Download Corrections in Data sheets (CORRIGENDUM PORTION ONLY)

Transcript
National Council for Cement and Building Materials
Enclosure to: Ref: NCB/MM/2015-16/CQC-H/60(13)/04
Appendix –‘A’
Specification for Thermocouple Calibration System
i) Furnace with digital temperature controller
a. 15 T/c can be calibrated simultaneously
b. Accuracy is between ± 0.250C and ± 0.10C at 10000C and 20C above 10000C
c. Depth of immersion: Approx 180 mm
d. Temperature Range: 1000C to 1300/15000C
e. Power: 3 kw, 208-240V AC, 50/60HZ
ii) Digital Temperature Indicator having both port for T/c & RTD
a. Accept 100 ohm resistance thermometer & conversion to ITS-90 & IEC-751
b. Temperature Range: 2000C to 23150C & resolution 0.0010C for RTD & 0.010C for T/c
c. Indicator Units: 0C, 0F, K
d. Display: LCD graphics panel with LED back light contrast control via key board
e. PC interface: RS 239 and software included
f. Power: 220-240 volts + 10% - 13% 47 to 63HZ max 40VA
g. Battery: Sealed lead acid rechargeable cell giving approx – 10 hours continuous operation
internal battery charges.
iii) Reference T/c “S” or “R” type with cold junction
Length: 360 mm
Dia: 4 mm to 6 mm
Probe wire length approx 2 m
Reference T/c should be calibrated from UKAS calibration
iv) Suitable adopter with the system
v) Semi standard thermocouple “S” or “R” type with high alumina sheath
Length: 300 mm
Dia: 4 mm
Probe wire length approx 2 m
vi) Standard 100 ohm suitable RTD with the system
Semi standard PRT: semi standard platinum resistance thermometer: general purpose, 2m cable
length, four wire with lemo plug fitted
Temperature Range: - 1000C to 6500C
Probe Diameter: 6 mm, Probe length: 350 mm
Sensing length: 25 mm, Handle: 19 × 120 mm, Cable: 2 m PTFE
Tender for NCB/MM/2015-16
National Council for Cement and Building Materials
Annexure-II (a)
Ref: NCB/MM/2015-16/CRT/9(14)/15
1. Equipment Name
:
Hot Air Oven
2. Broad Technical Specifications
:
Chamber size =500mmx500mmx500mm (Approx)
( WxHxD)
Temperature – Ambient 2500C + 500C
Insulation: Glass wool between the walls
Rack: Three ( Stainless Steel)
Electric Supply: 220V+10, 1 Phase
Controllers: Digital based Microprocessor
Controllers cum Indicator.
Certificate: Calibration Certificate from NABL
Accredited Lab.
3. Installation & Commissioning
:
NCB – Ballabgarh
4. Quantity Required
:
2 Nos.
5. End Use
:
For use in Physical & Chemical Lab
Tender for NCB/MM/2015-16
National Council for Cement and Building Materials
Enclosure to :NCB/MM/2015-16/CRT/16(14)/19
Appendix– A
PARTICLE SIZE ANALYZER
PRINCIPLE OF OPERATION
PARTIICLE SIZE RANGE
MODE OF SAMPLE ANMYSIS
SOURCE TYPE
TECHNOLOGY
DETECTORS
TYPE QF OPTICAL ALIGNMENT
TYPE
CONTROL AND OPERATION
SPECIFICATIONS FOR PARTICLE SIZE ANALYSER
TECHNOLOGY AND MAIN UNIT
Laser diffraction technology with Mie / Modified Mie /Fraunhofer
algorithm
0.1µ or less -1400µ or more
Wet mode & Dry mode
Solid state Laser/ Gas Laser
Multi Laser technology preferably with two or three laser sources
Silicon photo Diode detector with more t han 50 detector elements
Automatic alignment
SAMPLE DELIVERY SYSTEM WET
Circulation type with centrifugal pump with turbulance stirring
Fully Automatic and SOP controlled operation and SOP Controlled
Operation.
ULTRASONIC FACILITY
Should have built-in ultrasonic facility with variable power and time
which can be set and controlled through software
VOLUME OF SUSPENSIO
REOUIRED FOR MEASUREMENT
250 ml (approx.)
PARAMETER SE1TINGS
RSD
TYPE
SOFTWARE FEATURES
SAMPLE DELIVERY SYSTEM. DRY
Should have facility to set compressor and flow conditions.
1% or better
SOFTWARE
Operation and control of the particle size system and sample delivery
module.
One click operation with simple measurement icons
Real time display of particle distribution measurement
Possibility of simultaneous measurement and data processing.
Display of measurement conditions
Display of Histrograms, Tabular data, Peak summary, Percentile value,
size percentage information, Mean Volume, Mean Number and Mean
Area information etc.
Programming saving and recalling of measurements set ups in
accordance with Standard Operating Procedures (SOP’s)
128 Channel distribution
Data Export facility into ASCII, Excel and HTML formats
Data Tolerance feature.
Data Correlation with other techniques.
Customize Report generation
TRACEABILITY AND VALIDATION
NIST Traceable standards for both wet and dry applications.
BCR Validation standards
DATA PROCESING AND CONTROL
Tender for NCB/MM/2015-16
National Council for Cement and Building Materials
Computer & Printer with Pentium Core 2 Duo Process, 2 GB of DDR Ram, minimum 500 GB
UPS
Hard Disk Drive, DVD/CD-RW, 4 USB ports, 19 Inch-LCD Monitor,
HP Key Board, Optical Mouse and Color Inkjet Printer, OR Equivalent
Accessories
Analysis
for
Dry Suitable Air compressor and vacuum unit should be provided by the
supplier
Warranty
Provided upto 3 years
Accessories
Provided supplier upto 5 years
*
Transparent and durable dust cover for entire set up to be made supplied with equipment
*
Operation manual and service manual with schematic and circuit diagram and work instruction (framed for
display near the equipment) shall be supplied by the supplier.
Tender for NCB/MM/2015-16
National Council for Cement and Building Materials
Enclosure to :NCB/MM/2015-16/CRT/17(14)/20
Appendix– A
LAB ROOM TEMPERATURE CONTROLLER
1.
Name of the equipment
:
Lab Room Temperature Controller equipment.
2.
Broad technical specification
:
Capable for maintain the temp. of room at 270C + 2 0C
:
Auto cut compressor
:
Heating & cooling arrangement
:
Digital temp. reading
:
Thermocouple for room temp. control
:
Room size Approx. 25 x 20 feet x 10 feet ( height)
:
Tender for NCB/MM/2015-16
National Council for Cement and Building Materials
Annexure-II (a)
Ref: NCB/MM/2015-16/CDR/56(14)/38
1. Equipment Name
:
Hot Air Oven
2. Broad Technical Specifications
:
Having Internal size of – 95 cm height x 95 cm
width x 100 cm depth, Insulation of 7.5 cm of
suitable material
Range: Ambient to 110 0C, with digital
temperature indicator cum controller
Material of construction :(i)
Inner chamber made of Stainless Steel
Grade -304
(ii)
Outer body made up of Mild Steel having
powder coating.
3. Installation & Commissioning
:
NCB – Ballabgarh
4. Quantity Required
:
4 Nos.
5. End Use
:
For sample conditioning purpose in MPI
Laboratory
Tender for NCB/MM/2015-16
National Council for Cement and Building Materials
Enclosure: NCB/MM/2015-16/CRT/32(14)/43
APPENDIX –‘A’
Automatic Bomb Calorimeter
Description: 1.
2.
3.
4.
5.
6.
7.
Mode of Analysis
:
Fully Automatic (Isoperibol, Adiabatic & Dynamic)
Measurement Range
:
500-10,000 Kcal/Kg
Precision
:
0.05% RSD
Resolution
:
0.0001 Cal/Gm
Protection class to DINEN60529 or other International Standard
Temperature Resolution
:
0.0010C
Jacket Type
:
Water jacket with temp. controller with a precision of
+0.10C as per (ASTM/BIS/DIN)
8. Display
:
Graphic display on LCD Screen & soft touch key pads
integrated with system.
9. Correction
:
Nitrogen or Acid, Fuse wire, Sulphur & Halogen.
10. Memory
:
More than 1000 test results
11. Sample Crucible Size
:
One gram
12. Measurement Precision
:
As per ASTM D-5865/04 & BIS1350 Pt 2 DIN 519000
13. Oxygen Bomb
:
Removable bomb type not fixed bomb concept, No
additional bucket in system.
14. Water Handling system/ Chiller:
Fully automated integrated close loop water handling system
from original equipment manufacturer. No External water
handling system should be attached with the system.
15. Oxygen Filling Vessel & Venting: Automatic Oxygen filling & venting
16. System Structure
:
Inner vessel with water filling and draining automatically
with level sensor.
17. Bomb
:
SS Heavy duty with pressure Test Certificate (230 bar) to
suit up to 9500 cal vol. 260ml.
18. Standard Accessories required:
(1) Regulator
(2) Pressure Gauge
(3) Cotton Thread
(4) Ignition wire
(5) Crucibles Metal (25 No.)
(6) Certified Benzoic Tablets etc.
19. Transparent and durable dust cover for entire set up to be made supplied with equipment
Tender for NCB/MM/2015-16
National Council for Cement and Building Materials
Annexure –II (a)
Ref: NCB/MM/2015-16/CRT-H/35/48
1.
Equipment Name
2.
Broad Technical Specification
:
3.
Installation and Commissioning
:
No. of Programmer -Two
NCB – Hyderabad
4.
Quantity Required
:
1 No.
5.
End Use
:
For Burnability Studies
High Temperature Furnace
Upto 1600 0C programmable heating;
cooling and soaking
Chamber Size: 250mm x 250mm x 300mm
(Approx.)
Tender for NCB/MM/2015-16
National Council for Cement and Building Materials
Enclosure to: NCB/MM/2015-16/CRT-H/36/49
Appendix –‘A’
Technical specification- X-Ray Fluorescence Spectrometer
Primary Application:
Quantitative and Qualitative analysis of following materials;
1. Cement, Kiln feed, Limestone, Sandstone, Gypsum, Iron ore, dolomite, Concrete,
Mortar etc.
2. Industrial waste materials such as Flyash, Slag, mine waste, aggregates, etc.
Model: Sequential wavelength dispersive X ray fluorescence spectrometer ( Fully
Computerised ). Should not require external water chiller.
Elements to be analysed:Na11U92specifically to Na, Mg, Al, Si, P, S, Cl, K, Ca, Fe in raw
materials and cement.
Additionally Mn, Ti, Pb, Zn, Ba, Cd, Cr, Hg etc.
Accuracy and precision: Should meet criteria laid as per ASTM–C–114 and ISO- 29581-2.
Automatic sample changer: To hold 8 or more sample in powder press pallet or fused glass
bead form.
Optics: Wavelength Dispersive with high sensitivity, resolution and high speed.
H.T Generator: Power of 200 W or more &air cooled.Stability against fluctuation to be stated
X- Ray Tube: Suitable target for analytical requirement as stated in 1 and 2 of primary
application.End window.
Spectrometer:
1. Vacuum path.
2. Sample size of 30 mm to 40 mm.
3. Measurement time to be <15 min.
4. Samplehandling capability of such as solid, powder pallet and glass fused bead.
5. Sample inlet of air lock system type
6. Sample rotation of ~ 30 rpm ( pls. specify)
7. Goniometer which is θ/2θ independent drive mechanism with high angular
reproducibilty.
8. Crystal changer ofminimum 3 position, bidirectional and automatic.
9. Cabinet Temperature stability and it details to stated.
10. Detectors: Scintillation counter and flow proportional counter.
11. Safety standard against X Ray as per international guidelines.
Tender for NCB/MM/2015-16
National Council for Cement and Building Materials
12. Data processing software with a computer of latest operating system with IPS LED
monitor with laser printer.
Software requirement:
Total control of spectrometer function- Perfomance check, drift correction and operation.
Data evaluation and acquisition capability.
All program for quantitative and qualitative analysis of cement and related material, including
databases (details to be specified).
Matrix correction program with linear and high order line overlap correction etc.
Fundamental parameter method ( for standard less quantitative analysis of unknown
materials).
SPC module for trend analysis
UPS:
Online, generator compatible UPS (reputed make) of adequate capacity with minimum 30
minutes
backup.
Sample Holder:
10 Samples holder and 50 steel rings for sample preparation.
Spare Parts:
Spare parts for smooth functioning of equipment for up to 5 years. Please quote X Ray tube
also.
Operational training:
Two weeks training on operation, calibration, analysis and maintenance of equipment to be
provided at the supplier’s application center for two scientist.
Waranty:
1. XRF unit warranty and service as per NCB norms and requirements
2. Practical demonstration required at the technical bid evaluation stage.
Tender for NCB/MM/2015-16
National Council for Cement and Building Materials
Annexure-‘A’
Enclosure to : NCB/MM/2015-16/CRT-H/45/52
Detailed Specifications for 100 kN ACTM ( Fully Automatic)
1. Capacity……………………….100 kN (with overload safety protection)
2. Least count / Resolution…………..0.01 kN or less
3. Minimum Clearance between Platens ……390 mm
4. Minimum Distance between side columns -260mm
5. Platen size ……………………140 mm dia ( maximum)
6. Specimen size ………………..50 or 25 mm Cube or Cylinder
7. Pumping unit ………………..single phase 220V or three phase 415V
8. It should confirm to IS : 14858
9. Computer with Data analysis Software
a. Software should be windows based with automatic peak load record and with no manual
adjustment.
b. Data acquisition, storage, management and analysis should be freely programmable and
fully automated.
c. Full Address of Manufacturer with Serial No., Model No. and ISI Mark.
Tender for NCB/MM/2015-16
National Council for Cement and Building Materials
Annexure-‘A’
Enclosure to : NCB/MM/2015-16/CRT-H/46/53
Detailed Specifications for 50 kN ACTM ( Fully Automatic)
2. Capacity……………………….50 kN (with overload safety protection)
3. Least count / Resolution…………..0.01 kN or less
4. Min Clearance between Platens ……390 mm
5. Min Distance between side columns -260mm
6. Platen size ……………………140 mm dia ( maximum)
7. Specimen size ………………..50 or 25 mm Cube or Cylinder
8. Pumping unit ………………..Single phase 220V or three phase 415V
9. It should confirm to IS : 14858
10. Computer with Data analysis Software
a. Software should be windows based with automatic peak load record and with no manual
adjustment.
b. Data acquisition, storage, management and analysis should be freely programmable and
fully automated.
c. Full Address of Manufacturer with Serial No., Model No. and ISI Mark.
Tender for NCB/MM/2015-16
National Council for Cement and Building Materials
Annexure –II (a)
Ref: NCB/MM/2015-16/CDR/124/105/59
1.
Equipment Name
2.
Broad Technical Specification
Hot Air Oven ( Conditioning Equipment)
:
-
Having Internal Size of -100 cm height X
100 cm depth X 100 cm width
-
Range from ambient to -1100C, with digital
temperature indicator cum controller.
-
It should also have a heat circulation fan.
Accuracy: +10C.
-
It should also have a glass thermometer for
visual display of temperature.
-
There should be provision of inserting the
sensor externally for calibration purpose.
-
3.
Installation and Commissioning
:
Internal chamber made of stainless steel of
grade 304 and outer body should be of Mild
Steel having Powder Coating.
NCB - BALLABGARH
4.
Quantity Required
:
08 Nos.
5.
End Use
:
For sample conditioning purpose in MPI
Laboratory.
Tender for NCB/MM/2015-16
National Council for Cement and Building Materials
Annexure –II (a)
Ref: NCB/MM/2015-16/CQC/147/71
1.
Equipment Name
2.
Broad Technical
Specification
Hot Air Oven with Micro processed PID Digital
Temp. Controller cum Indicator
:
Condition: Calibration of Oven Calibrated by NABL
accredited Lab /
(i)
Range: Ambient to 2500C
(ii)
Resolution of Indicator : 0.010C
(iii)
Stability + 0.050 C
(iv)
Dimensions : 300x300x300mm
(v)
Capacity: 28 Ltr.
As per BIS -1269 Part -2
Power Supply
220 /230 Volts
3.
Installation and
Commissioning
:
NCB – BALLABGARH
4.
Quantity Required
:
(01 No.)
5.
End Use
:
For purpose to use on artifact for calibration.
Tender for NCB/MM/2015-16
National Council for Cement and Building Materials
Annexure-II (a)
Ref: NCB/MM/2015-16/CME/178/85
1.
Equipment Name
2.
Broad Technical Specification
:
Care and carry strap, hand held remove unit,
(O2 / CO/ CO2) sensor, battery charger, water
/ SO2 trap, integral printer, high storage
memory
Measuring range O2= 0 - 21.7%
CO: 0-10000PPm
CO2: 0-30.7%
Accuracy of ± 0.5 % and instant content
readings of parameters
Pump suction min upto 5000 mmwG and
operatable at high flue gas temp (upto 10000C
and dust laden gases.
Selectable key for different fuel types
Rugged combustion probe with detachable
steel shaft Remote hand set with protective
rubber indicates (O2 / CO/ CO2), combustion
efficiency, excess air, battery life.
3.
Installation and Commissioning
:
NCB-BALLABGARH
4.
Quantity Required
:
3 Nos.
5.
End Use
:
For Measurement of Flue Gas
Portable Flue Gas Analyzer (O2 /CO/CO2)
Tender for NCB/MM/2015-16
National Council for Cement and Building Materials
Annexure –II (a)
Ref: NCB/MM/2015-16/CQC/429/87
1.
Equipment Name
2.
Broad Technical
Specification
:
To maintain lab temperature between 20oC to 27oC
and humidity between 40 to 50% RH
3.
Installation and
Commissioning
:
NCB – BALLABGARH
4.
Quantity Required
:
09 No.
5.
End Use
:
To maintain temp. & RH conditioning
Split AC Type temperature controller i.e. Heating
& cooling to maintain the laboratory temp.
Tender for NCB/MM/2015-16
National Council for Cement and Building Materials
Annexure –II (a)
Ref: NCB/MM/2015-16/CRT/563/96
1.
Equipment Name
2.
Broad Technical Specification
:
3.
Installation and Commissioning
:
Inner chamber size =18”x18”x18” (Approx)
Range Ambient 2500C
Heating Element must be insides and
bottom ribs
Double wall inside stainless steel out wall
should be coated rust proof paint.
With pengorated adjustable shelves
Glass wool insulation
Light weight
220 Volt AC on/off switch
NCB-BALLABARH
4.
Quantity Required
:
2 Nos.
5.
End Use
:
For testing purpose.
Hot Air Oven
Tender for NCB/MM/2015-16
National Council for Cement and Building Materials
Annexure –II (a)
Ref: NCB/MM/2015-16/ CDR-A/573/99
1.
Equipment Name
:
Ductility apparatus with Mould for
ductility testing.
2.
Broad Technical Specification
:
As per ASTM D – 113
Length: 75.0 + 0.5 mm
Distance between clips: 30.0 + 0.3 mm
Width at mouth of clip: 20.0 + 0.2 mm
Thickness: 10.0 + 0.1 mm
The low temperature system should confirm
to ASTMD-113.
3.
Installation and Commissioning
:
NCB – BALLABGARH
4.
Quantity Required
:
01 No.
5.
End Use
:
For Bitumen’s Testing
Tender for NCB/MM/2015-16