6 Cephalometric procedures (optional)
After the exposure, release the patient by opening the ear rods and
guide him/her out. Pay attention that the patient does not hit the
secondary slot when stepping out from the cephalostat. Remove the
disposables.
6.4
Using the carpus support
Turn the rotation chassis so that the nasion support is against the
cameraholder. Turn the nasion support sideway and fasten the carpus
holder to the nasion support holder.
Press the white lever to attach the carpus support tightly.
Ask patient to remove any metal objects and place hand against the
carpus holder. Select technique factors manually.
Directive technique factors for carpus imaging
70 kV
3,2 mA
8sek
Verify that “READY” light is on. Press the exposure button to make the
exposure.
60785-4A