C) Equivalent Value (mV) Measured Value (mV) (TC connector) Deviation % 90-Day Spec. (mV or %) 0 0.000 mV 0.003 mV 100 1.000 0.305% -100 -1.000 0.305% 1000 10.000 0.035% -1000 -10.000 0.035% 10000 100.000 0.008% -10000 -100.000 0.008% 3-38. Thermocouple Measuring Accuracy The Thermocouple Measuring Accuracy test checks the accuracy of the thermocouple measuring circuitry. For this test, input a dc voltage into the 5500A front panel TC terminals using copper plugs and wire (observe polarity on the TC connector), select External Reference, and select the linear output 10 µV/°C as the thermocouple “type.” The Fluke 5500A/Leads test lead kit contains a copper TC plug and wire for this purpose. Table 3-24 shows the test points. (Optional: You can also source a known temperature from a temperature calibrator using a J-type thermocouple connection and Internal Reference. Source 0 °C, 100 °C, 1000 °C, and -200 °C.) 3-31