Download Chapter 2 Installing the Micro-Tech 3100
Transcript
Micro-Tech 3100 3.9.3 Two Point Calibration Another way to calibrate the instrument is to use the two point calibration method. This is a method that allows calculating zero and span without knowing the exact zero value. This function is activated by pressing the SPAN CAL key in MAIN MENU 1 after “2 points” calibration mode has been selected in CAL DATA SCROLL 1. 1 Starting Two Point Calibration AUTOSPAN Weights Two points calib. press START START Password: Operator SCALE # MANUAL Press START to begin the calibration. 2. Defining First Point S# ENTER 1st POINT weight 0.00 lb Actual 0.00 lb END Password: Operator ACQ. Enter the weight corresponding to the first point, then press ACQ. 3. Acquiring First Point In this phase, the instrument acquires the A/D raw data for a minute. The procedure can be shorted by pressing the END key. The following scroll is displayed during this phase : S# ACQUIRING 1st PNT Time remaining 0000 Weight 000.0 lb END ABORT 4. Defining Second Point S# ENTER 2nd POINT weight 0.00 lb Actual 0.00 lb ENTER Password: Operator ACQ. Enter the weight corresponding to the second point, then press ACQ REC XXXX 3-17