Micro-Tech 3100
Do the same with the second point shown below.
- I/O DEF SCROLL 2B Moisture input calibr. #2
5.0 %M = 5.0 V
Volt
ENTER %Moist
Password: Service
Default: 5.0 % 5.0 V
Min:
1.0 % 1.0 V
Max:
100.0 % 5.0 V
A.6.1.4
Define Digital Inputs
Digital inputs can be programmed. The following screen shows one logical
function per time, and allows the user to assign it to a physical input. The
NEXT key scrolls between the logical functions. The NC/NO key selects the
Normally Open (NO) or Normally Closed (NC) status of the input. Normally
Open means the input is inactive when disconnected. To program a function,
scroll with NEXT until the function is displayed, then enter the physical input
number and confirm with ENTER; finally scroll with NC/NO until the desired
mode is displayed. By assigning a function to 0, the function is disabled.
- I/O DEF SCROLL 4 Dig. Input def.
Ext Alarm 1
___ NC
ENTER NC/NO
NEXT
Password: Service
The following table shows the available logical selections that can be assigned
to any available physical input. Typical field wiring drawings and customer
specific field wiring drawings show Reset alarms defaulted to # 1 NO, Refill
defaulted to #2 NO and Running defaulted to #3 NC. Default inputs can be
reassigned to any physical input if desired. External alarms 1, 2 and 3 can be
assigned to logical functions not on the list.
Logical selections should not be reassigned after the physical inputs have been
wired.
CAUTION
LOGICAL INPUTS RETURN TO THE
INSTRUMENT IS COLD STARTED.
DEFAULT IF THE
Appendix Table A-1: Available Logical Selections
Selections:
Default:
External alarm 1
0 NO
External alarm 2
0 NO
External alarm 3
0 NO
REC XXXX
(0 = function disabled)
A-37