Calibration and Verification Performance Verification Tests 3 Table 3-20. AC Current Amplitude Accuracy Test (cont.) Nominal Value Frequency Measured Value (A) (AUX) Deviation % 90-Day Spec. (%) 32.9 mA 5 kHz 0.159 32.9 mA 10 kHz 0.459 33 mA 1 kHz 0.161 33 mA 5 kHz 0.241 190 mA 1 kHz 0.086 190 mA 10 kHz 0.466 329 mA 10 Hz 0.159 329 mA 45 Hz 0.080 329 mA 1 kHz 0.080 329 mA 5 kHz 0.159% 329 mA 10 kHz 0.459 0.33 A 1 kHz 0.171 0.33 A 5 kHz 0.791 2.19 A 45 Hz 0.094 2.19 A 1 kHz 0.094 2.19 A 5 kHz 0.714 2.2 A 500 Hz 0.171 2.2 A 1 kHz 0.471 11 A 45 Hz 0.068 11 A 500 Hz 0.098 11 A 1 kHz 0.268 3-35. Capacitance Accuracy The Capacitance Accuracy test verifies the accuracy of the synthesized capacitance output at the 5500A Calibrator front panel AUX terminals. Table 3-21 shows the test points. Use the Fluke 6304C LCR Meter with PM9540/BAN output cable as shown in Figure 3-5. This cable eliminates the need for a four-wire connection. Note Make sure there are no other connections to the 5500A, especially the SCOPE BNC. Connecting any additional grounds to the 5500A can cause erroneous capacitance outputs. To overcome a noise problem, increase the meter’s signal current by increasing the voltage or frequency. 3-29