Download OPTI CCA Service Manual, English
Transcript
MEDICAL INSTRUMENTS Service Manual AVL OPTI pH / Blood Gas Analyzer OPERATOR’S MANUAL REVISION LOG (Please record any changes made to this manual) Revision Number Release Date Approved By Description Third Edition, September 1998 Copyright 1998, AVL Scientific Corporation. All rights reserved. Unless otherwise noted, the contents of this document may not be reproduced, transmitted, transcribed, translated, stored in a retrieval system or translated into any language in any form without the written permission of AVL. While every effort is made to ensure its correctness, AVL assumes no responsibility for errors or omissions which may occur in this document. This document is subject to change without notice. For information contact: AVL Scientific Corporation AVL Medical Instruments AVL List GmbH 33 Mansell Court Stettemerstrasse 28 Kleiststrasse 48 P.O. Box 337 CH-8207 Schaffhausen A-8020 Graz Roswell, Georgia, USA 30077 Switzerland Austria 1-800-526-2272 41-848-800-885 43-316-987-0 PD7005 REV D - Important Information! - Important Information! - This Service Manual contains important warnings and safety instructions to be observed by the user. This instrument is only intended for one area of application which is described in the instructions. The most important prerequisites for installation, operation and safety, are explained to ensure safe and reliable operation. No warranty or liability claims will be covered if the instrument is installed in areas other than those described or if the necessary prerequisites and safety measures are not observed. The instrument is intended to be operated by qualified personnel capable of observing these prerequisites. Only accessories and supplies either delivered by or approved by AVL are to be used with the instrument. Due to the instrument operating principles, analytical accuracy depends on correct operation, function, and a variety of external influences beyond the manufacturer’s control. Therefore the test results from this instrument must be carefully examined by licensed physician, before further measures are taken based on the analytical results. Instrument adjustment and maintenance with the covers removed and connected to power, should only be performed by a qualified technician taking appropriate safety precautions and aware of the possible dangers of electrical shock. Instrument repairs are only to be performed by the manufacturer or authorized service personnel. Symbol ! Explanation: This symbol is located on the inside of the instrument: "Refer to the Operator's Manual / Service Manuals". Symbol for instrument type B: A B-type instrument falls under safety categories I, II or III, or has an internal power supply, providing the required insulation against discharge current and reliable ground connections. - Important Information! - Important Information! - - Operating Safety Information • • This instrument falls under Safety Category I. This instrument is a Type B device. This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interferences, and (2) this device must accept any interference received, including interference that may cause undesired operation. Warning: Changes or modifications to this unit not expressly appoved by the party responsible for compliance could void the user’s authority to operate the equipment. Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communication. However, there is no guarantee that intereference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: • Reorient or relocate the receiving antenna • Increase the separation between the equipment and the receiver • Connect the equipment into an outlet on a circuit different from that to which the receiver is connected • Consult the dealer or an experienced radio/TV technician for help CAUTION: • • • • • • The main power cord may only be plugged into a properly grounded socket. When using an extension cord, make sure it is properly grounded and fused. Disconnection of the ground lead or a loose ground connection, inside or outside the instrument, may yield unsafe or hazardous operation of the instrument. Intentional disconnection of the grounding should be avoided. When replacing fuses, make sure that the fuses used are of the specified type and rating. Never use repaired fuses or short-circuit the fuse holders. The instrument is designed as a closed conventional device which is not waterproof. Do not operate the instrument in an environment with explosive or hazardous gases or in the vicinity of anesthetic gas mixtures containing oxygen or nitrous oxide. The instrument is suitable for continous operation. - Operating Safety Information - Contents CONTENTS 1 INTRODUCTION General Information...................................................................................................................1-1 W arnings ..................................................................................................................................1-1 2 GENERAL DESCRIPTION Specifications............................................................................................................................2-1 Principles of Operation ..............................................................................................................2-5 Fluidics Block Diagram ..............................................................................................................2-9 OPTI Cassette ........................................................................................................................ 2-10 OPTI Cassette Valve Positions ................................................................................................ 2-11 Assemblies ............................................................................................................................. 2-12 Software Structure OPTI 1 ....................................................................................................... 2-14 Software Structure OPTI CCA .................................................................................................. 2-21 3 INSTALLATION / MAINTENANCE Environment ..............................................................................................................................3-1 Installation ................................................................................................................................3-2 Maintenance .............................................................................................................................3-4 4 MECHANICS Pump Cartridge .........................................................................................................................4-1 Printer Assembly .......................................................................................................................4-1 Housing ....................................................................................................................................4-2 Sample Measurement Chamber Cover .......................................................................................4-2 SMC Lower Module ...................................................................................................................4-2 Handle ......................................................................................................................................4-2 Gas Manifold Module .................................................................................................................4-3 Display Assembly ......................................................................................................................4-3 Keypad .....................................................................................................................................4-3 Exhaust Fan ..............................................................................................................................4-4 Storage Compartment ...............................................................................................................4-4 Peristaltic Pump Motor ..............................................................................................................4-4 Battery Receptacle ....................................................................................................................4-4 Main Board Assembly ................................................................................................................4-5 Barcode Reader ........................................................................................................................4-6 1 Contents 5 ELECTRONICS DESCRIPTION Main Board OPTI 1 - Electronic Circuits ....................................................................................5-1 Block Diagram - Sheet 1 ..........................................................................................................5-3 Component Location - Sheet 2 .................................................................................................5-3 RAM Battery Backup and Clock - Sheet 3 .................................................................................5-3 Microprocessor - Sheet 4.........................................................................................................5-3 16 Bit EPROM - Sheet 5 ..........................................................................................................5-4 16 Bit Ram - Sheet 6 ...............................................................................................................5-4 Barcode and Serial Interface - Sheet 7 .....................................................................................5-4 Keypad and Display - Sheet 8 ..................................................................................................5-5 Cassette Detection Logic - Sheet 9 ..........................................................................................5-5 Battery Charger Circuit - Sheet 10 ...........................................................................................5-5 Voltage Regulator - Sheet 11 ...................................................................................................5-6 SMC Temperature Control - Sheet 12 ......................................................................................5-6 SMC Heat Transistor Circuit - Sheet 13....................................................................................5-6 SMC Heat Control - Sheet 14 ...................................................................................................5-7 Digital to Analog Converter - Sheet 15 .....................................................................................5-7 Analog to Digital Converter - Sheet 16 .....................................................................................5-7 Barometer - Sheet 17 ..............................................................................................................5-8 Motor Drive Control - Sheet 18.................................................................................................5-8 Printer Controller - Sheet 19 ....................................................................................................5-8 Cassette Detection Circuit - Sheet 20.......................................................................................5-9 Interconnect - Sheet 21 ...........................................................................................................5-9 Main Board Test Points ........................................................................................................... 5-10 Connector Listing .................................................................................................................... 5-11 Main Board OPTI CCA - Electronic Circuits ............................................................................. 5-18 Main Board Test Points ........................................................................................................... 5-22 Connector Listing .................................................................................................................... 5-24 2 Contents 6 TEST PROGRAMS AND DIAGNOSTICS OPTI 1 Diagnostic 1..................................................................................................... 6-1 Barometer .........................................................................................................................6-1 Battery ..............................................................................................................................6-1 Temperature .....................................................................................................................6-2 Gas Pressure ....................................................................................................................6-2 Cleaning ...........................................................................................................................6-2 Flow Test ..........................................................................................................................6-3 Optics Test .......................................................................................................................6-3 Diagnostic 2..................................................................................................... 6-4 Printer Test .......................................................................................................................6-4 Keypad Test ......................................................................................................................6-4 Display Test ......................................................................................................................6-4 Barcode Test ....................................................................................................................6-4 RS232 Test .......................................................................................................................6-4 Cover Test ........................................................................................................................6-5 Fset ..................................................................................................................................6-5 Diagnostic 3..................................................................................................... 6-6 Valve Test.........................................................................................................................6-6 Valve Drive Test ................................................................................................................6-6 Peristaltic Pump Test ........................................................................................................6-6 LED Test...........................................................................................................................6-7 Memory Test .....................................................................................................................6-7 Software Version ...............................................................................................................6-7 Reports .....................................................................................................................................6-7 6 TEST PROGRAMS AND DIAGNOSTICS CONT'D OPTI CCA Diagnostic 1................................................................................................... 6-10 Barometer ....................................................................................................................... 6-10 Battery ............................................................................................................................ 6-10 Temperature ................................................................................................................... 6-11 Gas Pressure .................................................................................................................. 6-11 Version ........................................................................................................................... 6-11 Reports ........................................................................................................................... 6-11 Barcode .......................................................................................................................... 6-14 Diagnostic 2................................................................................................... 6-14 Printer Test ..................................................................................................................... 6-14 Optics Test ..................................................................................................................... 6-14 LED Test......................................................................................................................... 6-15 3 Contents Interface Test.................................................................................................................. 6-15 IR Test............................................................................................................................ 6-15 Fan Test ......................................................................................................................... 6-15 Fset ................................................................................................................................ 6-16 Diagnostic 3 ................................................................................................... 6-18 Valve Test....................................................................................................................... 6-18 Flow Test ........................................................................................................................ 6-18 Display Test .................................................................................................................... 6-19 VDrive Test ..................................................................................................................... 6-19 Peristaltic Pump Test ...................................................................................................... 6-19 Keypad Test .................................................................................................................... 6-19 7 ADJUSTMENTS OPTI 1Summary........................................................................................................................7-1 Barometric Pressure..................................................................................................................7-1 Sample Measuring Chamber Temperature..................................................................................7-1 OPTI CCA Summary..................................................................................................................7-2 Barometric Pressure..................................................................................................................7-2 Sample Measuring Chamber Temperature..................................................................................7-3 8 TROUBLESHOOTING Displayed Alarms / Message Summary......................................................................................8-1 System Stop Alarms ..................................................................................................................8-2 System Error Alarms .................................................................................................................8-4 System W arning Messages........................................................................................................8-8 System Information Messages ................................................................................................. 8-11 Troubleshooting ...................................................................................................................... 8-13 Troubleshooting Flowcharts ..................................................................................................... 8-24 4 Contents 9 ELECTRONIC DIAGRAMS Main Board OPTI 1 - Electronic Diagram ....................................................... 9-1 OPTI 1 Block Diagram - Sheet 1 ......................................................................................9-2 Layout - Sheet 2 ..............................................................................................................9-3 RAM Battery Backup and Clock - Sheet 3.........................................................................9-4 Microprocessor - Sheet 4.................................................................................................9-5 16 Bit EPROM - Sheet 5 ..................................................................................................9-6 16 Bit RAM - Sheet 6 .......................................................................................................9-7 Barcode and Serial Interface - Sheet 7 .............................................................................9-8 Keypad and Display - Sheet 8 ..........................................................................................9-9 Cassette Detection Logic - Sheet 9 ................................................................................ 9-10 Battery Charger Circuit - Sheet 10 ................................................................................. 9-11 Voltage Regulator - Sheet 11 ......................................................................................... 9-12 SMC Temperature Control - Sheet 12 ............................................................................ 9-13 SMC Heat Transistor Circuit - Sheet 13.......................................................................... 9-14 SMC Heater Control - Sheet 14...................................................................................... 9-15 Digital to Analog Converter - Sheet 15 ........................................................................... 9-16 Analog to Digital Converter - Sheet 16 ........................................................................... 9-17 Barometer - Sheet 17 .................................................................................................... 9-18 Motor Drive Control - Sheet 18....................................................................................... 9-19 Printer Controller - Sheet 19 .......................................................................................... 9-20 Cassette Detection Circuit Sheet 20 ................................................................................ 9-21 Interconnection - Sheet 21 ............................................................................................. 9-22 Optic Module Circuit Diagrams ..................................................................... 9-23 Microprocessor - Sheet 1............................................................................................... 9-23 Photo Detector - Sheet 2 ............................................................................................... 9-24 Photo Detector Gain - Sheet 3 ....................................................................................... 9-25 Analog to Digital Converter - Sheet 4 ............................................................................. 9-26 Cassette Detection Circuit - Sheet 5 .............................................................................. 9-27 Heater Board SMC Top Plate ......................................................................... 9-28 Display Board Interconnect ........................................................................... 9-29 Printer Board Interconnect ............................................................................ 9-30 5 Contents Electronic Diagrams OPTI CCA...................................................................... 9-31 OPTI CCA Block Diagram - ............................................................................................ 9-32 Component Location Layout............................................................................................. 9-33 Optic Module Circuit Diagrams ...................................................................... 9-34 Optics Optics Optics Optics Optics Optics Optics Optics Module Module Module Module Module Module Module Module - Digital Circuit Sheet 1 ............................................................................. 9-34 Amplifier Circuit Sheet 2 ......................................................................... 9-35 Amplifier Circuit Sheet 3 ......................................................................... 9-36 Inverter Circuit Sheet 4 ........................................................................... 9-37 A/D Converter Circuit Sheet 5 ................................................................. 9-38 LED's, Gasses & DAC Circuit Sheet 6 ..................................................... 9-39 LED's & Lytes Circuit Sheet 7 ................................................................. 9-40 Connector Circuit Sheet 8 ....................................................................... 9-41 Main Board Circuit Diagrams ......................................................................... 9-42 Main Main Main Main Main Main Main Main Main Main Main Main Main Main Main Main Main Main Main Main Main Main Main 6 Board Board Board Board Board Board Board Board Board Board Board Board Board Board Board Board Board Board Board Board Board Board Board - RAM Battery Backup / Clock Sheet 1 ........................................................ 9-42 Microprocessor Sheet 2............................................................................ 9-43 16 Bit EPROM Sheet 3 ............................................................................. 9-44 16 Bit RAM Sheet 4 .................................................................................. 9-45 Barcode Interface Sheet 5 ........................................................................ 9-46 Keypad and Display Sheet 6 ..................................................................... 9-47 Cassette Detection Logic Sheet 7 ............................................................. 9-48 Battery Charger Sheet 8 ........................................................................... 9-49 Voltage Regulator Sheet 9........................................................................ 9-50 SMC Temperature Control Sheet 10 ......................................................... 9-51 SMC Heater Transistor Sheet 11 .............................................................. 9-52 SMC Heater Control Sheet 12................................................................... 9-53 Digital to Analog Converter Sheet 13 ........................................................ 9-54 Analog to Digital Converter Sheet 14 ........................................................ 9-55 Barometer Sheet 15 ................................................................................. 9-56 Motor Drive Sheet 16 ............................................................................... 9-57 Printer Control Sheet 17 ............................................................................ 9-58 Cassette Detect Sheet 18 ......................................................................... 9-59 Interconnect Sheet 19 .............................................................................. 9-60 Serial Interface Sheet 20 .......................................................................... 9-61 tHb / SO2 Laser / LED Driver Sheet 21 ..................................................... 9-62 tHb / SO2 Amplifier Sheet 22.................................................................... 9-63 Over Temperature Protection Sheet 23 ..................................................... 9-64 Contents SMC Module .................................................................................................. 9-65 Optics Optics Optics Optics Module Module Module Module Interconnect - Sheet 1............................................................................. 9-65 LED's, Gasses & Heater - Sheet 2 ........................................................... 9-66 LED's, ION's & Heater - Sheet 3 .............................................................. 9-67 SMC Heater/tHb Board - Sheet 4 ............................................................. 9-68 Display Board IR............................................................................................ 9-69 Display IR Board - Interconnect Sheet 1 ........................................................................ 9-69 Display IR Board - IR Circuit Sheet 2 ............................................................................ 9-70 10 ILLUSTRATED PARTS LIST Spare Parts List ...................................................................................................................... 10-1 Main Upper Housing (Outer View) ............................................................................................ 10-2 Main Upper Housing (Inner View) ............................................................................................. 10-4 Lower Housing Assembly......................................................................................................... 10-6 Printer Assembly ..................................................................................................................... 10-8 SMC Assembly ...................................................................................................................... 10-10 Gas Module Assembly ........................................................................................................... 10-12 Display Assembly .................................................................................................................. 10-14 Battery Receptacle Assembly................................................................................................. 10-16 SMC Top Cover Assembly ..................................................................................................... 10-18 SMC Top Plate Assembly(OPTI 1 Only).................................................................................. 10-20 11 SERVICE SPARE PARTS Spare Parts List ...................................................................................................................... 11-1 7 1 Introduction 1 Introduction OPTI System General Information This Service Manual includes information and data necessary for repair and maintenance of the OPTI 1 and OPTI CCA analyzers. The manual is intended to be used with the Operator’s Manual where operating instructions and procedures are described. In order to fully utilize the described procedures in this manual, it is necessary to be familiar with the operation and handling described in the Operator’s Manual. To ensure proper operation and performance which meets the analyzer specifications, maintenance and repairs must be performed according to the instructions described in this service manual. The use of AVL original parts and recommended materials is required to achieve performance specifications. Spare parts and the proper order numbers are described in Chapters 10 and 11. Product warranties may vary by country. Specific warranty terms and conditions are described in documents provided at the time of installation. Warnings Warnings in this manual are marked with CAUTION and describe situations or potential dangers that may be hazardous to personnel performing maintenance or service activities. Information marked with NOTE describes situations or hazards, which can cause damage or analyzer malfunction and should be avoided. The following general operating conditions should be strictly adhered to: • Never operate the analyzer near flammable or explosive gases. • Check the supply voltage before connecting the analyzer to local AC power. • When operating the analyzer connected to primary AC power, always connect it to a properly grounded 3-pole power receptacle. • Replace damaged or worn power cables or plugs. • Before opening the rear cover, turn off the analyzer and disconnect the power cable from the primary AC source. • Replace fuses with approved or original types only. 1-1 1 Introduction 1-2 • Operate the analyzer away from sources of liquids such as sinks or wash basins. • Avoid leakage or spilling of fluids inside the analyzer, which may damage the electrical assemblies. • Clean the analyzer surfaces with only a mild soapy solution as necessary. Avoid strong or harsh chemical cleaning agents that may damage the analyzer housing and surfaces. • Use proper tools and test equipment as described in this manual to complete testing and repairs. • Analyzer surfaces may be contaminated from contact with blood. Always use precaution when contacting these surfaces. • Use approved protective gloves when handling blood specimens of contacting contaminated surfaces. • Adhere to local regulations when disposing of OPTI cassettes or contaminated parts. 2 General Description 2 General Description Specifications Measurement Parameters The specifications are described for both the OPTI 1 and OPTI CCA analyzers. Certain parameters are specific only to OPTI CCA analyzers. The specifications listed below are valid for human whole blood, pH of human plasma, serum, and approved QCmaterial. Validated Measurement Range pH PCO 2 PO 2 Na+ K+ ctHb SO 2 Barometer Range 6.6 - 7.7 10 - 120 20 - 500 100 - 180 0.8 – 10.0 5 – 25 60 – 100 300 - 800 Input Values Units 0.01 / 0.001 1 / 0.1 1 / 0.1 1 / 0.1 0.1 / 0.01 0.1 1 / 0.1 0.1 pH units mmHg mmHg mmol/L mmol/L g/dL % mmHg Range * Patient Temperature Temp Total Hemoglobin tHb Hemoglobin type MCHC P 50 FIO 2 Respiratory Quotient, RQ Patient Id Patient Sex Operator Id Display Resolution (Lo / Hi) 14 - 44 °C 58 - 111 °F 1 - 26 g/dL 1 - 16 mmol/L 1 - 260 g/L adult / fetal 30.0 – 37.0 15 - 40 mmHg 0.21 - 1.0 0.7 - 2.0 11 digits max. male / female / ? 11 digits max. Default Value 37.0 °C 98.6 °F 15.0 g/dL 150 g/L adult % 26.7 mmHg 0.21 0.84 ? (unknown) *SI units are also available. 2-1 2 General Description Calculated Values Range Display Resolution Base excess in vitro Base excess in vivo BE BEecf Base excess actual Buffer base BEact BB 0 - 100 0.1 mmol/L Actual bicarbonate HCO 3 - 1 - 200 0.1 mmol/L Total CO 2 TCO 2 1 - 200 0.1 mmol/L Standard bicarbonate stHCO 3 - 1 - 200 0.1 mmol/L Standard pH stpH 6.5 - 8.0 0.001 pH units Oxygen saturation O 2 sat 0 - 100 0.1 % Oxygen content O 2 ct 0 - 56 0.1 vol% 1000 - 10 0.1 nmol/L + -40 - +40 -40 - +40 -40 - +40 0.1 Units 0.1 mmol/L mmol/L 0.1 Hydrogen ion concentration Alveolar-arterial oxygen partial pressure difference cH AaD O2 0 - 800 0.1 mmHg P50 P50 15 - 35 0.1 mmHg Anion Gap AnGp Normalized Ca++ nCa++ Hematocrit, Calculated Hct(c ) 15 - 75 1 PCV% Data Management Printout Serial Interface built-in thermoprinter 1x RS 232, 9-pin SUBMIN D/F Infrared Interface 1XIR (Unidirectional or Bi-directional) Gas Supply Calibration Gas 2-2 14 % O 2 , 6 % CO 2 , balance N 2 , max. pressure 145 psi (10 bar) ASCII or ASTM ASTM 2 General Description Operating Parameters Sample type Sample input device Sample size Sample input Analysis time Type of measurement Units Heparinized whole blood, plasma/serum (pH only), AVL approved quality control materials syringe, capillary, microsampler OPTI 1 80µL OPTI CCA 125µL automatic aspiration < 2 minutes Optical fluorescence and reflectance conventional, standard international (SI) Temperature / Humidity Ambient temperature Measuring chamber temperature Relative humidity 15 °C - 32 °C 37 °C ± 0.1 °C (60 °F - 90 °F) (98.6 °F ± 0.18 °F) 5 - 95 % non-condensing Electrical Supply Voltage range Frequency Power consumption 90 - 250 VAC, 50 - 60 Hz typical 110 VA Classification Safety category Instrument type Operation type Protective system Ex - protection I B (following ÖVE - MG/EN 60 601-1, IEC 601-1 with optional power supply) For continuous operation IP20 The device is not specified for operation inside explosion hazardous areas. Dimensions / Weight Height Width Depth Weight 4.875 14.25 4.75 10.5 inches (12.4 cm) inches (36.2 cm) inches (27.8 cm) pounds (4.8 kg ) with battery 2-3 2 General Description Acoustic Noise Level standby ready wash/dry measurement 28 28 60 43 dbA dbA dbA dbA Test Certificates CSA CE 2-4 2 General Description Principles of Operation Luminescence is the emission of light energy resulting from "excited" molecules returning to a resting state. When luminescence is initiated by light, it is commonly referred to as fluorescence. When a fluorescent chemical is exposed to light energy of an appropriate "color", electrons in the molecules of the fluorescent chemical are "excited". A very short time later, the electrons return to a resting state and in this process sometimes emit a small amount of light energy. This energy is less than the excitation energy and thus has a different color. That is, the emitted light (fluorescence emission), is red-shifted from the excitation light and is much less intense. Fluorescent optodes (from optical electrodes) essentially measure the intensity of light emitted from fluorescent dyes. The emitted light is distinguished from excitation light by means of optical filters. Because the excitation light energy is kept constant, the small amount of light that results is changed only by the concentration of the analyte. The concentration of the analyte is determined by the calculation of the difference in fluorescence measured at a known calibration point and fluorescence measured with the unknown concentration of analyte. The P O 2 optode measurement principle is based upon luminescence, first documented in the 1930's, and commercially utilized to measure blood P O 2 in 1983. The relationship of luminescence to P O 2 is quantified by the Stern-Volmer equation which describes how the fluorescence emission intensity "I" is reduced as the PO 2 "P" is increased. Unlike conventional electrochemical "Clark" P O 2 electrodes, the oxygen optode does not consume oxygen molecules during the measurement. The pH optode measurement principle is based upon pH-dependent changes of the luminescence of a dye molecule immobilized in the optode. Chemists have used such pH indicator dyes for many years to perform acid-base titrations in turbid media. The relationship of luminescence to pH is quantified by a variant of the Mass-Action Law of chemistry that describes how the fluorescence emission intensity increases as the blood pH is increased above the dye's characteristic pKa. pH optodes do not need a reference electrode to measure pH; however, they exhibit a small sensitivity to the ionic strength of the sample being measured. The PC O 2 optode measurement principle is based upon placing a pH optode behind an ion-impermeable membrane, just as conventional PC O 2 blood gas electrodes employ the Severinghaus C O 2 electrode construction. As such, PC O 2 optodes suffer from non-selective interference from volatile acids and bases in blood just as conventional PC O 2 electrodes. 2-5 2 General Description During the measurement, light originating from lamps in the analyzer is passed through optical filters so that specific colors are transmitted to the sensors, causing them to emit fluorescence. The intensity of this emitted light depends upon the partial pressure of oxygen (P O 2 ), carbon dioxide (P O 2 ) or hydrogen ion concentration (pH) of the blood in direct contact with the sensors, as described above. The light emitted by the fluorescent sensors passes through lenses and additional optical components. A filter is used to isolate specific colors of interest from this returning light for measurement by a light detector. The Na and K ion optodes are based upon the principle of Ion Selective Electrodes (ISE’s). The optodes use ion selective recognition elements (ionophores) similar to those used in ISE’s, however the ionophores are linked to fluorescent dyes instead of electrodes. These types of dyes have been used since the 1970’s to visualize and quantify cellular ion levels in fluorescence microscopy and cell counters 8 . As the ion concentration increases, these ionophores bind larger amounts of ions and cause the fluorescence intensity to increase or decrease, depending on the particular ion. Like the pH optode, the ion optodes do not need a reference electrode, however they do exhibit a small pH sensitivity that is automatically compensated in the AVL OPTI using the measured pH. The measurement of total Hemoglobin (ctHb) and oxygen saturation (SO 2 ) uses the well-established principle of optical reflectance. Red and infrared light a three wavelengths is directed at whole, nonhemolyzed blood within a precisely defined part of the cassette over the O2 optode. The photons are partially absorbed and reflected by erythrocytes in a manner proportional to hemoglobin level; at low hemoglobin levels the unabsorbed photons strike the O2 optode’s pink overcoat and are reflected back up through the blood a second time. A portion of the reflected light exits the top of the cassette and is measured by a detector in the instrument. The infrared wavelengths are selected for the hemoglobin measurement because they are largely independent of SO2, that is, the predominate forms of adult and fetal hemoglobin absorb similarly within the 750 – 850 nm wavelength range. The red wavelength is utilized for the SO2 measurement because it is much more strongly absorbed by deoxyhemoglobin than all other hemoglobins, and is picked close to the isobestic point for oxy- and carboxyhemglobin. Maintaining high shear force just prior to measurement minimizes sensitivity to erythrocyte aggregation (rouleau formation). The optical signal of the detectors is converted by the microprocessor to a numerical readout in conventional units of measure and displayed on the front of the device. Other values commonly used for the assessment of oxygen and acid-base status are calculated from these measured values. 2-6 2 General Description The entire sample path of the OPTI cassette is filled with storage buffer to keep the sensors stable during storage. In addition, the cassette pouch is filled with a CO 2 atmosphere, which keeps the pH of the storage buffer stable. The cassette not only contains the sensors but also houses a distribution valve (cassette valve), the sample fill port, an input/output port (I/O port), a reservoir and a vent. The function of the I/O port is to connect the cassette to the peristaltic pump allowing sample aspiration and calibration gas to flow through the cassette. The storage buffer in the reservoir ensures stability over the shelf life of the cassette. The distribution valve has 2 functions: its hollow body serves as a waste container, referred to as “waste”, holding storage buffer and OPTI-trol. In addition, the cassette valve incorporates small channels, called links that connect the sensor channel, the fill port, the vent and the I/O port in different combinations. These different connections are achieved by turning the cassette valve to different positions. Each valve position has an assigned number. An unused cassette is set to position 0, the home position. The diagram on page 2-9 shows the valve positions and the connections for each position. E.g., position 2 connects the fill port via the valve link to the sensor channel and the reservoir to the I/O port. The OPTI 1 and OPTI CCA analyzers are microprocessor-based instruments measuring optical fluorescence. The disposable, single-use cassette contains all the elements needed for calibration, sample measurement and waste containment. After reading the calibration information specific to a cassette into the instrument by 'swiping' the cassette package through a convenient bar code reader, the cassette is placed in the measurement chamber. While the analyzer warms the cassette to 37.0 ± 0.1 °C, the cassette valve is turned to position 5. In this position, the I/O port is connected to the vent allowing the pump to purge calibration gas. At the same time, readings are taken from all sensors to ensure proper functioning. In addition, the fluid light gates L1, L2 and L3 are calibrated. Then the valve turns to position 1 which connects the I/O port to the reservoir and the cassette sensors to the vent via the waste. The pump starts pushing the storage buffer into the waste. L1 and L2 monitor the movement of the buffer and the pump rate. After the buffer is pushed into the waste, the valve turns to position 4, which connects the I/O port to the reservoir and the sensors directly to the vent. The pump 2-7 2 General Description first purges the cassette with calibration gas, then slows down during the actual gas calibration passing the gravimetrically prepared gas mix across the sensors. The pH sensor is calibrated via the precision storage buffer. Calibration is completed when stable readings of all sensors are obtained. The calibration process lasts from 15 to 80 seconds. After calibration verification, the cassette valve turns to position 8 closing off all ports. The instrument now waits for aspiration of sample fluid. After pressing the J key, the valve turns to position 2, which connects the fill port to the sensor channel and the reservoir to the I/O port via waste. The pump starts aspirating the sample. During aspiration, the sample light gates L1 and L2 check for sample type (blood or clear fluid), bubbles and sufficient sample volume. Once the sample has been aspirated successfully, the valve turns again to position 4. After a short warm-up period, readings are taken from all 3 sensors. Once stable readings are obtained, the valve turns to position 8 closing off all ports. The results are calculated and the cassette is ready for disposal. In case of an OPTI-trol measurement (OPTI 1 only), the user has the option of terminating the measurement or introducing a subsequent blood sample. If a blood sample is introduced, the OPTI-trol is purged into the waste and the blood sample is aspirated and analyzed. 2-8 Gas Valve Fig. 2-1: Fluidics Block Diagram NO Air Gas Manifold Assembly Low Pressure Switch Calibration Gas 6.0% CO2 14% O2 (140 psi) NC COM Fluidics Block Diagram Regulator (2.5 psi) Three Way Valve Peristaltic Pump and Cartridge Sample Casette 2 General Description 2-9 2 General Description OPTI Cassette Finger Grip Vent Valve/Waste Location of Fluid Light Gate (L1) pH Sensor O2 Sensor CO2 Sensor Location of Fluid Light Gate (L2) Key Fill Port Location of Cassette Detect Sensor (CDETECT) Valve Key Slots Na Sensor Finger Grip Input/Output Port Cl/Ca Sensor Location of Fluid Light gate (L3) Fig. 2-2: OPTI 1 Cassette 2-10 K Sensor Cassette Body 2 General Description OPTI Cassette Valve Positions Fig. 2-3: OPTI Cassette Valve Positions 2-11 2 General Description Assemblies The OPTI consists of several major components and assemblies that control the operation of the analyzer. A brief function description of the assemblies is provided below. Valves Modular Components Valve The gas valve module controls the delivery of the internal calibration gas. The gas valve module incorporates a pressure regulator, pressure sensor, gas valve and a threeway valve. Vdrive The valve drive assembly is a stepper motor that controls the positioning of the cassette valve. Pump The peristaltic pump assembly is a stepper motor and pump cartridge that provides for the pump flow used to transport sample and calibrator material through the cassette. • The sample measurement chamber (SMC) assembly provides thermostated heating control of the measuring chamber at 37 °C ± 0.1 °C (98.6 °F ± 0.18 °F). • The optics module is a microprocessor-based assembly that houses the optics detectors, preamplifiers and processing circuitry for optode sensor signals. • The main board includes the main microprocessor, RAM and EPROM memory and control circuitry for the LCD display, keyboard, printer motors, barcode and serial interface. • The OPTI Cassette contains the optode sensor used to measure the sample and also includes the reservoir for reagent and sample waste. • The OPTI CCA (only) SMC Cover provides optics and processing circuitry for ctHb/SO2 measurement. • A minimal amount of tubing is used in the OPTI system. • Storage compartment. • Liquid crystal display, 2 lines x 24 characters per line. • Keyboard with numeric keys, l and r keys, J and E keys. • Thermoprinter. • Serial RS 232 C - interface • Barcode reader 2-12 2 General Description Software Structure The OPTI software is designed to provide a simple user interface organized in a menu fashion. The operator may select a menu function by using the l and r keys to move to the desired menu selection. The menu function to be selected will be blinking. Pressing the J key will select the desired menu item. The following system software menus are provided in flow chart format and show the software menu structure: Main Menu Sampling Menu Run Menu Data Menu Setup Menu Diagnostic Menu QC - Range Setup Menu 2-13 2 General Description Main Menu – OPTI 1 RUN: SRCs Controls Data Time/Date Setup Diag SRC Controls Data Setup Run SRC Run Controls Pat Reports BP SRC Reports Patient Info Time/Date Diag Diag 1 Barometric Pressure Battery Voltage Patient Report Temperature Control Reports Print # of Copies Gas Pressure Cal Report Export Calc Param Clean Password Flow Test QC Lockout Optics Security Misc Diag 2 Units Printer Keypad Display Correlation Barcode Comm RS232 Language Cover FSet Battery Diag 3 Valve V Drive Pump LEDs Memory Version Reports Fig. 2-4: Main Menu 2-14 2 General Description Sampling Menu – OPTI 1 Power On AVL OPTI 1 Please Wait Calibrating 60 Input Pat. Data <ENTER> Patient Sample Pat ID: XXXXXXXXXXX Purging Please Wait Please Wait Rev AOPX1.10 Calibration OK Place Sample Now 599 66 Please Wait 89 Input Pat. Data <ENTER> Purging Gas System 55 Aspirating Sample Input Pat. Data <ENTER> Warm Up 37.0 C MENU <ENTER> 200 Measuring Sample 90 Input Pat. Data <ENTER> READY - Cassette Barcode MENU <ENTER> Patient Sample Op ID: XXXXXXXXXXX Patient Sample Temp: 37.0 C Patient Sample THb: 15.0 g/dL Patient Sample Sex: ? MALE FEMALE Measurement Completed Patient Sample FIO2: 0.21 Swipe Cassette Barcode pH PCO2 7.437 35.6 Run Opti-trol Controls? NO YES PO2 84.9 O2Sat 98.7 *YES (see Run Menu) Press ESC NO Insert Cassette Patient Sample Hb Type: Adult Fetal Patient Sample P50: 26.7 mmHg Please Remove Cassette Patient Sample RQ: 0.84 Cassette Inserted Next Sample <ESC> Input Pat. Data <ENTER> Please Wait Checking Cassette Calibrating 90 Place Sample <ENTER> Please Wait MENU <ENTER> __________________ AVL OPTI 1 Patient Report * if Opti-trol selected Fig. 2-5: Sampling-Menu 2-15 2 General Description Run Menu – OPTI 1 RUN: SRC Controls Data Time/Date Setup Diag SRC SRC Measurement Op ID: XXXX Swipe SRC Barcode Controls Data Run Controls Op ID: XXXXXXXX Data: Patient SRC Controls Export Run Controls Level: 1 2 3 Time/Date Setup Time: 08:51:23 Date: 15-Feb-95 Please enter PW2 Password <ENTER> To Data Menu To Setup Menu if OPTI-trol is selected Insert SRC Run Controls Level 1 Lot #:xxxxxx <ENTER> Run OPTI-trol Controls Op ID: XXXXXXXX Insert SRC Swipe Cassette Barcode to run Control DATA - Controls StatsReprt: All L1 L2 L3 SRC Inserted Insert Cassette Run OPTI-trol Level 1 Lot #:xxxxxx <ENTER> SRC Measurement 160 Optics Module Warmup Cassette Inserted Swipe Cassette Barcode to run Control SRC Measurement 73 Reading SRC Signals Please Wait Checking Cassette Insert Cassette pH PCO2 PO2 Test 7.354 43.6 103.4 Pass Calibrating Please Open Cover Calibration OK Place Control Now<Enter> Calibration OK XXX Place Control Now<Enter> Please Remove SRC Aspirating Control pH PCO2 PO2 Test 7.238 45.3 101.8 Pass Printed Results Measuring Control 99 Cassette Inserted Accept Results <ENTER> Reject / Edit <ESC> 90 pH PCO2 PO2 Test 7.238 45.3 101.8 Pass Edit <ESC> Reject Results <ENTER> Place Sample <ENTER> or End <ESC> <ENTER> Accept Results <ENTER> Reject / Edit <ESC> Please Remove Cassette Printed Results Esc Edit <ESC> Reject Results <ENTER> Aspirating Sample Input Pat. Data <ENTER> Run Controls Op ID: XXXXXXXXXXX Measuring Sample Input Pat. Data <ENTER> Run Controls Level: 1 2 3 pH PCO2 PO2 O2Sat 7.388 44.6 159.7 99.4 Run Controls Level 1 Lot #: XXXX <ENTER> Fig. 2-6: Run Menu 2-16 Diag <ESC> Please Remove Cassette To Diag Menu 2 General Description Data Menu – OPTI 1 Data: Patient SRC Controls Export Data - Patient Report: Last Last5 Data - SRC Report: Last5 Statistic Data - Controls Ranges Statistics Type Last: Displays and Prints last patient result Last 5: Prints last 5 SRC results Data - Control Ranges Control Level: 1 2 3 Last 5: Prints last 5 patient results Statistics: Prints statistic report for all levels Control Level x xxxxxx Change Lot No: N Y All: Prints all patient results in memory Please Wait Control Level x xxxxxx Print Old Data: N Y Data - SRC Delete All SRC Data N/Y Control Level x xxxxxx Delete Old Data: N Y All Data - Export Patient SRC Controls (Enter PW2) Please Enter PW2 Password <ENTER> SRC Database deleted Control Level x xxxxxx New Lot No: xxxxxx Control Level x xxxxxx Exp Date: 23-Feb-95 Control Level x xxxxxx pH: 6.900 - 7.700 Control Level x xxxxxx PCO2: 10.0 - 120.0 Control Level x xxxxxx PO2: 20.0 - 500.0 Data - Controls StatsReprt: All L1 L2 L3 Data - Controls - Type Normal OPTI-trol Fig. 2-7: Data Menu 2-17 2 General Description Setup Menu – OPTI 1 Setup Please enter PW2 Password <ENTER> Setup:Baro Pat.Info Prnt CalcPar. Security Misc. Barometer SETUP - Please enter the Baro: 727.3 mmHg Patient Info Print SETUP - Patient Info ON/OFF Defaults Limits Calculated Parameters SETUP - Print Patient Report: OFF ON SETUP - Calc Parameter BE: OFF ON Order 1 SETUP - Print Number Copies: 1 2 SETUP - Calc Parameter O2 Sat: OFF ON Order 2 SETUP - Print Cal.Report: OFF ON SETUP - Calc Parameter BB: OFF ON Order 7 SETUP - Calc Parameter HCO3: OFF ON Order 3 SETUP - Calc Parameter st.HCO3:OFF ON Order 8 SETUP - ON/OFF Patient ID: OFF ON SETUP - Defaults tHb: 15.0 g/dl SETUP - Limits pH: 7.350 - 7.450 SETUP - ON/OFF PatID: Optional Required SETUP - Defaults FIO2: 0.21 SETUP - Limits PCO2: 35.0 - 45.0 SETUP - ON/OFF Operator ID: OFF ON SETUP - Defaults RQ: 0.84 SETUP - Limits PO2: 80.0 - 100.0 SETUP - ON/OFF OpID: Optional Required SETUP - Defaults P50: 26.7 SETUP - Calc Parameter tCO2: OFF ON Order 4 Security Miscellaneous 1 SETUP - Security Password QC-LockOut SETUP - Password None PW1 PW2 PW1+2 SETUP - QC-Lockout None SRC Controls NewLot SETUP - Password Please Enter PW: XXXXX SETUP - Calc Parameter st.pH: OFF ON Order 9 SETUP - Calc Parameter AaDO2: OFF ON Order 5 SETUP - Calc Parameter BEecf: OFF ON Order 10 SETUP - Calc Parameter O2ct: OFF ON Order 6 SETUP - ON/OFF Temp: OFF ON SETUP - ON/OFF tHb: OFF ON SETUP - ON/OFF Sex: OFF ON 1 SETUP - Misc Unit Corr Comm Lang Bat SETUP - ON/OF FIO2: OFF ON SETUP - ON/OFF Hb Type: OFF ON SETUP - Units Units: Conventional SI SETUP - Correl. Factors pH Slope: 1.00 SETUP - Comm Baud Format SETUP - ON/OFF P50: OFF ON SETUP - Units Temp: C F SETUP - Correl. Factors pH Offset: 0.000 + - SETUP - Comm - Baud Baud Rate (<>): 9600 SETUP - ON/OFF RQ: OFF ON SETUP - Units Time: 12 hr 24 hr SETUP - Correl. Factors PCO2 Slope: 1.00 SETUP - Comm - Format ASCII ASTM SETUP - Units tHb: g/dL g/L mmol/L SETUP - Correl. Factors PCO2 Offset: 0.000 + - SETUP - Correl. Factors PO2 Slope: 1.00 SETUP - Correl. Factors PO2 Offset: 0.000 + - Fig. 2-8: Setup Menu 2-18 SETUP - Language Engl Deut Japn Fran Esp SETUP - Battery Saver AutoOff: 60 10 AlwaysOn SETUP - Display Light AutoOff AlwaysOn 2 General Description Diagnostic Menu – OPTI 1 RUN: SRC Controls Data TimeDate Setup Diag DIAG1:Baro Batt Temp Gas Clean Flow TestOptics > DIAG2: Printr Keypd Displ Barcode RS232 Cover Fset > DIAG3: Valve VDrive Pump LEDs Mem Vers Reports > DIAG1: - Barometer Baro: 731.9 mmHg DIAG2: - Printer Test DIAG3: - Valve Valve: ON OFF DIAG1: - Battery Battery Voltage: 12.9 V DIAG2: - Keypad Key: DIAG3: - VDrive Position: 0 DIAG1: - Temp. T3:31.48 T1:37.31 T2:37.01 DIAG2: - Display Test DIAG3: - Pump Pump Speed: 75 DIAG1: - Gas Gas Pressure: High DIAG2: - Barcode Barcode: DIAG3: - LEDs LT1:0 LT2:0 L1: 2614 L2: 2988 CD: 519 DIAG1: - Gas Purge <ENTER> or <ESC> DIAG2: - Interface RS232-Jumper2-3 <ENTER> DIAG3: - Memory Memory OK DIAG1: - Cleaning Remove PumpCart<ENTER> DIAG2: - Cover Cover: Closed DIAG3: - Version Version: AOPX1.10 DIAG1: - Cleaning Place Syringe <ENTER> Password DIAG3: - Report Errors Calibns SRCs Config. DIAG1: - Cleaning Perform Wash <ENTER> <ENTER> AVL OPTI 1 Error Report AVL OPTI 1 Calibration Report DIAG3: - Version Optics: OPTIC2B AVL OPTI 1 SRC Report AVL OPTI 1 Config. Report DIAG1: - Cleaning Reinsert Pump <ENTER> DIAG1: - Flow Test Insert Cassette <ENTER> DIAG1: - Flow Test Test in Process ...... DIAG1: - Flow Pass: 4.54 5.70 <ENTER> DIAG1: - Flow Test Discard Cassette DIAG1: - Test Optics DIAG1: - Optics Test Insert Cassette <Enter> DIAG1: - Test Optics Test in Process .. DIAG1: - Test Optics Remove Cassette Fig. 2-9: Diagnostic Menu 2-19 2 General Description QC-Range Setup – OPTI 1 Data - Controls Ranges Statistics Type Please Enter PW2 Password <ENTER> Data - Control Ranges Control Level: 1 2 3 Control Level 1 xxxxxx Change Lot #: N / Y Yes No Control Level 2 xxxxxx Change Lot #: N / Y Yes No Control Level 3 xxxxxx Change Lot #: N / Y Yes No Control Level 1 xxxxxx pH: 6.900 - 7.700 Control Level 1 xxxxxx Print Old Data: N / Y Control Level 2 xxxxxx pH: 6.900 - 7.700 Control Level 2 xxxxxx Print Old Data: N / Y Control Level 3 xxxxxx pH: 6.900 - 7.700 Control Level 3 xxxxxx Print Old Data: N / Y Control Level 1 xxxxxx PCO2: 10.0 - 120.0 Control Level 1 xxxxxx Delete Old Data: N / Y Control Level 2 xxxxxx PCO2: 10.0 - 120.0 Control Level 2 xxxxxx Delete Old Data: N / Y Control Level 3 xxxxxx PCO2: 10.0 - 120.0 Control Level 3 xxxxxx Delete Old Data: N / Y Control Level 1 xxxxxx PO2: 20.0 - 500.0 Control Database Deleted Control Level 2 xxxxxx PO2: 20.0 - 500.0 Control Database Deleted Control Level 3 xxxxxx PO2: 20.0 - 500.0 Control Database Deleted Control Level 1 xxxxxx New Lot No: 111111 Control Level 2 xxxxxx New Lot No: 222222 Control Level 3 xxxxxx New Lot No: 333333 Control Level 1 xxxxxx Exp. Date: 31-Dec-95 Control Level 2 xxxxxx Exp. Date: 31-Dec-95 Control Level 3 xxxxxx Exp. Date: 31-Dec-95 Fig. 2-10: QC Ranges Setup 2-20 2 General Description Main Menu – OPTI CCA RUN: SRCs Controls Data Time/Date Setup Diag SRC Controls Data Setup Run SRC Run Controls Pat Reports BP SRC Reports HbCal Time/Date Diag Diag 1 Barometric Pressure Battery Voltage Patient Report Temperature Control Reports Print # of Copies Gas Pressure Cal Report Export Vers QC Reports Format Barcode Pat. Info Diag 2 Printer Optics Misc Password LEDs QC Lockout RS232 Security Units IR Fan Correlation FSET Comm Diag 3 Language Gvalve Flow Disp Battery VDrive Pump Keypad Fig. 2-11: Main Menu 2-21 2 General Description Sampling Menu – OPTI CCA Power On AVL OPTI CCA AOPX 2.0 Measuring - To input xxx Pat. Data press <ENTER> Patient Sample Pat ID: XXXXXXXXXXX AVL OPTI CCA Please Wait Patient Sample Op ID: XXXXXXXXXXX Please Wait Warming up Purging Gas System Patient Sample Temp: 37.0 C xx Measurement Completed Warm Up 37.0 C MENU <ENTER> MCHC 200 pH PCO2 7.437 35.6 PO2 84.9 BE 1.5 READY-Swipe Cass.Barcode For Menu press <ENTER> Patient Sample THb: 15.0 g/dL NA K+ 156.7 5.23 tCO2p 20.7 Swipe Cassette Barcode PO2 tHb SO2 147.4 15.4 95.6 Open cover - wipe and insert Cass. - close cover Calibrating For READY press Edit Pat. Data xx Hct(c) 46.2 Patient Sample Sex: ? MALE FEMALE Patient Sample FIO2: 0.21 <ESC> <ENTER> Please Remove Cassette Patient Sample Hb Type: Adult Fetal Patient Sample P50: 26.7 mmHg Please Wait Checking Cassette Patient Sample RQ: 0.84 Aspirating Sample Please wait. Measuring Sample Please wait __________________ AVL OPTI CCA Patient Report Fig. 2-12: Sampling Menu 2-22 33.0% 2 General Description Run Menu – OPTI CCA RUN: SRC Controls Data Time/Date Setup Diag SRC Controls SRC Measurement Op ID: XXXX Data Run Controls Op ID: XXXXXXXX Data: Patient SRC Controls Export Run Controls Level: 1 2 3 Please Wait To Data Menu Open Cover - Insert SRC Close Cover Control Level x Lot# To confirm press<ENTER> SRC Measurement In Progress Swipe Cassette Barcode to run Control SRC Level x Lot Number Continue <ENTER>or <ESC> Open cover - Wipe and insert cass.-close cover SRC Measurement in process Calibrating Please Wait xx pH PCO2 PO2 Test 7.100 70.0 60.1 Pass Na+ K+ tHb Test 124.9 2.50 20.0 Pass SO2 70.0 Test Pass Time: 08:51:23 Date: 15-Feb-95 Setup Diag Please enter PW2 Password <ENTER> To Setup Menu To Diag Menu xx Place Control and press <ENTER> xxx Aspirating Control Measuring Control Please Open Cover pH PCO2 PO2 Test 7.238 45.3 101.8 Pass Please Remove SRC Na+ K+ 167.4 8.61 Printed Results Time/Date Test Pass Please Open Cover and Remove Cassette Control Results: Accept Rej/Edit Review Pass Review Please Remove Cassette Printed Results Rej/ Edit Reject Results <ESC> Edit Level,ID <ENTER> ESC ENTER Run Controls Op ID: XXXXXXXXXXX Run Controls Level: 1 2 3 Control Level x Lot# To confirm press<ENTER> Fig. 2-13: Run Menu 2-23 2 General Description Data Menu – OPTI CCA Data: Patient SRC Controls Export Data - Patient Report: Last Last5 Data - SRC Report: Last5 Statistic Data - Controls StatsReprt: ALL L1 L2 L3 Last: Displays and Prints last patient result Last 5: Prints last 5 SRC results Data - Controls - Lx or ALL Print Delete_Lx or _ALL Last 5: Prints last 5 patient results Data - SRC - Statistics Print Delete_All Print control data for particular level or for all. All: Prints all patient results in memory Please Enter PW2 Password <ENTER> DATA - Controls - Lx or ALL Delete Printed Data: N Y SRC Database deleted Please Enter PW2 Password <ENTER> Print Statistics for all levels Delete control data for particular level or for all. All DATA - SRC Delete All SRC Data: N Y Fig. 2-14: Data Menu 2-24 Data - Export Patient SRC Controls 2 General Description Setup Menu – OPTI CCA Setup Please enter PW2 Password <ENTER> Setup: Baro HbCal Print QC Format Pat.Info Misc. SETUP - HbCal AVL-Calibrator LastBlood Barometer SETUP - Please enter the Baro: 727.3 mmHg Print Format QC Clean Optics and inside of Cover <ENTER> SETUP - Print Patient Report: OFF ON SETUP-QC Swipe Barcode or <ENTER> for menu Please Open Cover SETUP - Print Number Copies: 1 2 SRC Level x Lot No.: xxxxxx Please Wipe and Insert Calibrator; Close Cover SETUP - Print Cal.Report: OFF ON SRC Level x xxxxxx Exp. Date: xxx-xx Control Level x Lot No: xxxx SETUP - Patient Info ON/OFF Defaults Limits SETUP - ON/OFF Patient ID: OFF ON SETUP - Defaults tHb: 15.0 g/dl SETUP - Limits pH: 7.350 - 7.450 SETUP - ON/OFF PatID: Optional Required SETUP - Defaults FIO2: 0.21 SETUP - Limits PCO2: 35.0 - 45.0 SETUP - ON/OFF Operator ID: OFF ON SETUP - Defaults RQ: 0.84 SETUP - Limits PO2: 80.0 - 100.0 SETUP - ON/OFF OpID: Optional Required SETUP - Defaults P50: 26.7 1 Control Level x xxxx Swipe Second Barcode Now Patient Info Miscellaneous SETUP - Format Printout Display SETUP - Format Print xxx : OFF ON SETUP - QC SRC Control Control Level x xxxx Exp. Date: xxx-xx SETUP - Format Display Show with ABG : BE SETUP - SRC or Control Level: 1 2 3 Control Level x xxxx Type: OPTI-check SETUP - Format Display Show with Lytes : tCO2p Control Level x xxxx pH: 7.360 - 7.440 SETUP - Format Display Show with Hb : Hct(c) Control Level x xxxx PCO2: 41.0 - 51.0 mmHg Control Level x xxxx PO2: 90.0 - 114.0 mmHg SETUP - ON/OFF Temp: OFF ON Control Level x xxxx Na: 138.0 - 147.0 mmol/L SETUP - ON/OFF tHb: OFF ON Control Level x xxxx K: 4.10 - 4.60 mmol/L SETUP - ON/OFF Sex: OFF ON Control Level x xxxx tHb: 11.7 - 14.6 g/dL SETUP - ON/OF FIO2: OFF ON Control Level x xxxx SO2: 86.0 - 94.0 % SETUP - ON/OFF Hb Type: OFF ON SETUP - ON/OFF P50: OFF ON 1 SETUP - ON/OFF RQ: OFF ON SETUP - Misc Unit Corr Comm Lang Bat Security SETUP - Security Password QC-LockOut SETUP - Password None PW1 PW2 PW1+2 SETUP - Password Please Enter PW: XXXXX SETUP - QC-Lockout None SRC NewLot Both SETUP - Units Units: Conventional SI SETUP - Correl. Factors xxx Slope: 1.00 SETUP - Comm Format Auto-Export SETUP - Units Temp: C F SETUP - Correl. Factors xxx Offset: 0.000 + - SETUP - Comm - Format1 RS232 IR SETUP - Units Time: 12 hr 24 hr SETUP - Language Engl Deut Japn Fran Espa SETUP - Battery Saver AutoOff: 60 10 AlwaysOn SETUP - Display Light AutoOff AlwaysOn SETUP - Comm - Format2 ASCII ASTM Mobile-ASTM SETUP - Units tHb: g/dL g/L mmol/L SETUP - Units Resolution: Low High Fig. 2-15: Setup Menu 2-25 2 General Description Diagnostic Menu – OPTI CCA RUN: SRC Controls Data TimeDate Setup Diag DIAG1:Baro Batt Temp Gas Vers Reports BCode DIAG2 DIAG2: Printr Optics LEDs RS232 IR Fan FSET DIAG3 DIAG3: GValve Flow Disp VDrive Pump keypad DIAG1 DIAG1: - Barometer Baro: 731.9 mmHg DIAG2: - Printer Test DIAG3: - Valve Valve: ON OFF DIAG1: - Battery Battery Voltage: 12.9 V DIAG2: - Test Optics DIAG3: - Flow Test Insert Cassette <ENTER> DIAG1: - Temp. T3:31.48 T1:37.31 T2:37.01 DIAG2: - Optics Test Insert SRC <Enter> DIAG3: - Flow Test Test in Process ...... DIAG2: - Optics Test Test in Process .. DIAG1: - Flow Pass: DIAG1: - Gas Gas Pressure: High DIAG1: - Gas Purge <ENTER> or <ESC> 4.54 5.70 <ENTER> DIAG2: - Optics Test Remove SRC DIAG1: - Flow Test Discard Cassette DIAG1: - Version Version: AOPX2.0 DIAG2: - LEDs LT1:0 LT2:0 L1: 2614 L2: 2988 CD: 519 DIAG2: - Display Test DIAG1: - Version Optics: OPTIC2H DIAG2: - Interface RS232-Jumper2-3 <ENTER> DIAG3: - VDrive Position: 0 DIAG2: - IR IR Loopback Test: PASS DIAG3: - Pump Pump Speed: 800 DIAG2: - Fan Cooling Fan: Off On DIAG2: - Keypad Key: DIAG1: - Report Errors Cals SRCs Config Error Report Password <ENTER> SRC Report Configuration Report Calibration Report DIAG1: - Barcode Barcode: DIAG1: - Cleaning Remove PumpCart<ENTER> DIAG1: - Cleaning Place Syringe <ENTER> DIAG1: - Cleaning Perform Wash <ENTER> DIAG1: - Cleaning Reinsert Pump <ENTER> Fig. 2-16: Diagnostic Menu 2-26 xxx 2 General Description QC Range Setup Menu – OPTI CCA Data - Controls Ranges Statistics Type Please Enter PW2 Password <ENTER> Data - Control Ranges Control Level: 1 2 3 Control Level 1 xxxxxx Change Lot #: N / Y Yes No Control Level 2 xxxxxx Change Lot #: N / Y Yes No Control Level 3 xxxxxx Change Lot #: N / Y Yes No Control Level 1 xxxxxx pH: 6.900 - 7.700 Control Level 1 xxxxxx Print Old Data: N / Y Control Level 2 xxxxxx pH: 6.900 - 7.700 Control Level 2 xxxxxx Print Old Data: N / Y Control Level 3 xxxxxx pH: 6.900 - 7.700 Control Level 3 xxxxxx Print Old Data: N / Y Control Level 1 xxxxxx PCO2: 10.0 - 120.0 Control Level 1 xxxxxx Delete Old Data: N / Y Control Level 2 xxxxxx PCO2: 10.0 - 120.0 Control Level 2 xxxxxx Delete Old Data: N / Y Control Level 3 xxxxxx PCO2: 10.0 - 120.0 Control Level 3 xxxxxx Delete Old Data: N / Y Control Level 1 xxxxxx PO2: 20.0 - 500.0 Control Database Deleted Control Level 2 xxxxxx PO2: 20.0 - 500.0 Control Database Deleted Control Level 3 xxxxxx PO2: 20.0 - 500.0 Control Database Deleted Control Level 1 xxxxxx New Lot No: 111111 Control Level 2 xxxxxx New Lot No: 222222 Control Level 3 xxxxxx New Lot No: 333333 Control Level 1 xxxxxx Exp. Date: 31-Dec-95 Control Level 2 xxxxxx Exp. Date: 31-Dec-95 Control Level 3 xxxxxx Exp. Date: 31-Dec-95 Fig. 2-17: QC-Range Setup Menu 2-27 3 Installation and Maintenance 3 Installation and Maintenance OPTI 1 System Environment Location is important for trouble-free operation of your analyzer. Before beginning setup of the analyzer, choose a location that provides a stable flat surface and meets the following environmental requirements for the unit: • Grounded electrical outlet if operated with optional power supply. • Away from direct sunlight. • Room temperature between 15 ºC and 32 ºC (59 - 90 °F). • Maximum relative humidity of 95% non-condensing. • Sufficient space to allow air to circulate around the unit. • Away from strong electromagnetic fields, such as those created by electric motors and X-ray equipment. • Away from explosive gases or vapors. • Secure and level work surface. Note: Above requirements also apply when the OPTI 1 Analyzer operates on battery power outside a laboratory setting. Before beginning installation of the analyzer, check the contents to ensure that the required accessories have been included and are free from shipping damage. Check for these items which are necessary to complete the installation of the OPTI 1 system: • Power Cord • 2 Standard Reference Cassettes (SRC) • Battery Pack • Power Supply (Battery Charger) • OPTI 1 Cassette Packs • Thermal Printer Paper • Blood Gas Controls • Peristaltic Pump Cartridge • Gas Cylinder • Capillary Tubes 3-1 3 Installation and Maintenance Analyzer Installation Begin by placing the analyzer on a secure table top that allows sufficient work space and is near an electrical power outlet. 1. Plug in the Battery Charger • • Plug the battery charger power cord into the receptacle on the left rear side of the analyzer. Plug the power cord into a grounded electrical outlet. 2. Install the battery pack in the battery receptacle • Push the battery pack into the receptacle on the left hand side of the OPTI 1. Note: The battery will need to be charged for at least 6 hours prior to use without the AC Adapter. It will be charged automatically whenever the analyzer is plugged into an electrical outlet. 3. Turn on the Power • Locate and turn on the power switch on the left side of the analyzer. The AVL OPTI 1 Analyzer will now begin warming up. During this time keypad entries can be made to configure the analyzer settings. • • • • Use the numeric keys to make changes to the information on the display. Use the J key to accept the displayed values or information. The arrow keys, l and r are used to move the cursor and select menus. The E key returns you to the previous software level. 4. Install the Gas Cylinder • • • • Remove the protective cap on the gas cylinder. Swipe the barcode of the gas bottle. Insert the cylinder into the gas receptacle located on the right hand side of the analyzer. Turn clockwise (to the right) until ‘finger’ tight. Be sure the tank fits tightly to avoid any gas leaks. Note: 3-2 A wrench is not needed to tighten the gas cylinder. The low pressure seal is made with an O-ring. 3 Installation and Maintenance 5. Install the Printer Paper • Place paper into the paper tray. • • Thread the paper into the paper feed slot, as shown in the diagram on the analyzer. Press the paper advance button to bring the paper completely through the feeder. 6. Enter Date and Time • • • Move the cursor to the correct position (with arrow key). Enter correct date, press J. Follow the same procedure to enter the correct time, then use the arrow key to make AM or PM blinking, press J Note: The Date and Time need only be entered when the system is installed. The system purges the gas system for approximately one minute. The system will complete its warm-up. Once the warm-up is complete, the analyzer is ready for sample analysis. Prior to analyzing patient samples or Quality Control materials, a complete check should be performed. In addition, run the Standard Reference Cassettes (SRC’s) to validate system performance. Refer to Chapter 4, Sec 4.5.1 “Running an SRC Measurement” of the Operator’s Manual for instructions. 3-3 3 Installation and Maintenance Maintenance The OPTI 1 analyzer requires minimal routine operator maintenance to provide trouble-free operation. The routine maintenance required is described in this section. Daily Maintenance No daily maintenance is required for the OPTI 1 analyzer. Weekly Maintenance Once per week, the Sample Measurement Chamber (SMC) must be cleaned. Open the top cover and clean the optics surfaces with a lint-free cloth, slightly moistened with a dilute alcohol or ammonia-based cleaner as needed. Make sure to remove any blood residue. Use a cotton swab for cleaning cramped or restricted areas such as near the valve drive. Annual Maintenance Once a year, the peristaltic pump cartridge needs to be replaced. As Needed Maintenance The gas cylinder is designed to provide up to 200 sample measurements. The analyzer provides a display message to inform the user when it is necessary to change the gas cylinder. The thermal printer paper supplied with the OPTI 1 includes an indicator strip to alert the user when the paper roll needs to be replaced. Clean the outside housing surfaces of the OPTI 1 using a soft damp cloth as necessary to preserve the appearance of the analyzer. Note: 3-4 Never use strong chemical or abrasive cleaners on the OPTI 1 to avoid damage to the housing surfaces. 4 Mechanics 4 Mechanics OPTI System The AVL OPTI is completely housed in a molded plastic case, which provides a chassis for the electronic and mechanical assemblies. The analyzer is designed for quick removal of all major assemblies. Some of the components such as the printer, pump cartridge and top plate are accessible from the outside of the analyzer. The analyzer interior is accessible, after removing the bottom of the analyzer. The following chapter describes the removal of the major assemblies of the OPTI analyzer. Caution: Before disconnecting any analyzer assemblies, disconnect the analyzer from primary power and remove the installed battery to avoid electrical damage or electrical shock. Pump Cartridge The pump cartridge can be removed easily. Open the printer compartment cover to access the pump cartridge. Grasp the pump cartridge and pull straight up to free it from the pump motor shaft. To reinstall the pump cartridge, first check whether the flat in the pump spool is aligned with the flat on the pump motor shaft. If the flat on the pump spool is not aligned, a coin can be used to turn the spool within the pump cartridge housing and align the pump spool with the motor shaft. The motor shaft may also be turned for easier reinstallation of the pump cartridge. For easiest alignment of the motor shaft, turn the motor shaft so the flat on the motor shaft is parallel to the right side of the analyzer housing with motor shaft flat to the right. Insert the pump cartridge and press down with a slight force so the cartridge fits onto the motor shaft and ensure the cartridge is seated into the pump ports. Press down on the pump spool to prevent the pump roller from binding on the pump cartridge housing. Printer Assembly To remove the printer assembly, first open the printer cover to gain access to the printer assembly. First remove the pump cartridge to allow enough free space for removing the printer assembly without interference in the printer compartment. The pump cartridge can be removed by grasping it and pulling it up. Next remove the printer paper and set aside. Two thumbscrews that hold the printer assembly in place can now be loosened. After the thumbscrews have been loosened, the printer can be slid towards the rear of the analyzer. Disconnect the printer ribbon cable that attaches to the printer. The printer assembly can now be removed. 4-1 4 Mechanics Analyzer Housing The main analyzer housing is comprised of two parts, the lower and upper housing. The assemblies located inside the analyzer can be removed. To gain access to these assemblies, the analyzer housings must be separated. To open the analyzer housing, first remove the battery on the left side of the analyzer. Turn the analyzer over to access the seven (7) screws securing the two housing sections together. Remove the screws and separate the upper and lower housing sections. Caution: When turning the analyzer over to access the lower housing section, place the analyzer on a padded surface to prevent scratches or other damage to the upper housing section. Sample Measurement Chamber Cover To remove the SMC cover, the bottom of the analyzer must first be opened to separate the top and bottom portions of the analyzer housing. Locate the cable and disconnect the cable from the main board. Open the top cover of the SMC and remove the two (2) screws at the cover hinge. Next remove the hinge plate. Then pull the cover to remove it. The cable will be attached to the SMC cover. For reinstallation follow the above steps in reverse order. Note: The OPTI CCA SMC Cover is not field replaceable. The complete unit must be returned the local service repair facility. SMC Lower Module The SMC top cover must first be removed. Disconnect the cables for the valve drive motor, sample light gates and optics. Next remove the tygon tubing from the peristaltic pump port. For easier access, remove the battery housing and the gas module. Three (3) screws hold the module in place. Pull out the module from the bottom of the analyzer. For reinstallation follow the above steps in reverse order. Spare SMC modules are delivered with a calibration sheet. Upon installation of the new SMC module, the values indicated on the calibration sheet have to be programmed into the instrument using the Fset menu (see Chapter 6 - Fset). 4-2 4 Mechanics OPTI Handle The OPTI is equipped with a carrying handle for easy transport. To remove the handle, first separate the lower portion of the analyzer. To access the handle, remove the battery compartment, gas manifold module and storage compartment. Next remove the four (4) self-tapping screws which hold the handle in place. Gas Manifold Module The gas manifold module is located on the right rear side of the analyzer. The calibration gas cylinder is screwed into the module to form a gas-tight seal. The gas manifold module provides pressure regulation and controls the flow of gas when required for cassette calibration. To remove the module, first open the lower portion of the analyzer. Disconnect the tygon tubes, which connect to the peristaltic pump and clean-out port. Next disconnect the cable connector on the Main Board. There are two (2) retaining screws that hold the module in place. After removal of these screws, the module may be pulled out of the analyzer. Upon reassembly, make sure the tubes from the clean-out port and the peristaltic pump are connected to the correct nipples. The tubes should not be crossed. There are two types of gas manifolds. The older type is marked with a red dot, the newer type with a green dot (do not confuse the dot with the calibration label which is white or green and contains a 3-digit number). After installation, the correct type - red or green - must be programmed into the instrument, using the Fset menu (see Chapter 6 Fset). Display Assembly The OPTI uses a backlit LCD display module to display operating instructions and analyzer results. To remove the LCD display, first open the lower portion of the analyzer and disconnect the ribbon cable on the main board. The storage compartment and gas manifold module must also be removed. Next remove the four (4) screws which hold the LCD display assembly in place. The assembly can now be removed. To reinstall the assembly, follow the removal instructions in reverse order. Before tightening the mounting screws, make sure to center the display over the window. 4-3 4 Mechanics Keypad The keypad fitted on the OPTI can be removed by first opening the lower portion of the analyzer. The flex cable from the keypad will now need to be disconnected from the display board. Use a knife to lift off the keypad from the analyzer housing. Before replacement with a new keypad, the surface must be cleaned. Use alcohol to remove any adhesive, which may be on the surface. The surface must be clean and dry before installing a new keypad. Take the new replacement keypad and remove the backing to expose the adhesive side of the keypad. Carefully apply the new keypad to ensure proper alignment on the analyzer housing. Caution: The keypad is attached to the housing of the analyzer with an adhesive backing and can not be re-used once removed from the analyzer. Exhaust Fan The OPTI is equipped with an internal exhaust fan, which is used for cooling the electronic circuitry. To remove the fan, first open the lower portion of the analyzer. Disconnect connector that attaches to the main board. One (1) screw fastens to the fan retainer and must be removed. Next remove the retainer and slide the fan out of the analyzer. The software controls the fan. T3 measures the air temperature inside the OPTI housing and is displayed in the DIAG1 - Temp menu. If T3 rises above 31.5 ° C, the fan is activated. The fan turns off, when T3 drops below 30.5 ° C. Storage Compartment Located on the rear panel of the OPTI is a storage compartment. The storage compartment is included to provide a convenient location for storing small accessories during transport of the analyzer. To disassemble the storage compartment, first separate the lower portion of the analyzer. Then, slide the storage compartment out of the upper housing. To replace the storage compartment, fit the grooved edges of the compartment over the upper housing and slide it into place. Peristaltic Pump Motor The peristaltic pump assembly is comprised of the pump motor and the peristaltic pump cartridge and during operation provides the necessary pump flow to move reagents and samples into the OPTI. The peristaltic pump uses a stepper motor, which is controlled by the main board and turns the pump roller in the pump cartridge to generate a small vacuum. To remove the pump motor, open the lower portion of the analyzer housing for access to the assembly. Remove the pump cartridge and disconnect the pump motor cable from the main board. Locate the pump motor and remove the four (4) mounting screws. The motor may now be removed. For reinstallation, reverse above steps. 4-4 4 Mechanics Battery Receptacle The OPTI can be operated by either battery power or by connection to a supplied battery charger unit connected to AC main power. The battery can easily be exchanged with a fully charged battery as described in the Operator’s Manual. To remove the battery receptacle, first remove the battery from the analyzer. The lower housing portion of the analyzer must be opened for access. Locate the battery receptacle and disconnect the internal cable from the main board. Next remove the two (2) retaining screws and pull the receptacle out of the analyzer. Main Board Assembly The OPTI analyzer is controlled and powered by the main board assembly. This board contains the microprocessor and associated control circuitry. Interconnections for the major assemblies are by cables attached directly to the board. For board removal first open the lower portion of the analyzer housing to access the board. Disconnect all the cables attached to the board. Remove the two (2) retaining screws that secure the board to the lower portion of the analyzer housing and pull out the circuit board. Located on the back of the main board are two (2) EPROM’s which can also be replaced in the event of failure or software revision. It is not necessary to separate the analyzer housing to access these EPROM’s. The OPTI analyzer has been designed and is equipped with an EPROM cover located on the bottom of the analyzer. Remove the cover to access the EPROM’s. A special EPROM carrier can be used for easy installation and removal of the EPROM’s. This special carrier allows for customer replacement of the EPROM’s. To remove the EPROM’s after opening the EPROM cover, insert a coin under the tabs of the carrier and pry out the EPROM’s. Replacement EPROM’s are supplied with the EPROM carrier for easy installation. Caution: Always disconnect power and remove the battery before handling electronic assemblies. When replacing the main board, perform the adjustment procedures outlined in Section 7 of this manual before returning to operation. To reset the OPTI software after installing new software or in the event of system corruption, open the cover, hold in the paper advance button and cycle the main power. Caution: When the OPTI software is reset, all patient results, QC data, SRC data and setup parameters stored in system memory will be lost. 4-5 4 Mechanics To replace the main board, separate the lower housing from the upper housing by removing 7 screws. Next, disconnect all cables from the main board. For easy access, place the OPTI upside down on a work surface. Then, flip the lower housing over the handle. After all cables have been disconnected, remove the lower housing. Remove the two screws that hold the main board in place and install the new board in reverse order. The EEPROM on the main board holds all factory settings for the OPTI. To avoid reprogramming all these values, install EEPROM from the old board onto the new board. As an alternate method, record all Fset values (see Chapter 6 - Fset) prior to removal of the old board and reprogram the values after installing the new board. Note: The OPTI CCA Main Board is not field replaceable. The complete unit must be returned the local service repair facility. Barcode Reader The internal barcode reader assembly is used to read the barcode data supplied with each cassette and Standard Reference Cassette (SRC). To access the internal barcode reader, first separate the upper and lower analyzer housing sections. Locate the barcode reader guide. Push on the latch and slide the guide downward. The module is held in place by retaining screws, which must be removed. Disconnect cable from the main board. Pry out the window of the barcode reader with a screwdriver. 4-6 5 Electronic Description 5 Electronic Description OPTI 1 Main Board The AVL OPTI 1 Main Board consists of the following electronic circuits. The schematic diagram for each part of the electronic description is located in Section 9 and is referenced to a sheet number corresponding to the electronic descriptions found in this section. SYSTEM FUNCTIONAL BLOCK DIAGRAM SHEET 1 CIRCUIT DIAGRAM SHEET 2 Component Location CIRCUIT DIAGRAM SHEET 3 Real Time Clock RAM Battery Back-up Address Decoder CIRCUIT DIAGRAM SHEET 4 Microprocessor Clock CIRCUIT DIAGRAM SHEET 5 16 Bit EPROM CIRCUIT DIAGRAM SHEET 6 16 Bit RAM CIRCUIT DIAGRAM SHEET 7 RS232 Controller Barcode Controller Buzzer CIRCUIT DIAGRAM SHEET 8 Keypad Controller Keypad Decoder Parallel Display Register CIRCUIT DIAGRAM SHEET 9 Status LED Driver Cassette Detection Motor Drive Detection Gas Pressure Input 7 Stage Ripple Counter CIRCUIT DIAGRAM SHEET 10 Input Power and Power Switch Main Power Fuse F1 Battery Charge Controller Battery Status LED 5-1 5 Electronic Description OPTI 1 Main Board Continued CIRCUIT DIAGRAM SHEET 11 Fuses F2 and F3 Switching +5 Volt Regulator + 5 Volt Regulator + 12 Volt Regulator Positive to Negative Voltage Converter - 5 Volt Regulator SSMC Heater Supply CIRCUIT DIAGRAM SHEET 12 +5 Volt Reference Voltage Regulator SSMC Temperature Controller CIRCUIT DIAGRAM SHEET 13 SSMC Heater Drivers CIRCUIT DIAGRAM SHEET 14 SMC Bottom Plate NTC SMC Top Plate NTC Battery Back-up Reference CIRCUIT DIAGRAM SHEET 15 Sample Sensor LED D/A Controllers Barometric Pressure Reference Voltage CIRCUIT DIAGRAM SHEET 16 A/D Converter CIRCUIT DIAGRAM SHEET 17 Barometric Pressure Converter CIRCUIT DIAGRAM SHEET 18 Peristaltic Pump Controller Driver Valve Drive Motor Driver Gas Valve Driver CIRCUIT DIAGRAM SHEET 19 Printer Motor Controller Driver Printer Serial to Parallel Converter CIRCUIT DIAGRAM SHEET 20 Cassette Detection CIRCUIT DIAGRAM SHEET 21 Connectors to Assemblies 5-2 5 Electronic Description Main Board Circuit Diagram Sheet 1 The Main Board circuitry is divided into 20 separate circuit schematic diagrams. The division of the circuit schematic is done to facilitate component identification and signal tracing. Each diagram contains signal labels allowing service personnel to trace signals throughout the board. Located near the signal labels are small case numbers. These numbers identify the circuit sheet where the signal is connected. Main Board Circuit Diagram Sheet 2 This diagram shows the component location on the main board. Main Board Circuit Diagram Sheet 3 IC U41, a non-volatile controller, detects analyzer power supplied either by the internal battery system or the external power source. In the event of a loss of power, U41 switches to provide RAM battery backup from battery BT1. IC U38 is a real-time clock and provides time and date information used within the AVL OPTI 1. Crystal Y3 is connected to IC U38 to provide a constant operating frequency and control of the real-time clock. IC U19 and U13 are 2 of 8 decoders providing various signal outputs as labeled on the right side of the schematic diagram. These two decoders utilize signals PCS0 through PCS3 as a control signal to decode 14 outputs, as two outputs are not used. Main Board Circuit Diagram Sheet 4 A 132 pin MOTOROLA MC68372 microprocessor, U24, controls the AVL OPTI 1. Crystal Y2 is connected directly to the microprocessor to provide a stable frequency control to the microprocessor. U54B, a quad gate is also connected to the crystal and is used as a buffer to provide a 32KHz signal to other portions of the electronic circuitry. U30 is an in-line buffer used to buffer some of the microprocessor inputs. 5-3 5 Electronic Description Main Board Circuit Diagram Sheet 5 U9 and U10 are 512K by 8 EPROMS and together comprise a 16-bit data bus used by the OPTI 1. The user fits these two EPROMS to the main board in a special socket, which allows for easy removal of the EPROMS to allow software replacement. U8 is a serial 128 by 16 Serial EEPROM that stores instrument specific data. Main Board Circuit Diagram Sheet 6 The circuitry on this sheet, IC’s U7 and U11 are 128K by 8 RAM chips that together provide a 16 bit data byte. Main Board Circuit Diagram Sheet 7 IC U3 is a Dual RS232 Driver and provides the serial interface control within the OPTI 1. Connected to U3 and mounted to the Main Board is P1, a standard 9-pin D-type connector. U5 is a barcode controller. The barcode controller is connected to the internal barcode reader through J14 and an optional external barcode wand which can be connected through J2, a RJ45 type connector mounted on the main board. The barcode input from either barcode device is fed to pin 8 of U5 and an 8-bit data output is buffered by U6. BZ1 driven by transistor Q8 is a piezo speaker and provides an audible alarm when activated. 5-4 5 Electronic Description Main Board Circuit Diagram Sheet 8 Plug J13 provides input from the Keypad and is fed to U49, an 8 bit serial input latch, which decodes the selected keypad entries. The serial output is buffered and fed directly to the microprocessor. U40A and U40B are dual D Flip-Flops, which provide input to U51, one of eight decoders. U51 is fed to U49 to determine the correct keypad row. U48A, B, C are part of a quad AND Gate package and provide debounce circuitry for the keypad. Serial data from the microprocessor is fed to the input of U50, a serial 8-bit shift register to convert the serial display data to parallel data. The parallel data is connected to the display through J13. Main Board Circuit Diagram Sheet 9 Serial data from the microprocessor is fed to U53, an 8-bit serial to parallel shift register. This data provides the logic to control cassette detection LED’s and provides output signals to the status LED. U44 is a seven-stage ripple counter used to divide the 32 kHz clock to a frequency of 1024 Hz. This divided clock and data from U53 are fed to AND Gate U54A. This output provides the signal to drive the cassette light gates. U52 is a parallel load to serial 8-bit shift register, which receives input from the cassette light gates and other signals and provides serial data to the microprocessor. Main Board Circuit Diagram Sheet 10 Switch S1 is the main power switch and switches power to the OPTI 1 circuitry. Relay K1 is used to switch input power either from the internal battery or main external power supply. K2 is shown in the battery mode. When input power is supplied by the external power supply, fuse F1 provides overcurrent protection for the input power. U1 is a +5 Volt voltage regulator used to provide a regulated voltage supply to operate U2. IC U2 is a battery charge controller with LED D46 indicating the battery charge status. When D46 is solidly on, the battery is being charged and as the battery is fully charged, D46 will blink rapidly to indicate the charge status. If D46 blinks slowly, it indicates a defective battery or the battery temperature range has been exceeded. A thermistor is included in the internal battery pack to provide temperature input to the battery charge controller. 5-5 5 Electronic Description Main Board Circuit Diagram Sheet 11 Switched power, either from the external power supply or internal battery is fed to this circuit and is fused by F2 and F3. The +12 or +16 volts (battery/external power supply) is fed to Q4, a Darlington Pair, which provides supply voltage used to heat the sample measurement chamber (SMC). This input voltage is also fed to U12 a switching 5-Volt regulator, which supplies +5 volt power to the digital circuitry (ssv) and printer (pvcc). U14 is a +12V voltage regulator and feeds U20, a positive to negative voltage converter. The negative voltage output from U20 is fed to U25 to provide a regulated 5 volt output. An additional +5V voltage regulator U21 is also used. Main Board Circuit Diagram Sheet 12 The SMC temperature circuitry comprises two temperature control circuits that control the top plate heater and the optics module heater (lower plate). U26, a +5 volt regulator, provides a reference voltage used by both circuits. Operational amplifier U33D along with potentiometer R97 and R95 provide adjustment and control for both heater circuits. IC U32A and U32B provide the optic module heater control while U32C and U32D control the top plate heater. Main Board Circuit Diagram Sheet 13 This circuitry with inputs from sheet 11 controls the heater transistors in the SMC assembly. U34D and U43C are connected to the base and emitter of the optics module heater transistors. The signal OPTICCTRL (from sheet 11) causes the optics module heater transistors to conduct more when the temperature of the optics module needs to be increased to maintain regulation. IC U43A operates similarly as U34D and is connected to the top plate heater transistor. 5-6 5 Electronic Description Main Board Circuit Diagram Sheet 14 This circuit converts resistance provided by the NTC’s to a voltage for the temperature control of the top and bottom plate heaters located in the SMC. Op Amp U33A is fed an input from the NTC located in the optics module heater (bottom plate), an output voltage representative of the input NTC resistance. U33C is an identical circuit and receives the NTC input from the top plate NTC. U33B is fed the reference voltage input and provides the battery backup reference voltage (signal BCHRG2) which is fed to the battery charge controller located on sheet 1. Main Board Circuit Diagram Sheet 15 The two digital to analog converters (DAC’s) U17 and U23 control the sample sensor brightness or intensity for the sample sensor LED’s, LG1 and LG2 located in the SMC assembly. A comparison is made with the values stored in memory and provides a self-calibration. This signal is fed to each sample sensor anode, which controls the intensity of the sample sensor LED’s. The reference voltage signal is fed to IC U22B, which changes the impedance for the barometer circuit located on sheet 16. Main Board Circuit Diagram Sheet 16 The analog to digital converter U15 receives a parallel input of 8 analog signals and converts these voltages to a serial digital signal which is fed directly to the microprocessor via signal HIS0. The parallel input signals, which are labeled on the schematic, can be observed on the analyzer display when the diagnostic menu is selected. 5-7 5 Electronic Description Main Board Circuit Diagram Sheet 17 The internal barometer circuitry is comprised of the pressure transducer (BARO1) and the associated circuitry. The output voltage of BARO1 changes according to the changing atmospheric pressure. Op Amps U16 A, B, and C are used to calibrate the output pressure and correlate this pressure to barometric pressure standard. The output voltage at TP 15 represents barometric pressure in mmHg according to the following formula: Voltage + 3000 = BP mmHg 10 For calibration and adjustment use the procedure in Section 7, Adjustments. Main Board Circuit Diagram Sheet 18 The valves and stepper motors used in the OPTI 1 analyzer are controlled by the circuitry contained on this schematic sheet. U47, a serial to parallel latch, receives input from the microprocessor and provides the control signal for the gas valve along with other signals as labeled. U28 is a serial to parallel decoder which decodes the direction and enable signal for the valve drive, peristaltic pump and the printer. U36 is the driver for the valve drive motor and is fed signals directly from the microprocessor (TP0 - TP3) and an enable signal from U28 to provide four motor drive signals to valve drive stepper motor. The peristaltic pump is controlled by U27, a stepper motor controller, which feeds U35, the driver for the peristaltic pump stepper motor. Main Board Circuit Diagram Sheet 19 The internal printer is controlled by this circuitry. U45, a serial to parallel printer driver, receives an enable signal and converts serial data from the microprocessor to a parallel output which is fed to the printer print heads. The printer motor is controlled and driven by U31 and U42. U31, the printer controller receives an enable and direction signal which feeds U42, the motor driver for the printer stepper motor. 5-8 5 Electronic Description Main Board Circuit Diagram Sheet 20 The cassette detection infrared LED is monitored by U29 C and D and associated circuitry. The signal from the cassette light gate is fed to U29D and further converted by U29C. The resultant output is fed to the analog to digital converter on sheet 15 which indicates whether a cassette is in place. Main Board Circuit Diagram Sheet 21 This section of the main board circuitry provides signal identification for the connectors and interconnection to the other assemblies within the OPTI 1 analyzer. Also located on this schematic section is relay K2. Relay K2 is switched dependent on whether full windings or half windings are required for the peristaltic pump motor. 5-9 5 Electronic Description Main Board TEST Points Test Point TP1 TP2 TP3 TP4 TP5 TP6 TP7 TP8 TP9 TP10 TP11 TP12 TP13 TP14 TP15 TP16 TP17 TP18 TP19 TP20 TP21 TP22 TP23 TP24 TP25 TP26 TP27 TP28 5-10 Circuit Sheet 10 10 10 11 10 10 17 17 15 11 11 11 16 16 17 11 17 15 11 16 16 12 12 18 12 12 12 10 Measurement +5.00 ± 0.25 V 0.00 ± 0.02 V N/A +5.00 ± 0.2 V N/A N/A -3.419 ± 0.020 V +5.000 ± 0.005 V N/A N/A N/A N/A N/A N/A Ref. Formula +5.00 ± 0.20 V N/A N/A -5.00 ± 0.20 V N/A N/A +5.000 V ± .005 V N/A N/A N/A N/A N/A N/A Description Regulated + 5 Volt Ground Regulated + 5 Volt (VCC) Barometer Reference Voltage Precision Reference Voltage SMC Heater Supply Voltage (VHTR) Over Temperature Unregulated Supply Voltage (VS) Barometer Voltage Output Regulated + 5 Volt (+5V) Regulated - 5 Volt (-5V) Precision +5 volt reference voltage 5 Electronic Description Main Board Connectors J1 Pin 1 2 3 J2 Pin 1 2 3 4 5 6 7 8 J3 Pin 1 2 3 4 5 6 7 8 J4 Pin 1 2 3 4 Main Input Power Signal Description Positive Voltage from External Power Supply/Charger Ground Sleeve (Ground) Schematic Sheet Location Sheet 10 Sheet 10 Sheet 10 External Barcode (RJ45) Signal Description NC Shield (Connected to ground) Ground Supply Voltage NC NC Schematic Sheet Location Sheet 7 Sheet 7 Sheet 7 Sheet 7 Sheet 7 Sheet 7 Sheet 7 Sheet 7 Debug Connector Signal Description Not Used Schematic Sheet Location Sheet 4 Battery Connector Signal Description Battery Positive NC Battery Thermistor Battery Negative Schematic Sheet Location Sheet 10 Sheet 10 Sheet 10 Sheet 10 5-11 5 Electronic Description J5 Pin 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 5-12 Upgrade Board Connector Signal Description Ground Ground Ground Supply Voltage Ground Ground Supply Voltage +5 Volts -5 Volts Ground NC NC NC NC Schematic Sheet Location Sheet 4 Sheet 4 Sheet 3, 4 Sheet 4 Sheet 4 Sheet 4 Sheet 4 Sheet 4 Sheet 3 Sheet 4 Sheet 3 Sheet 4 Sheet 4 Sheet 4 Sheet 4 Sheet 4 Sheet 4 Sheet 4 Sheet 4 Sheet 4 5 Electronic Description J6 Pin 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 Printer Connector Signal Description Print Head 1 Print Head 2 Print Head 3 Print Head 4 Print Head 5 Print Head 6 Print Head 7 Print Head 8 Printer Supply Voltage PVcc Printer Supply Voltage PVcc Printer Supply Voltage PVcc Printer Home Position Ground Motor Drive 4 Motor Drive 2 Motor Drive 1 Motor Drive 4 HRA Supply Voltage Vcc HRC NC NC NC Keyboard Row NC Keyboard Column Schematic Sheet Location Sheet 11 Sheet 11 Sheet 11 Sheet 11 Sheet 11 Sheet 11 Sheet 11 Sheet 11 Bus Bus Bus Sheet 4 Sheet 11 Sheet 11 Sheet 11 Sheet 11 Sheet 11 Sheet 9 Bus Sheet 9 Sheet 11 Sheet 11 Sheet 11 Sheet 8, 9 Sheet 11 Sheet 8 5-13 5 Electronic Description J7 Pin 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 5-14 SMC Top Plate Signal Description Top Plate Heat Transistor Emitter 1 Heater Supply Voltage Top Plate Heat Transistor Base Top Plate NTC 1 Top Plate NTC 2 Top Plate Heat Transistor Emitter 2 NC NC NC Heater Supply Voltage Cassette Front Anode Cassette Rear Anode Cassette Front Collector Cassette Rear Collector Sense Rear Anode Sense Front Anode Sense Rear Collector Sense Front Collector Ground Ground Schematic Sheet Location Sheet 13 Sheet 11 Sheet 13 Sheet 12, 14 Sheet 12 Sheet 13 Sheet 11 Sheet 9 Sheet 9 Sheet 9 Sheet 9 Sheet 15 Sheet 15 Sheet 15 Sheet 15 Ground Bus Ground Bus 5 Electronic Description J8 Pin 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 J9 Pin 1 2 3 4 5 6 Optic Module Connector Signal Description Supply Voltage Vcc Reset +5 Volts - 5 Volts Ground Optics Transmit Data Optics Receive Data Ground Ground Optics Top Heater Transistor Emitter 1 Heater Voltage Vhtr Optics Top Heater Transistor Base Optics Top Heater Transistor Emitter 2 Optics Block Heater Transistor Emitter 1 Heater Voltage Vhtr Optics Block Heater Transistor Base Optics Block Heater Transistor Emitter 2 Optics Heater NTC 1 Optics Heater NTC 2 N/C N/C Cassette Light Gate Driver Cassette Light Gate Collector Sense Front Emitter Sense Rear Emitter Schematic Sheet Location Bus Sheet 4 Sheet 4 Bus Bus Bus Sheet 7 Sheet 7 Bus Bus Sheet 13 Bus Sheet 13 Sheet 13 Sheet 13 Bus Sheet 13 Sheet 13 Sheet 12, 14 Sheet 12 Sheet 9 Sheet 20 Sheet 16 Sheet 16 Pump Motor Connector Signal Description Pump -D Pump -C Relay K2 Pump -C Pump -B Pump -B Relay K2 Pump -A Schematic Sheet Location Sheet 18 Sheet 18 Sheet 18 Sheet 18 Sheet 18 Sheet 18 5-15 5 Electronic Description J10 Pin 1 2 J11 Pin 1 2 3 4 5 J12 Pin 1 2 3 4 5 6 7 8 9 10 11 12 13 14 5-16 Fan Control Connector Signal Description +12 Volts Fan Control Schematic Sheet Location Bus Sheet 18 Gas Valve 1 Connector Signal Description Supply Voltage Vs Gas Pressure 1 Gas Valve 1 (GV1) Supply Voltage Vs Auxiliary Schematic Sheet Location Bus Sheet 9 Sheet 18 Bus Sheet 18 Sample Measurement Chamber Connector Signal Description Ground Ground Home Position Light Gate SMC Light Gate Ground Ground Home Position LED Supply Voltage Vvss NC NC Valve -D Valve -B Valve -A Valve -C Schematic Sheet Location Bus Bus Sheet 9 Sheet 9 Bus Bus Sheet 9 Bus Sheet 18 Sheet 18 Sheet 18 Sheet 18 5 Electronic Description J13 Pin 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 LCD / LED Connector Signal Description Keypad Row A Keypad Column A Keypad Row B Keypad Column B Keypad Row C Keypad Column C Keypad Row D Keypad Column D LCD BL Status LED Green Status LED Yellow Status LED Red Ground Vcc Status LED Blink LCD Enable LCD Read Write LCD Reset LCD Data 0 LCD Data 1 LCD Data 2 LCD Data 3 LCD Data 4 LCD Data 5 LCD Data 6 LCD Data 7 Schematic Sheet Location Sheet 18 Sheet 9 Sheet 9 Sheet 9 Bus Bus Sheet 4 5-17 5 Electronic Description OPTI CCA Circuits The AVL OPTI CCA Optics Module and Main Board consist of the following electronic circuits. The schematic diagram for each part of the electronic description is located in Section 9 and is referenced to a sheet number corresponding to the electronic descriptions found in this section. SYSTEM FUNCTIONAL BLOCK DIAGRAM SHEET 1 CIRCUIT DIAGRAM Component Location CIRCUIT DIAGRAM SHEET 1 Optics Module Digital Circuit CIRCUIT DIAGRAM SHEET 2 Optics Module Amplifier Circuit CIRCUIT DIAGRAM SHEET 3 Optics Module Amplifier Circuit CIRCUIT DIAGRAM SHEET 4 Optics Module Inverter Circuit CIRCUIT DIAGRAM SHEET 5 Optics Module A/D Converter Circuit CIRCUIT DIAGRAM SHEET 6 Optics Module LED’s, Gasses & DAC Circuit CIRCUIT DIAGRAM SHEET 7 Optics Module LED’s and Lytes Circuit CIRCUIT DIAGRAM SHEET 8 Optics Module Interconnect Circuit 5-18 5 Electronic Description OPTI CCA Circuits Continued CIRCUIT DIAGRAM SHEET 1 Main Board RAM Battery Backup Circuit Clock Circuit CIRCUIT DIAGRAM SHEET 2 Main Board Microprocessor Circuit CIRCUIT DIAGRAM SHEET 3 Main Board 16 Bit EPROM Circuit CIRCUIT DIAGRAM SHEET 4 Main Board 16 Bit RAM Circuit CIRCUIT DIAGRAM SHEET 5 Main Board Barcode Interface Circuit CIRCUIT DIAGRAM SHEET 6 Main Board Keypad and Display Circuit CIRCUIT DIAGRAM SHEET 7 Main Board Cassette Detection Logic Circuit CIRCUIT DIAGRAM SHEET 8 Main Board Battery Charger Circuit CIRCUIT DIAGRAM SHEET 9 Main Board Voltage Regulator Circuit CIRCUIT DIAGRAM SHEET 10 Main Board SMC Temperature Control Circuit 5-19 5 Electronic Description OPTI CCA Circuits Continued CIRCUIT DIAGRAM SHEET 11 Main Board SMC Heater Transistor Circuit CIRCUIT DIAGRAM SHEET 12 Main Board SMC Heater Control Circuit CIRCUIT DIAGRAM SHEET 13 Main Board Digital to Analog Converter Circuit CIRCUIT DIAGRAM SHEET 14 Main Board A/D Converter Circuit CIRCUIT DIAGRAM SHEET 15 Main Board Barcode Circuit CIRCUIT DIAGRAM SHEET 16 Main Board Motor Drive Circuit CIRCUIT DIAGRAM SHEET 17 Main Board Printer Controller Circuit CIRCUIT DIAGRAM SHEET 18 Main Board Cassette Detect Circuit CIRCUIT DIAGRAM SHEET 19 Main Board Interconnect Circuit CIRCUIT DIAGRAM SHEET 20 Main Board Serial Interface Circuit CIRCUIT DIAGRAM SHEET 21 Main Board tHb/SO2 Laser/LED Driver Circuit 5-20 5 Electronic Description OPTI CCA Circuits Continued CIRCUIT DIAGRAM SHEET 22 Main Board tHb/SO2 Amplifier Circuit CIRCUIT DIAGRAM SHEET 23 Main Board Over Temperature Protection Circuit CIRCUIT DIAGRAM SHEET 1 Optics Module Interconnect Circuit CIRCUIT DIAGRAM SHEET 2 Optics Module LED’s, Gasses & Heater Circuit CIRCUIT DIAGRAM SHEET 3 Optics Module LED’s, ION & Heater Circuit CIRCUIT DIAGRAM SHEET 1 Optics Module SMC Heater / tHb Circuit CIRCUIT DIAGRAM SHEET 1 Display IR Board Interconnect Circuit CIRCUIT DIAGRAM SHEET 2 Display IR Board IR Circuit 5-21 5 Electronic Description Main Board - OPTI CCA TEST Points Test Point TP1 TP2 TP3 TP4 TP5 TP6 TP7 TP8 TP9 TP10 TP11 TP12 TP13 TP14 TP15 TP16 TP17 TP18 TP19 TP20 TP21 TP22 TP23 TP24 TP25 TP26 TP27 TP28 TP29 TP30 TP31 TP32 TP33 TP34 TP35 TP36 5-22 Circuit Sheet 8 8 8 8 8 14 9 17 14 14 14 14 14 8 8 9 17 15 13 15 15 13 9 9 9 12 14 14 14 14 9 10 10 10 16 10 Measurement +5.00 ± 0.25 V N/A N/A N/A N/A N/A +5.00 ± 0.2 V +5.000 ± 0.005 V N/A N/A N/A N/A N/A N/A N/A N/A N/A -3.57 ± 0.020 V N/A Ref. Formula N/A N/A N/A N/A +5.00 ± 0.20 V N/A N/A N/A N/A N/A -5.00 ± 0.20 V +5.000 V ± .005 V Description Regulated + 5 Volt Regulated + 5 Volt (VCC) Precision Reference Voltage SMC Heater Supply Voltage (VHTR) Over Temperature Unregulated Supply Voltage (VS) Barometer Reference Voltage Barometer Voltage Output Regulated + 5 Volt (+5V) Regulated - 5 Volt (-5V) Precision +5 volt reference voltage 5 Electronic Description Main Board - OPTI CCA TEST Points Test Point TP37 TP38 TP39 TP40 TP41 TP42 TP43 TP44 TP45 TP46 TP47 TP48 TP49 TP50 Circuit Sheet 9 8 10 21 21 21 21 21 21 21 23 23 22 22 Measurement 0.00 ± 0.02 V Description Ground 670 nm LED 670 nm LED + 850 nm Laser 850 nm Laser + 780 nm Laser 780 nm Laser + 5-23 5 Electronic Description Main Board Connectors J1 Pin 9 10 11 12 13 14 15 16 J2 Pin 4 5 6 J3 Pin 5 6 7 8 5-24 External Barcode (RJ45) Signal Description NC Shield (Connected to ground) Ground Supply Voltage NC NC Schematic Sheet Location Sheet 7 Sheet 7 Sheet 7 Sheet 7 Sheet 7 Sheet 7 Sheet 7 Sheet 7 Main Input Power Signal Description Positive Voltage from External Power Supply/Charger Ground Sleeve (Ground) Schematic Sheet Location Sheet 10 Sheet 10 Sheet 10 Battery Connector Signal Description Battery Positive NC Battery Thermistor Battery Negative Schematic Sheet Location Sheet 10 Sheet 10 Sheet 10 Sheet 10 5 Electronic Description J4 Pin 1 2 J5 Pin 1 2 J6 Pin 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 tHb/SO2 Connector Signal Description Schematic Sheet Location Sheet Sheet IRDA Connector Signal Description Schematic Sheet Location Sheet Sheet SMC Top Plate Signal Description Top Plate Heat Transistor Emitter 1 Heater Supply Voltage Top Plate Heat Transistor Base Top Plate NTC 1 Top Plate NTC 2 Top Plate Heat Transistor Emitter 2 NC NC NC Heater Supply Voltage Cassette Front Anode Cassette Rear Anode Cassette Front Collector Cassette Rear Collector Sense Rear Anode Sense Front Anode Sense Rear Collector Sense Front Collector Ground Ground Schematic Sheet Location Sheet 13 Sheet 11 Sheet 13 Sheet 12, 14 Sheet 12 Sheet 13 Sheet 11 Sheet 9 Sheet 9 Sheet 9 Sheet 9 Sheet 15 Sheet 15 Sheet 15 Sheet 15 Ground Bus Ground Bus 5-25 5 Electronic Description J7 Pin 7 8 9 10 11 12 J8 Pin 3 4 J9 Pin 6 7 8 9 10 J10 Pin 15 16 17 18 19 20 21 22 23 24 25 26 27 28 5-26 Pump Motor Connector Signal Description Pump -D Pump -C Relay K2 Pump -C Pump -B Pump -B Relay K2 Pump -A Schematic Sheet Location Sheet 18 Sheet 18 Sheet 18 Sheet 18 Sheet 18 Sheet 18 Fan Control Connector Signal Description +12 Volts Fan Control Schematic Sheet Location Bus Sheet 18 Gas Valve 1 Connector Signal Description Supply Voltage Vs Gas Pressure 1 Gas Valve 1 (GV1) Supply Voltage Vs Auxiliary Schematic Sheet Location Bus Sheet 9 Sheet 18 Bus Sheet 18 Sample Measurement Chamber Connector Signal Description Ground Ground Home Position Light Gate SMC Light Gate Ground Ground Home Position LED Supply Voltage Vvss NC NC Valve -D Valve -B Valve -A Valve -C Schematic Sheet Location Bus Bus Sheet 9 Sheet 9 Bus Bus Sheet 9 Bus Sheet 18 Sheet 18 Sheet 18 Sheet 18 5 Electronic Description J11 Pin 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 Optic Module Connector Signal Description Supply Voltage Vcc Reset +5 Volts - 5 Volts Ground Optics Transmit Data Optics Receive Data Ground Ground Optics Top Heater Transistor Emitter 1 Heater Voltage Vhtr Optics Top Heater Transistor Base Optics Top Heater Transistor Emitter 2 Optics Block Heater Transistor Emitter 1 Heater Voltage Vhtr Optics Block Heater Transistor Base Optics Block Heater Transistor Emitter 2 Optics Heater NTC 1 Optics Heater NTC 2 N/C N/C Cassette Light Gate Driver Cassette Light Gate Collector Sense Front Emitter Sense Rear Emitter Schematic Sheet Location Bus Sheet 4 Sheet 4 Bus Bus Bus Sheet 7 Sheet 7 Bus Bus Sheet 13 Bus Sheet 13 Sheet 13 Sheet 13 Bus Sheet 13 Sheet 13 Sheet 12, 14 Sheet 12 Sheet 9 Sheet 20 Sheet 16 Sheet 16 5-27 5 Electronic Description J12 Pin 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 5-28 Printer Connector Signal Description Print Head 1 Print Head 2 Print Head 3 Print Head 4 Print Head 5 Print Head 6 Print Head 7 Print Head 8 Printer Supply Voltage PVcc Printer Supply Voltage PVcc Printer Supply Voltage PVcc Printer Home Position Ground Motor Drive 4 Motor Drive 2 Motor Drive 1 Motor Drive 4 HRA Supply Voltage Vcc HRC NC NC NC Keyboard Row NC Keyboard Column Schematic Sheet Location Sheet 11 Sheet 11 Sheet 11 Sheet 11 Sheet 11 Sheet 11 Sheet 11 Sheet 11 Bus Bus Bus Sheet 4 Sheet 11 Sheet 11 Sheet 11 Sheet 11 Sheet 11 Sheet 9 Bus Sheet 9 Sheet 11 Sheet 11 Sheet 11 Sheet 8, 9 Sheet 11 Sheet 8 5 Electronic Description J13 Pin 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 LCD / LED Connector Signal Description Keypad Row A Keypad Column A Keypad Row B Keypad Column B Keypad Row C Keypad Column C Keypad Row D Keypad Column D LCD BL Status LED Green Status LED Yellow Status LED Red Ground Vcc Status LED Blink LCD Enable LCD Read Write LCD Reset LCD Data 0 LCD Data 1 LCD Data 2 LCD Data 3 LCD Data 4 LCD Data 5 LCD Data 6 LCD Data 7 Schematic Sheet Location Sheet 18 Sheet 9 Sheet 9 Sheet 9 Bus Bus Sheet 4 5-29 6 Test Programs and Diagnostics 6 Test Programs and Diagnostics OPTI 1 System The AVL OPTI 1 is equipped with diagnostic routines designed to assist the user in testing the operation of the each system component. These tests can be selected in the Diagnostics Menu. An overview of the diagnostic tests available is located in Section 2. This section describes each diagnostic test and the expected result. Diagnostic 1 Barometer The BARO test displays the system barometric pressure in a real-time mode. This test can be used to verify the internal pressure transducer is measuring correctly. The barometric pressure should be compared to the laboratory reference standard and calibrated accordingly. The operator can adjust the barometric pressure by entering the correct value in the SETUP - BARO menu. To calibrate the internal barometer follow the adjustment procedure outlined in Chapter 7. Battery The OPTI 1 may be powered by an internal battery pack or by an external battery charger, which is supplied with the analyzer. This test allows the user to determine the voltage of either the battery or external battery charger. When the charger is plugged in and connected to the analyzer, the display indicates the battery charger voltage. When the analyzer is powered from the internal battery, the displayed voltage is the battery voltage. When the internal battery powers the OPTI 1, the voltage must be greater than 11.2 volts for correct operation. While the external battery charger is connected, the internal battery is being charged. The main board contains the required circuitry to monitor and control the battery charging system. An LED is visible through the rear cover indicating battery charge status. A solid ON light indicates the battery is charging, while a fast flashing light (4 times per second) indicates the battery is fully charged. The battery LED will be OFF when the analyzer is not connected to the external charger. A slowly blinking light (once per second) indicates a very cold, very hot or defective battery. 6-1 6 Test Programs and Diagnostics Temperature The temperature diagnostic test displays three system-operating temperatures. T1 displays the SMC top plate temperature, T2 displays the SMC lower plate temperature and T3 displays the ambient temperature of the internal chassis and is used for the fan control. Circuitry on the main board, optics module, and SMC heater board provides temperature control for the analyzer. The following specifications are required for proper operation: T1 = 37.00° C ±0.05° C T2 = 37.15° C ±0.05° C T3 = Internal analyzer temperature Adjustment of the temperature circuitry is described in Chapter 7, Adjustments. Gas Pressure The newly installed calibration gas cylinder contains a gas pressure of 140-psi (9.7 bar) and is regulated and monitored by the gas module assembly. The gas pressure sensor portion of the module provides gas pressure status to the OPTI 1 analyzer. If the gas bottle pressure falls below approx. 35 psi, the switch closes (low pressure). The pressure needs to be above approx. 50 psi for the switch to open (high pressure). This diagnostic test indicates HIGH PRESSURE or LOW PRESSURE depending on the remaining gas cylinder pressure. When this diagnostic test is completed, an additional screen is displayed to allow replacement of the gas bottle and a gas purge sequence to be initiated by the user. Cleaning The OPTI 1 provides an automated cleaning routine that can be performed by selecting the diagnostic cleaning menu. Operator instructions are displayed to provide the user with a step-by-step guide for completion of the cleaning routine. First, the operator will be instructed to remove the peristaltic pump cartridge followed by pressing J. Next the operator will be instructed to place a syringe in the front peristaltic pump receptacle. The syringe should contain a mild cleaning solution such as a 30% bleach solution. Press J upon completion. The OPTI 1 will now instruct the operator to perform a wash. To do this, move the cleaning solution back and forth to flush the fluidic tubing. A tissue should also be placed at the inlet port of the SMC to catch any cleaning solution, which may be pushed, through the port. Press J upon completion. The operator will now be instructed to reinsert the pump cartridge. Press J to confirm this action and complete the cleaning procedure. 6-2 6 Test Programs and Diagnostics Flow Performing a flow test requires the use of a new cassette. The purpose of the flow test is to check proper operation of the pump, gas module and to detect leaks downstream from the pump. At the end of the flow test, a pass/fail as well as two flow numbers are displayed. If the flow numbers are 99.99, this part of the test failed and no number could be obtained. Press J to start the test. First, the gas pressure is turned on (tubing upstream from pump is pressurized). With the pump stopped, the fluid light gates L1 and L2 check for the buffer to remain in the cassette. If the buffer is pushed out, the test fails and the pump cartridge needs to be replaced. Now the pump starts pumping slowly, while L2 is looking for the trailing edge of the buffer. If the trailing edge is not detected within a certain time-out, the test fails. Check for leaks downstream from the pump. Replace the SMC gas seal and the pump cartridge receptacles. Once L2 has detected the trailing edge, L1 is looking for the trailing edge of the buffer. L1 must detect the trailing edge between 1.2 seconds and 5.2 seconds. The actual time is indicated by the first number on the display at the end of the test. If the number is less than 1.2 seconds, replace the gas bottle and check the output pressure of the gas module (2.5 psi ± 0.5 psi). Next, the system reverses the pump direction and measures the time of the buffer movement between LG1 and LG2. Since now the pump transports the buffer against the gas pressure, the time needs to be between 4.0 seconds and 8.0 seconds to pass the test. The second number displayed at the end of the flow test is the actual time. If the time is below 4.0 seconds, replace the gas bottle and check the output pressure of the gas module. If the time exceeds 8.0 seconds, replace the pump cartridge. Optics Test During the optics test, the optics module is warmed for 30 seconds to stabilize the optics system. Readings are then taken for each of the three channels for one minute. The test results are then calculated and printed to include the mean value of each channel’s intensity, noise and CV %. To test the proper function of the optics module and electronics, start the Optics test and insert an SRC (Low, Medium or High level). The values printed at the end of the test should be within the following limits: pH: PCO2: PO2: AVG 40’000 to 160’000 AVG 40’000 to 160’000 AVG 25’000 to 160’000 %DR: -1.0% to +1.0% %DR: -1.0% to +1.0% %DR: -1.0% to +1.0% %SEE: 0.0% to 0.1% %SEE: 0.0% to 0.1% %SEE: 0.0% to 0.1% If a value is out of range, repeat the test with another SRC. 6-3 6 Test Programs and Diagnostics Diagnostic 2 Printer Test The internal thermal printer can be tested using the printer test diagnostic. Selection of the printer test prints the following message: !”#$%’()*+,-./012345678 9 : ; < = > ? & ABCDEFGHIJKLMNOPQ RSTUVWXYZ abcdefghij klmnopqrstuvwxyz Keypad Test The keypad test when selected responds with the pressed key’s information displayed in the analyzer’s LCD display. Each key can be pressed successively until all keys have been tested. To exit this diagnostic routine, press E twice to return to the diagnostic menu. Display Test The display test can be selected to test the LCD display module. As this test is selected, each of the 24 characters in both rows of the display will be illuminated to indicate the function of the entire display module. The test will repeat every several seconds until E is pressed to end the test. Barcode Test The barcode test is designed to test the function of either the internal barcode or the optional external barcode. To perform the barcode test press J to start the test. Swipe a valid barcode label and the analyzer will display the barcode information on the LCD display. RS232 Test The RS232 Serial Interface test is included to test the serial interface function. To perform the test, first select the RS232 test. Displayed instructions will request the user to jumper pin 2 & 3 of the 9-pin serial interface connector on the rear panel of the analyzer. When the jumper is in place press J to start the test. Test results will be displayed indicating either PASS or FAIL. 6-4 6 Test Programs and Diagnostics Cover Test The cover test is used to test the function of the SMC cover. When the cover is closed the display will indicate Closed. Open the SMC Cover and the display should indicate Open. To exit this test, press the E key and return to the diagnostic menu. Fset The Fset menu permits viewing and modification of factory-set values stored in the EEPROM. At the prompt Password <ENTER> type 3350 <ENTER>. IDAC1 and IDAC2 are default DAC values for LG1 and LG2. The default DAC values are used only if no valid DAC values are available from the last measurement. The serial number displayed should match the last four digits of the serial number printed on the serial number label. pH, CO2 and O2 false light values are subtracted from the signal intensities and compensate for amplifier offset and photodiode dark currents. CDetect Low Limit is used as the reference value for the cassette detect sensor with SMC cover open and no cassette inserted. CDetect Up Limit is used as the reference value for the cassette detect sensor with SMC cover closed and cassette inserted. CDetect Low Offset is added to the CDetect Low Limit and is the threshold for a ‘cassette present’ condition. The CDetect Low Offset is always 100. CDectect Up Factor is multiplied with the CDetect Up Limit and is the threshold for a ‘cassette not present’ condition. The CDectect Up Factor is always 0.6. The Home Offset value indicates the number of motor steps between the home light gate position and the cassette valve home position. Gas Module Type: GR RE allows selection between ‘green dot’ and ‘red dot’ gas modules. Depending on its revision level, the gas module MB7001 has either a green or red dot applied. Generally, instruments with S/N < 1530 are marked with a red dot. PHR Correction: 160. These are factory-set optics calibration values which should not be changed. PCR Correction: 75. These are factory-set optics calibration values which should not be changed. 6-5 6 Test Programs and Diagnostics Diagnostic 3 Valve Test The gas module incorporates two gas valves: an On/Off valve on the high-pressure side which enables/disables the gas flow, and a 3-way valve, which connects the pump to the vent or to the gas module. Both valves are wired in parallel, i.e. they are activated/deactivated simultaneously. The valve test allows the user to test the gas valves. When the test is started, the analyzer displays the valve status. Pressing the right or left arrow key will turn OFF and ON the valve. Press E to exit the valve test. This test can be used for checking adequate gas flow and the output pressure of the gas module (2.5 psi ± 0.5 psi). Remove the pump cartridge and turn the valve on. The gas will free-flow out the rear pump cartridge receptacle. Connect a piece of tubing and submerge it in water to check for adequate flow. Vigorous bubbles indicate adequate flow. If flow is inadequate, replace the gas module. To check the output pressure, connect a calibrated, digital pressure meter to the rear pump receptacle. If the pressure is outside the specified limits, replace the gas module. Note that output pressure is factory-set and cannot be adjusted. Vdrive Test The valve drive test is used to test and determine the position and movement of the valve drive. Select the Vdrive menu to begin the test. Position 0 will be displayed. Press the right arrow key to select valve position 1 - 12 and press J. The valve drive will move to the desired position. To test the home position, select position 0 and press J. When the valve drive has reached home position, insert a new cassette. The cassette should fit over the valve drive pins without any resistance. The home position can be adjusted using the Fset menu. Increasing the number will adjust the home position counter-clockwise, decreasing the number will adjust it clockwise. Peristaltic Pump Test The pump test is used to test the function of the peristaltic pump. When the pump test is selected, the OPTI 1 will perform an automatic test of the peristaltic pump including clockwise and counter-clockwise rotation and a cycling of each pump speed used in normal operation. Once the test routine is started, correct pump operation can be observed by opening the printer access cover. Note that the pump speed is computer-controlled and adjustment is therefore not necessary. 6-6 6 Test Programs and Diagnostics LED Test The LED test displays the operating voltage of the internal LED’s and allows for verification of operation and performance. The LED test will display the following LED’s: LT1 LT2 L1 L2 CD Cassette seat detect front (SMC Top Cover) Cassette seat detect rear (SMC Top Cover) Front fluid detect LED Rear fluid detect LED Cassette detect LED Memory Test The Memory Test can be selected and the analyzer performs a check of the system memory. The analyzer displays MEMORY OK when the system successfully verifies the memory. Software Version Selecting the Software Version menu displays the currently installed software revision level. Pressing the right arrow key displays the software version for the optics module. Reports With the Reports menu, the user can select one of four reports to be printed. The Error Report saves the last 50 system errors and can be printed. When the number of system errors exceeds 50 the oldest error is deleted and replaced by the most recent error. A Calibration Report can also be printed which provides a calibration data summary of the stored patient and QC measurements. The third report is an SRC Report, which provides a summary of all stored SRC results. The Configuration Report prints out all user selectable settings. A printout of this report is particularly helpful prior to erasing memory (e.g. software update). 6-7 6 Test Programs and Diagnostics Error Report: AVL OPTI 1 ERROR REPORT 30-AUG-96 14:00 S/N: 1000 6-JUL- 96 08:45 ERROR - Cassette Misseat 2 Date/Time of error occurrence Type of error SRC Report: AVL OPTI 1 SRC Diagnostic Report 30-AUG-96 14:02 S/N: 1000 30Aug96 09:09 SRCID:742104 mV: 90.4 95.9 74.0 %DR: 0.06 0.09 0.06 %SEE: 0.01 0.02 0.01 Date/Time of start of SRC run Intensity/1000 must be > 7mV Drift must be < 0.3% for all parameters Noise must be < 0.2% for all parameters Calibration Report: AVL OPTI 1 Calibration Report 30-APR-98 14:00 S/N: 1000 30-AUG-96 11:18 LOT: 63407 mV Drift T W Ph 223.0 -2.4 33 102 PCO2 301.3 -0.4 33 103 PO2 625.9 -2.8 33 112 6-8 Date/Time of start of measurement Intensity/1000 must be > 7mV Drift during gas calibration, no limits apply T: time needed for calibration in sec. T must be less than 75 seconds W: ratio between warm QC and gas cal. pH and CO2 must be with 5 units of each other O2 needs to be >93 and < 130 If T is out of limits: Bad Calibration Error If W is out of limits: Bad Cassette Error 6 Test Programs and Diagnostics Configuration Report: AVL OPTI 1 Configuration Report 30-AUG-97 14:02 S/N: 1000 This report prints the current programmed instrument configuration. This should be printed before changing instrument software. Version: AOPX1.50dp Current Software Version Baro. Factor : 0.648 Patient Info Pat. ID : ON / Opt. Oper. ID : ON / Opt. Temp. : ON THb : ON Sex : ON Age : ON FIO2 : ON Hb Type : ON P50 : ON RQ : ON Def. THb : 15.0 g/dL Def. FIO2 : 0.21 Def. RQ : 0.84 Def. P50 : 26.7 mmHg Reference Limits - Barometric Pressure Sensor factor Patient Information parameters programming Controls Info Printouts Calculated Parameters Security Miscellaneous Correlation Factors Communications Language : English Battery Saver - 6-9 6 Test Programs and Diagnostics OPTI CCA System The AVL OPTI CCA is equipped with diagnostic routines designed to assist the user in testing the operation of the each system component. These tests can be selected in the Diagnostics Menu. An overview of the diagnostic tests available is located in Section 2. This section describes each diagnostic test and the expected result. Diagnostic 1 Barometer The BARO test displays the system barometric pressure in a real-time mode. This test can be used to verify the internal pressure transducer is measuring correctly. The barometric pressure should be compared to the laboratory reference standard and calibrated accordingly. The operator can adjust the barometric pressure by entering the correct value in the SETUP – BARO menu. To calibrate the internal barometer follow the adjustment procedure outlined in Chapter 7. Battery The OPTI CCA may be powered by an internal battery pack or by an external battery charger that is supplied with the analyzer. This test allows the user to determine the voltage of either the battery or external battery charger. When the charger is plugged in and connected to the analyzer, the display indicates the battery charger voltage. When the analyzer is powered from the internal battery, the displayed voltage is the battery voltage. When the internal battery powers the OPTI CCA, the voltage must be greater than 11.2 volts for correct operation. While the external battery charger is connected, the internal battery is being charged. The main board contains the required circuitry to monitor and control the battery charging system. An LED is visible through the rear cover indicating battery charge status. A solid ON light indicates the battery is charging, while a fast flashing light (4 times per second) indicates the battery is fully charged. The battery LED will be OFF when the analyzer is not connected to the external charger. A slowly blinking light (once per second) indicates a very cold, very hot or defective battery. 6-10 6 Test Programs and Diagnostics Temperature The temperature diagnostic test displays three system-operating temperatures. T1 displays the SMC top plate temperature, T2 displays the SMC lower plate temperature and T3 displays the ambient temperature of the internal chassis and is used for the fan control. Circuitry on the main board, optics module, and SMC heater board provides temperature control for the analyzer. The following specifications are required for proper operation: T1 = 37.00° C ±0.05° C T2 = 37.15° C ±0.05° C T3 = Internal analyzer temperature Adjustment of the temperature circuitry is described in Chapter 7, Adjustments. Gas Pressure The newly installed calibration gas cylinder contains a gas pressure of 140 psi (9.7 bar) and is regulated and monitored by the gas module assembly. The gas pressure sensor portion of the module provides gas pressure status to the OPTI CCA analyzer. If the gas bottle pressure falls below approx. 35 psi, the switch closes (low pressure). The pressure needs to be above approx. 50 psi for the switch to open (high pressure). This diagnostic test indicates HIGH PRESSURE or LOW PRESSURE depending on the remaining gas cylinder pressure. When this diagnostic test is completed, an additional screen is displayed to allow replacement of the gas bottle and a gas purge sequence to be initiated by the user. Version Selecting the Version menu displays the currently installed software revision level. Pressing the right arrow key displays the software version for the optics module. Reports With the Reports menu, the user can select one of four reports to be printed. The Error Report saves the last 50 system errors and can be printed. When the number of system errors exceeds 50 the oldest error is deleted and replaced by the most recent error. A Calibration Report can also be printed which provides a calibration data summary of the stored patient and QC measurements. The third report is an SRC Report, which provides a summary of all stored SRC results. The Configuration Report prints out all user selectable settings. A printout of this report is particularly helpful prior to erasing memory (e.g. software update). 6-11 6 Test Programs and Diagnostics Error Report: AVL OPTI CCA ERROR REPORT 30-APR-98 14:00 S/N: 1000 6-JUL- 96 08:45 ERROR - Cassette Misseat 2 Date/Time of error occurrence Type of error SRC Report: AVL OPTI CCA SRC Diagnostic Report 30-APR-98 14:00 S/N: 1000 30Aug96 09:09 SRCID:742104 mV: 90.4 95.9 74.0 %DR: 0.06 0.09 0.06 %SEE: 0.01 0.02 0.01 Date/Time of start of SRC run Intensity/1000 must be > 7mV Drift must be < 0.3% for all parameters Noise must be < 0.2% for all parameters Calibration Report: AVL OPTI CCA Calibration Report 30-APR-98 14:00 S/N: 1000 30-AUG-96 11:18 LOT: 63407 mV Drift T W Ph 223.0 -2.4 33 102 PCO2 301.3 -0.4 33 103 PO2 625.9 -2.8 33 112 Na+ 180.1 0.0 33 105 K+ 452.4 7.2 33 93 6-12 Date/Time of start of measurement Intensity/1000 must be > 7mV Drift during gas calibration, no limits apply T: time needed for calibration in sec. T must be less than 75 seconds W: ratio between warm QC and gas cal. pH and CO2 must be with 5 units of each other O2 needs to be >93 and < 130 If T is out of limits: Bad Calibration Error If W is out of limits: Bad Cassette Error 6 Test Programs and Diagnostics Configuration Report: AVL OPTI CCA Configuration Report 30-APR-98 14:00 S/N: 1000 This report prints the current programmed instrument configuration. This should be printed before changing instrument software. Version: AOPX1.50dp Current Software Version Baro. Factor : 0.648 Patient Info Pat. ID : ON / Opt. Oper. ID : ON / Opt. Temp. : ON THb : ON Sex : ON Age : ON FIO2 : ON Hb Type : ON P50 : ON RQ : ON Def. THb : 15.0 g/dL Def. FIO2 : 0.21 Def. RQ : 0.84 Def. P50 : 26.7 mmHg Reference Limits - Barometric Pressure Sensor factor Patient Information parameters programming Controls Info Printouts Calculated Parameters Security Miscellaneous Correlation Factors Communications Language : English Battery Saver - 6-13 6 Test Programs and Diagnostics Barcode The barcode test is designed to test the function of either the internal barcode or the optional external barcode. To perform the barcode test press J to start the test. Swipe a valid barcode label and the analyzer will display the barcode information on the LCD display. Diagnostic 2 Printer Test The internal thermal printer can be tested using the printer test diagnostic. Selection of the printer test prints the following message: !”#$%’()*+,-./012345678 9 : ; < = > ? & ABCDEFGHIJKLMNOPQ RSTUVWXYZ abcdefghij klmnopqrstuvwxyz Optics Test During the optics test, the optics module is warmed for 30 seconds to stabilize the optics system. Readings are then taken for each of the three channels for one minute. The test results are then calculated and printed to include the mean value of each channel’s intensity, noise and CV %. To test the proper function of the optics module and electronics, start the Optics test and insert an SRC (Low, Medium or High level). The values printed at the end of the test should be within the following limits: pH: PCO2: PO2: Na K Ca AVG 110,000 to 300,000 AVG 120,000 to 250,000 AVG 130,000 to 300,000 AVG 100,000 to 270,000 AVG 100,000 to 250,000 AVG 300,000 to 560,000 %DR: -1.0% to +1.0% %DR: -1.0% to +1.0% %DR: -1.0% to +1.0% %DR: -1.0% to +1.0% %DR: -1.0% to +1.0% %DR: -1.0% to +1.0% If a value is out of range, repeat the test with another SRC. 6-14 %SEE: 0.0% to 0.1% %SEE: 0.0% to 0.1% %SEE: 0.0% to 0.1% %SEE: 0.0% to 0.1% %SEE: 0.0% to 0.1% %SEE: 0.0% to 0.1% 6 Test Programs and Diagnostics LED Test The LED test displays the operating voltage of the internal LED’s and allows for verification of operation and performance. The LED 1 test will display the following LED’s: LT1 L1 L2 L3 Cassette seat detect front (SMC Top Cover) Front fluid detect LED Rear fluid detect LED Rear fluid detect LED Press ENTER to view the second portion of the LED test. The LED 2 test will display the following LED’s: Cover: Open or Closed SMC Cover CD Cassette detect LED Interface Test The RS232 Serial Interface test is included to test the serial interface function. To perform the test first select the RS232 test. Displayed instructions will request the user to jumper pin 2 & 3 of the 9-pin serial interface connector on the rear panel of the analyzer. When the jumper is in place press J to start the test. Test results will be displayed indicating either PASS or FAIL. IR Test The IR Loopback test is included to test the IR interface function. To perform the test first select the IR test. Test results will be displayed indicating either PASS or FAIL. Fan Test The FAN test allows the cooling fan to be turned ON and OFF using the analyzer keypad. Turn the FAN ON and check for proper airflow at the fan vents on the rear of the analyzer. 6-15 6 Test Programs and Diagnostics Fset The Fset menu permits viewing and modification of factory-set values stored in the EEPROM. At the prompt Password <ENTER> type 3350 <ENTER>. Select Instrument IDAC1, IDAC2 and, IDAC3 are default DAC values for LG1, LG2 and LG3. The default DAC values are used only if no valid DAC values are available from the last measurement. The serial number displayed should match the last four digits of the serial number printed on the serial number label. pH, O2 and CO2, Ca, K, Na false light values are subtracted from the signal intensities and compensate for amplifier offset and photodiode dark currents. CDetect Low Limit is used as the reference value for the cassette detect sensor with SMC cover open and no cassette inserted. CDetect Up Limit is used as the reference value for the cassette detect sensor with SMC cover closed and cassette inserted. CDetect Low Offset is added to the CDetect Low Limit and is the threshold for a ‘cassette present’ condition. The CDetect Low Offset is always 500. CDectect Up Factor is multiplied with the CDetect Up Limit and is the threshold for a ‘cassette not present’ condition. The CDectect Up Factor is always 0.8. Gas Module Type: GR This selection must always be set to GR. Home Offset value indicates the number of motor steps between the home light gate position and the cassette valve home position. PHR Correction: 160. These are factory-set optics calibration values that should not be changed. PCR Correction: 75. These are factory-set optics calibration values that should not be changed. 6-16 6 Test Programs and Diagnostics Fset Cont’d. Laser Parameters: WQC_SETTLE: WQC_NUM: SAM_SETTLE: SAM_NUM: K1: K2: K3: K4: K5: K6: K7: K8: K9: K10: K11: K12: F1: 0.000 F2: 0.000 F3: 0.000 F4: 1.000 F5: 1.000 F6: F7: F8: F9: F10: F11: F12: Note: Except F1 – F5, these are factory-set optics calibration values, which should not be changed. F1 through F5 are values, which are update during the tHb Calibration. The values, which are listed for F1 through F5, are default values and can be entered through FSET 6-17 6 Test Programs and Diagnostics Diagnostic 3 Valve Test The gas module incorporates two gas valves: an On/Off valve on the high-pressure side which enables/disables the gas flow, and a 3-way valve that connects the pump to the vent or to the gas module. Both valves are wired in parallel and they are activated/ deactivated simultaneously. The valve test allows the user to test the gas valves. When the test is started, the analyzer displays the valve status. Pressing the right or left arrow key will turn OFF and ON the valve. Press E to exit the valve test. This test can be used for checking adequate gas flow and the output pressure of the gas module (2.5 psi ± 0.5 psi). Remove the pump cartridge and turn the valve on. The gas will free-flow out the rear pump cartridge receptacle. Connect a piece of tubing and submerge it in water to check for adequate flow. Vigorous bubbles indicate adequate flow. If flow is inadequate, replace the gas module. To check the output pressure, connect a calibrated, digital pressure meter to the rear pump receptacle. If the pressure is outside the specified limits, replace the gas module. Note that output pressure is factory-set and cannot be adjusted. Flow Test Performing a flow test requires the use of a new cassette. The purpose of the flow test is to check proper operation of the pump, gas module and to detect leaks downstream from the pump. At the end of the flow test, a pass/fail as well as two flow numbers are displayed. If the flow numbers are 99.99, this part of the test failed and no number could be obtained. Press J to start the test. First, the gas pressure is turned on (tubing upstream from pump is pressurized). With the pump stopped, the fluid light gates L1 and L2 check for the buffer to remain in the cassette. If the buffer is pushed out, the test fails and the pump cartridge needs to be replaced. Now the pump starts pumping slowly, while L2 is looking for the trailing edge of the buffer. If the trailing edge is not detected within a certain time-out, the test fails. Check for leaks downstream from the pump. Replace the SMC gas seal and the pump cartridge receptacles. Once L2 has detected the trailing edge, L1 is looking for the trailing edge of the buffer. L1 must detect the trailing edge between 1.2 seconds and 5.2 seconds. The actual time is indicated by the first number on the display at the end of the test. If the number is less than 1.2 seconds, replace the gas bottle and check the output pressure of the gas module (2.5 psi ± 0.5 psi). 6-18 6 Test Programs and Diagnostics Next, the system reverses the pump direction and measures the time of the buffer movement between LG1 and LG2. Since now the pump transports the buffer against the gas pressure, the time needs to be between 4.0 seconds and 8.0 seconds to pass the test. The second number displayed at the end of the flow test is the actual time. If the time is below 4.0 seconds, replace the gas bottle and check the output pressure of the gas module. If the time exceeds 8.0 seconds, replace the pump cartridge. Display Test The display test can be selected to test the LCD display module. As this test is selected, each of the 24 characters in both rows of the display will be illuminated to indicate the function of the entire display module. The test will repeat every several seconds until E is pressed to end the test. VDrive Test The valve drive test is used to test and determine the position and movement of the valve drive. Select the VDrive menu to begin the test. Position 0 will be displayed. Press the right arrow key to select valve position 1 - 12 and press J. The valve drive will move to the desired position. To test the home position, select position 0 and press J. When the valve drive has reached home position, insert a new cassette. The cassette should fit over the valve drive pins without any resistance. The home position can be adjusted using the Fset menu. Increasing the number will adjust the home position counter-clockwise, decreasing the number will adjust it clockwise. Peristaltic Pump Test The pump test is used to test the function of the peristaltic pump. When the pump test is selected, the OPTI CCA will perform an automatic test of the peristaltic pump including clockwise and counter-clockwise rotation and a cycling of each pump speed used in normal operation. Once the test routine is started, correct pump operation can be observed by opening the printer access cover. Note that the pump speed is computer-controlled and adjustment is therefore not necessary. Keypad Test The keypad test when selected responds with the pressed key’s information displayed in the analyzer’s LCD display. Each key can be pressed successively until all keys have been tested. To exit this diagnostic routine, press E twice to return to the diagnostic menu. 6-19 7 Adjustments 7 Adjustments OPTI 1 System The AVL OPTI 1 is designed to require only minimal adjustments by field service personnel. The main board contains 6 electrical adjustments; however, field service personnel may make only three adjustments. Three potentiometers (R40, R100, and R136) are sealed and are factory adjustments only. These factory-only adjustments require special test fixtures and can only be performed at the factory. Field personnel may make the following adjustments: Barometric Pressure (R35) Sample Measurement Chamber Temperature (R97 and R95) For all adjustments first turn the analyzer power off. Open the analyzer lower housing by removing seven screws on the bottom of the analyzer. Set the analyzer upper housing on its side to access the keypad. Turn the analyzer power on. After completion of the adjustments, turn off the analyzer power before reassembly of the analyzer. Once reassembly is completed, turn the analyzer power on to resume normal operation. Barometric Pressure The operator through keyboard entry in the Setup mode may adjust barometric pressure. The displayed pressure may be adjusted ± 50 mmHg through keyboard entry. Use the following procedure if barometric pressure requires adjustment beyond the span of the keyboard entry. Using the analyzer keypad, select the Diagnostics Menu and then the barometer test to enable the barometric pressure. Locate R35 and adjust for the correct barometric pressure on the analyzer display. Sample Measurement Chamber Temperature Note: Prior to performing temperature adjustment turn on the analyzer power and allow a 10-minute warm-up to ensure temperature stability. Using the analyzer keypad, select the Diagnostics Menu and then the temperature test to enable the display of analyzer temperature. For temperature T1, locate R97 and adjust for a display of 37.00 ±. 0.03 ° C. After adjustment of R97 allow 5 minutes for temperature to stabilize. Locate R95 and adjust for a display of 37.15 ±0.03 ° C for temperature T2. Note: Always adjust temperature T1 before adjusting T2 as temperature T2 depends on temperature T1. R95 adjusts the difference between T1 and T2 which must be 0.15 ° C. 7-1 7 Adjustments OPTI CCA System The AVL OPTI CCA is designed to require only minimal adjustments by field service personnel. The main board contains nine electrical adjustments; however, field service personnel may make only three adjustments. Six potentiometers are factory adjustments only: R74 Barometer Slope R129 Top Cover Temperature R161 Optics Temperature R215 Laser Intensity R217 Laser Intensity R277 LED Intensity (tHb/SO2) These factory-only adjustments require special test fixtures and can only be performed at the factory. Field personnel may make the following adjustments: Barometric Pressure (R67) Sample Measurement Chamber Temperature (R126 and R124) For all adjustments first turn the analyzer power off. Open the analyzer lower housing by removing seven screws on the bottom of the analyzer. Set the analyzer upper housing on its side to access the keypad. Turn the analyzer power on. After completion of the adjustments, turn off the analyzer power before reassembly of the analyzer. Once reassembly is completed, turn the analyzer power on to resume normal operation. Barometric Pressure The operator through keyboard entry in the Setup mode may adjust barometric pressure. The displayed pressure may be adjusted ± 50 mmHg through keyboard entry. Use the following procedure if barometric pressure requires adjustment beyond the span of the keyboard entry. Using the analyzer keypad, select the Diagnostics Menu and then the barometer test to enable the barometric pressure. Locate R67 and adjust for the correct barometric pressure on the analyzer display. 7-2 7 Adjustments Sample Measurement Chamber Temperature Note: Prior to performing temperature adjustment turn on the analyzer power and allow a 10-minute warm-up to ensure temperature stability. Using the analyzer keypad, select the Diagnostics Menu and then the temperature test to enable the display of analyzer temperature. Locate R126 and adjust for a display of 37.00 ±. 0.03 ° C for temperature T1. After adjustment of R126 allow 5 minutes for temperature to stabilize. Locate R124 and adjust for a display of 37.15 ±. 0.03 ° C for temperature T2. Note: Always adjust temperature T1 before adjusting T2 as temperature T2 depends on temperature T1. R124 adjusts the difference between T1 and T2 which must be 0.15 ° C. 7-3 8 Troubleshooting 8 Troubleshooting OPTI System The AVL OPTI and OPTI CCA analyzers are equipped with sensors and diagnostic software that control, monitor and detect analyzer system status. Operator messages indicating system status are displayed to provide the operator with useful information to determine the cause of detected system errors or malfunctions. System errors are also stored in memory and an error report can be printed that lists the last 50 status messages. Displayed Alarms There are four types of displayed system status messages: SYSTEM STOP ALARMS These alarms indicate system conditions that must be resolved before system operation can be continued. SYSTEM ERROR ALARMS These errors occur during sample analysis and are specific to the current sample being analyzed. Error alarms indicate the status of the current measurement or additional required operator entry. SYSTEM WARNING MESSAGES System warning messages notify the operator of conditions requiring operator intervention to complete the current measurement. SYSTEM INFORMATION MESSAGES These messages describe operator instructions or confirm completion of operator selections. The following section describes each displayed message and the causes and conditions that generate this message. Further in the chapter probable causes and fault isolation are provided for each status message. 8-1 8 Troubleshooting SYSTEM STOP MESSAGES STOP - Temp Out of Range Please Wait Menu<ENTER> When the sample measurement chamber bottom plate temperature is outside the range of 37.10° C to 37.20° C or the top plate is outside the range of 36.95° C to 37.05° C this error is displayed. Normal operation can only be resumed when the temperature is in range. To check the system temperature, access the DIAG1 Temperature menu to display system temperature. Temperature T1 indicates top plate temperature while T2 indicates bottom plate temperature. If this error condition remains, perform temperature adjustment procedure as described in Section 7, Adjustments. STOP - Low Battery Recharge Now During prolonged periods of use, the battery charge may become depleted and normal operation can no longer be continued. This message will be displayed when the battery voltage is less than 11.2 volts. System operation is halted until the error condition is resolved in order to prevent the possible memory corruption. Replace the battery pack with a fully charged spare or connect the battery charger to recharge the battery pack. STOP If the battery voltage drops below 10.5 V, the unit will lock up to prevent memory corruption. To resume operation, connect the instrument to the power supply and cycle the power. - Low Battery Recharge and Cycle Power STOP - Low Gas Replace Now 8-2 <ESC> An internal gas cylinder is used to provide a calibration gas source that is used when calibrating each sample cassette. When the gas pressure sensed by the internal gas manifold assembly detects pressure below 35 psi, too low to perform a successful calibration, this error is displayed. Replace the gas cylinder with a full cylinder to continue normal operation. Access DIAG1 Gas to display the current gas pressure status. 8 Troubleshooting STOP - Need New SRC Data Must Run SRCs STOP - <ESC> System Error Cycle Power for Reset STOP - New Cassette Lot Must Run Controls <ESC> STOP - Memory Error Database Deleted <ESC> Standard Reference Cassettes (SRC’s) are measured to verify the analyzer’s measuring performance. In analyzer programming, the key operator may select QC Lockout for 1, 2 or 3 levels. Depending on the selection, 1, 2 or 3 levels of SRC’s are required to be run daily. If no SRC records for the last 24 hour period exist in the database, this message is displayed. Performing SRC measurements of the levels required will remove this error message. The STOP System Error is displayed when the operating software detects incorrect communication between the main microprocessor and the optics module microprocessor and halts routine operation. This indication may occur if the system software becomes corrupted. Generally, cycling power OFF and then ON can restart the system. The OPTI will restart and complete the start-up initializing routine for normal operation. When “New Lot” or “Controls” is selected under QCLockout, the OPTI will detect a difference in the cassette lot number from the previous cassette. When the difference is detected, the Stop - New Cassette Lot message is displayed. To continue the operator must verify the new lot of cassettes by measuring QC material. Vital analyzer parameters are stored in the system’s EEPROM. Upon each power on, this EEPROM data is checked against its copy in RAM. If the values do not match due to memory corruption, this error is displayed. The analyzer will perform a complete reinitialization which includes deletion of the stored data. In addition, for diagnostic purposes, a report outlining the old values in RAM and the new initialized values can be printed if J is pressed. 8-3 8 Troubleshooting SYSTEM ERROR MESSAGES ERROR - Unstable Sensors Discard Cassette <ESC> ERROR-Cassette MisSeat 1 Reinsert or <ESC> ERROR-Cassette MisSeat 2 Reinsert or <ESC> ERROR-Cassette MisSeat 2 Discard Cassette ERROR - Bad Cassette Discard Cassette 8-4 <ESC> <ESC> This error occurs during a sample measurement. In this condition the cassette successfully calibrated and detected correct aspiration of the sample. During a sample measurement, the OPTI monitors each of the sensor’s output for an endpoint. When two or more parameters continue to drift and do not reach endpoint this message is displayed. When the OPTI detects that the SMC Cover has been closed, the SMC top light gates LT1 and LT2 are checked to ensure proper seating of the cassette. If a cassette has been misseated, the spring loaded top plate is not in its correct position and causes either of the SMC top light gates to sense this misseat condition. Open the SMC cover and reposition the cassette. If error persists, insert a new Once a cassette has been inserted into the SMC and the cover is closed, the front and rear light gates LG1 and LG2 detect buffer is present in the cassette. Normally the buffer is removed during the calibration process. When this message is displayed, the OPTI assumes the cassette is seated improperly or the cassette is not sealed firmly to the SMC port. Open the SMC cover and reposition the cassette. If reinserting the cassette does not eliminate the Cassette Misseat Error, this message will be displayed. During cassette calibration, each sensor is checked for an endpoint. If all sensors reach endpoint, calculations are performed for each sensor comparing calibration information to initial optics data. This message is displayed if two or more of the sensor calculations are outside the limits. 8 Troubleshooting ERROR - Bad Calibration Discard Cassette ERROR - Cal Expired Discard Cassette ERROR - <ESC> <ESC> Dirty Optics Clean Cass./Optics <ESC> ERROR - Bad Sensors Discard Cassette <ESC> ERROR - Invalid Barcode Continue <ENTER> ERROR - Invalid Barcode Check TimeDate <ENTER> If two or more sensors did not reach a stable endpoint, the calibration is considered “failed” and this message is displayed. Following insertion of a cassette, a one-point cassette calibration will occur. The cassette will remain calibrated for a 10-minute period. If no measurement occurs within 10 minutes, the cassette will be recalibrated. This process will continue for up to 30 minutes. If no sample has been aspirated during this thirty-minute interval, the cassette must be discarded. For OPTI CCA calibration expires in 10 minutes. After the cassette has been repositioned following an initial Misseat 2 Error, the sample light gates LG1 and LG2/(LG3) monitor the transition in the cassette from buffer to air. When the cassette buffer is removed, the light gates can not detect this buffer-to-air transition. The OPTI assumes that the optics are dirty when no buffer transition is detected. After insertion of a new cassette, an optics test is performed. Each sensor must provide a reading of greater than 10,000 counts for an adequate sensor output. If any sensor fails to meet these criteria, no sample measurement can be performed. A manual optics test can be performed in System Diagnostics to confirm the cassette performance. During sample analysis the barcode is read. If the analyzer is unable to determine the barcode information or the system software does not support the barcode format this message will be displayed. The barcode information is checked to determine if the barcode is a cassette, compares the checksum, and compares calibration parameters to expected ranges. A barcode test can be performed by accessing DIAG Barcode and swiping a valid barcode. The barcode data will be displayed. The OPTI currently supports the following barcode formats: Codabar, Code39, Interleaved 2 of 5, UPC, EAN. If the expiration date does not match the current date/time setting, this message is displayed. 8-5 8 Troubleshooting ERROR - No Pat. ID Entry Edit Data<ENTER>or<ESC> ERROR - No Op. ID Entry Edit Data<ENTER>or<ESC> ERROR - Expired Barcode Check Time Date <ENTER> ERROR - Possible Clot Discard Cassette ERROR - <ESC> Invalid Sample Discard Cassette <ENTER> 8-6 During analyzer setup and programming a selection can be made to require entry of patient ID for each sample analyzed. The analyzer will not report sample results until a patient ID is entered. Enter a patient ID to obtain a printed sample report. During analyzer setup and programming a selection can be made to require entry of operator ID for each sample analyzed. The analyzer will not report sample results until an operator ID is entered. Enter an operator ID to obtain a printed sample report. Each measurement cassette and Standard Reference Cassette (SRC) is provided with a barcode label that must be read prior to insertion of the cassette into the analyzer. The barcode contains measurement information, lot number and expiration date. This message is displayed when the expiration date indicated in the barcode information has been exceeded. To resolve this error, first check analyzer time and date settings. If the settings are correct remove the SRC or measurement cassette and discard the expired cassette. Select an in-date cassette to perform the measurement. During sample aspiration the SMC light gates LG1 and LG2 (LG3 for OPTI CCA) are used to detect proper sample aspiration. When LG1 detects sample has been aspirated at the beginning of the sensor path and LG2/LG3 fails to detect sample aspiration at the end of the sensor path, the warning of a possible clot is displayed. In the OPTI-trol mode, a blood sample must be measured after the OPTI-trol measurement. 8 Troubleshooting ERROR - Gas Expired <ESC> ERROR - No ID Entry Edit Data<ENTER>or<ESC> Each gas bottle is provided with a barcode label, which must be read prior to insertion of the gas bottle. This message is displayed when the expiration date on the barcode has been exceeded. First check whether the analyzer time/date setting is correct. Then replace gas bottle with an in-date bottle (check expiration date on new bottle). Note: The in-use expiration date may occur prior to the expiration date printed on the gas bottle label. During analyzer setup, a selection can be made to require entry of patient ID and operator ID for each sample analyzed. If this function is selected, the analyzer will not report sample results, until a patient and/or operator ID is entered. 8-7 8 Troubleshooting SYSTEM WARNING MESSAGES WARNING - Low Gas 1 Test Remaining <ESC> WARNING – Gas Expires Soon! Continue <ENTER> No Sample Detected - Low Battery 1 Test Remaining 8-8 The Gas expiration date is monitored based on the barcode that was swiped into the OPTI’s memory at the time of the last gas canister change. This message alerts the operator that there is 2 weeks usage remaining for the gas tank. The operator can continue normal operation by pressing ENTER. The OPTI requires a minimum of 80µL (OPTI CCA 120µL) of blood or a minimum of 110µL (OPTI CCA 120µL) of clear fluid such as QC material to perform a measurement. Sample light gates LG1 and LG2/(LG3) in the sample measurement chamber (SMC) detect the leading and trailing edge of the sample in order to determine if the minimum amount of sample is aspirated. When LG1 does not detect sample aspiration, this message is displayed. WARNING WARNING The gas manifold assembly continually monitors gas pressure when the OPTI is in use. When the pressure is less than 25 psi, the Low Gas warning is displayed to indicate that the remaining gas pressure can perform only one additional sample measurement. <ESC> During battery operation of the analyzer, the OPTI continually monitors the battery voltage. When the operating voltage of the battery is less than 11.8 volts and greater than 11.2 volts a battery warning is displayed. This informs the operator to plug in the battery charger or install a fully charged battery prior to an interruption of system operation. 8 Troubleshooting WARNING-Bad Sensor - pH Continue<ENTER>or<ESC> WARNING-Bad Sensor- CO2 Continue<ENTER>or<ESC> These warning messages indicate a single sensor has an output of less than 7,000 counts. The warning message may appear either during the calibration or sample measurement process. If this message occurs during sample measurement, check sample preparation. These warning messages are displayed and are specified for each measurement parameter. When this warning is displayed no sample measurement results will be reported for the parameter identified, however results will be reported for the other parameters. WARNING-Bad Sensor - O2 Continue<ENTER>or<ESC> WARNING-Bad Sensor- Na Continue<ENTER>or<ESC> WARNING-Bad Sensor - K Continue<ENTER>or<ESC> WARNING - Unstable pH Check Cassette WARNING - Unstable PCO2 Check Cassette WARNING - This warning may be displayed during a calibration or measurement sequence and indicates a single sensor failed to reach an acceptable endpoint. The warning is displayed and indicates the sensor that does not reach endpoint in the right column of the display. If this warning occurs during sample analysis, check the cassette for air bubbles in the cassette upon the completion of the measurement sequence. If the warning is displayed during calibration, the parameter having the warning will not provide results. Unstable PO2 Check Cassette WARNING - Unstable Na Check Cassette WARNING - Unstable K Check Cassette 8-9 8 Troubleshooting WARNING- Bubble Detected Check Cassette <ENTER> WARNING - Check Pump Continue <ENTER> or<ESC> WARNING - Check Pump Please remove cassette WARNING - No PatID Entry Continue <ENTER> WARNING - No Op ID Entry Continue <ENTER> WARNING - No ID Entry Continue 8-10 <ENTER> This message is displayed when the OPTI has detected an air bubble during sample aspiration. This message is displayed before the final results are displayed so the operator can be alerted to check the cassette for possible bubbles. If an air bubble is sensed during sample aspiration, the aspiration routine is retried for up to 3 additional cycles. If all aspiration retries fail, the analyzer will display a No Sample Error. For a description of the aspiration cycle and bubble detect routine, refer to the “ERROR - No Sample” description. During a cassette calibration cycle, the peristaltic pump will pump calibration gas into the cassette. When the gas is being pumped into the cassette, the buffer in the cassette is pumped into the cassette reservoir. SMC light gates LG1 and LG2 are monitored during this time to detect the buffer evacuation in a precise time interval. If the light gates detect an improper evacuation of the buffer, the pump warning message is displayed. If pressing J twice to the above error message does not resolve the Check Pump condition, this message will be displayed. During analyzer setup and programming a selection can be made to require entry of a patient and/or operator ID for each sample analyzed. The analyzer will not report sample results until a patient and/or operator ID is entered. Enter the required ID to continue sample measurement. 8 Troubleshooting SYSTEM INFORMATION MESSAGES No SRC Record (Level X) SRC Database Deleted (Level X) Please Wait Printing Stopped Please Leave Cassette & Close the Cover Terminate Operation: N Y When the operator has requested a printed report of the SRC measurement statistics and no records are stored in the SRC database this status message will be displayed. The displayed message also identifies the applicable SRC Level for which no SRC records exist. When the OPTI analyzer detects a new SRC ID, the operator is prompted to delete all current SRC data (password protected). As the operator enters the correct password and replies E to delete the SRC data, the analyzer prints the SRC statistical report and confirms deletion of the SRC data with this displayed message. While the OPTI internal thermal printer is printing a report, the operator may press E to abort printing. The analyzer will display this message until the current report is restarted or the operator presses E a second time to abort the print cycle. Following a sample measurement, the cassette reservoir is locked and the sample waste is sealed inside the cassette. This reservoir locking prevents the biohazardous waste from leaking and forms a sealed container for the sample waste. If the SMC Cover is opened prematurely in order to discard the cassette and the valve drive has not been turned to the locking position, this message is displayed. Close the cover and wait for the analyzer to prompt for cassette removal. During an operating sequence of the OPTI, the operator may select to abort the current operation. To do this, press E and then “0”. The analyzer then displays this message and waits for a key response (J or E )before terminating the current operation. 8-11 8 Troubleshooting Not Enough Records (Level X) Control Database Deleted (Level X) Operation: Halted ! Please Close the Cover The OPTI stores QC results and SRC results in an internal database. Cumulative statistics printed or displayed in the system software are calculated based on the stored data. A minimum of 2 data points is required in order to calculate statistical results. This message is displayed when too few records exist in the database to provide a result. Enter additional QC or SRC measurement data when this message is displayed. When a new lot of QC material is to be run, the assay or package insert values must first be entered (under password control). As the operator enters a new lot number, the OPTI will prompt the operator to print and then delete the old QC data. As the QC records are deleted, this message is displayed to confirm the deletion of the QC database has occurred. If the sample measurement chamber (SMC) cover is opened during a critical time (during calibration, aspiration or measurements), this message will be displayed. Close the SMC cover to continue normal operation. New SRC Lot ! During an SRC measurement, the SRC ID number is checked by the barcode information. If the OPTI analyzer detects an ID number that differs from that of previous SRC measurements, the New SRC warning is displayed. The operator may select J to continue or E to exit the SRC measurement process. Cassette Expired If an operator swipes a cassette barcode and the expiration date of the cassette has been exceeded, this message will be displayed. Check the analyzer TIME and DATE for the correct entry. If the time and date are correct replace the cassette lot with an in date lot. <ESC> 8-12 8 Troubleshooting Troubleshooting The AVL OPTI provides display messages indicating the status of the analyzer. The previous section describes the conditions in which error, warning or system information messages may occur. This section contains troubleshooting tables to define possible causes and instructions for remedial action. In addition, troubleshooting flowcharts are provided for the most common error occurrences. Error Message WARNING -Low Battery 1 Test Remaining <ESC> Possible Causes Battery voltage >11.2 and < 11.8 volts. Remedial Action Replace battery pack with a fully charged battery. Plug in battery charger, check Charge LED is ON. The charger will charge a depleted battery while connected to AC power in approximately 6 hours. If the battery will no longer accept a charge replace the battery with a new battery pack. Check the charge LED status; a solid ON indicates the battery is charging, a fast flashing ON LED (4 times/second) indicates battery is fully charged. If the charge LED is OFF this indicates the charger is not connected. Replace charger or replace the main board. A slow flash (1 flash per second) indicates the battery voltage is below 9.5 volts or the battery thermistor senses an excessively hot or cold battery. Replace the battery if this indication persists. 8-13 8 Troubleshooting Error Message STOP - Low Battery Recharge Now Possible Causes Battery voltage <11.2 volts and the analyzer will stop operating. Remedial Action See above troubleshooting hints. When the battery voltage rises above 11.2 volts the analyzer will resume operation. STOP - Low Battery Recharge and Cycle Power Battery voltage drops below 10.5 V. Connect power supply and cycle power. STOP - System Error Cycle Power for Reset At power-up software is not initialized correctly. Turn power OFF and ON to retry. Check interconnect cables for a secure connection. Perform system initialization. Replace SMC module. Replace main board. STOP - Temp Out of Range Please Wait Menu <Enter> SMC top plate temperature T1 must be 37.00° C ± 0.05 and bottom plate T2 must be 37.15° C ± 0.05. Check fuse F2. Perform Temperature Diagnostic check to determine current temperature. Perform Temperature circuit adjustment. Install SW Version ≥ 1.10 Replace SMC module Replace main board. Replace heater board in SMC cover. 8-14 8 Troubleshooting Error Message Possible Causes Remedial Action WARNING - Low Gas 1 Test Remaining The gas cylinder is near empty Replace gas cylinder with a full and has only enough pressure to cylinder. perform one additional measurement. Replace gas module assembly. STOP - Low Gas Replace Now <ESC> The pressure of the gas cylinder is empty. No measurements can be made. See troubleshooting hints above. STOP - Memory Error Database Deleted <ESC> RAM data has been changed, corrupted, or lost, or a new software version has been installed Press ESC and all factory settings will be copied from EEPROM to RAM. ERROR - Invalid Barcode Continue <ENTER> During barcode swipe, the read barcode information is not in the correct format (i.e. SRC barcode is swiped for a normal measurement). Check the correct menu is selected. Select a cassette barcode and swipe to determine barcode function. Perform barcode diagnostic test. Perform system initialization. Check barcode format is supported by the OPTI. Clean barcode lens. Replace internal or external barcode assembly. Replace main board. ERROR - Invalid Barcode Check Time/Date <ENTER> The expiration date does not match the current date setting Check correct Time/Date setting 8-15 8 Troubleshooting Error Message Possible Causes Remedial Action STOP - Need New SRC Data Must run SRC’s <ESC> If QC lockout has been selected to require daily SRC measurement, SRC’s must be run at least once per day. STOP - New Cassette Lot Must Run Controls <ESC> New Lot has been selected in QC Lockout and a new lot of cassettes was detected. The new lot of cassettes must be verified with quality control material. Select a cassette from the current lot and run measurement. Top cover light gates LT1 and LT2 detect a cassette not seated properly. Open cover and reposition the cassette. ERROR - Cassette Misseat 1 Reinsert or <ESC> Run required SRC daily measurements. Run QC material to verify the new lot of cassettes. Retry with new cassette. Perform LED test in system Diagnostics. Replace SMC cover. ERROR - Cassette Misseat 2 Reinsert or <ESC> During displacement of buffer, the rear light gate LG2 fails to detect a transition from buffer to air. Open cover and reposition the cassette. Retry. Discard cassette and retry with a new cassette. Replace SMC seal. Perform flowtest. Check pump cartridge is seated correctly, replace if necessary. Check valve drive position. Check for adequate gas flow 8-16 8 Troubleshooting Error Message Possible Causes Remedial Action ERROR - Cassette Misseat 2 Discard Cassette <ESC> If above error occurs repeatedly, this message is displayed. Discard cassette. ERROR - Bad Sensors Discard Cassette <ESC> During a cassette calibration, the optode sensors did not have sufficient signal output. Check cassette foil pouch for tears or leaks. Make sure cassette pouch was opened within 10 minutes from start of measurement. Clean glass optics plate. Try a new cassette. ERROR - Bad Calibration Discard Cassette <ESC> Two or more sensors failed to reach endpoint during calibration cycle. Discard cassette and retry with a new cassette. Check for leaks in fluidic tubing. Check for proper seating of pump cartridge. ERROR - Unstable Sensors Discard Cassette Sample preparation or sample with high or extreme measurement characteristics. Check cassette for air bubbles or voids over sensors. Thoroughly mix sample prior to measurement. ERROR - Expired Barcode Check Time/Date <ENTER> Cassette date code has expired. Use cassette with an in-date expiration. Check for current date and time. ERROR - Bad Cassette Discard Cassette <ESC> Cassette optics characteristics failed during cassette check. Check cassette foil pouch for leaks or tears. 8-17 8 Troubleshooting Error Message Possible Causes Remedial Action Make sure cassette pouch was opened within 10 minutes from start of measurement. Discard cassette. ERROR - Cal Expired Discard Cassette <ESC> ERROR - Dirty Optics Clean Cass./Optics <ESC> After calibration of the cassette a sample must be run within 10 minutes. If no sample is entered, the analyzer will perform another calibration. This process will be repeated for up to 30 minutes, then the cassette must be discarded. OPTI CCA calibrations expire After the first 10 minute period. Run a sample measurement within 30 minutes after cassette insertion. Sample light gates LG1 and LG 2 fail to calibrate. Wipe residual from cassette prior to insertion. SMC Optics may have become dirty with blood residual. Clean glass optics plate. Open a new cassette and retry. Discard cassette and try new cassette. Perform Diagnostic LED Test. Make sure green LED’s in SMC cover light up. Replace SMC cover. ERROR - Gas Expired <ESC> Gas bottle barcode expired. Replace with new gas bottle. In-use life expired. Set correct time/date. Instrument time/date setting incorrect. 8-18 8 Troubleshooting Error Message Possible Causes Remedial Action ERROR - No Pat. ID Entry Edit Data <ENTER> or <ESC> Patient ID is selected as a required entry for all measurements. Press <ENTER> to continue and enter a Patient ID for the sample. Results will be displayed and will not be printed until an ID has been entered. ERROR - No Op. ID Entry Edit Data <ENTER> or <ESC> Operator ID is selected as a required entry for all measurements. Press <ENTER> to continue and enter an Operator ID for the sample. Results will be displayed and will not be printed until an ID has been entered. ERROR - No ID Entry Edit Data <ENTER> or <ESC> Patient and operator ID was selected as a required entry for all measurements. Press <ENTER> to continue and enter an operator and patient ID for sample results. WARNING - Bubble Detected Check Cassette LG 1 detected sample aspiration and detected air prior to sample reaching LG 2. Check cassette for air bubbles or insufficient sample volume. ERROR Possible Clot Discard cassette <ESC> LG 1 detected sample aspiration and analyzer timed out before sample reached LG 2. Check sample cassette for clots. Run properly heparinized sample. Check peri pump cartridge and replace as necessary. ERROR - Invalid Sample Discard Cassette <ENTER> During OPTI-trol measurement, the sample detected was not a blood sample. Measure a blood sample after OPTI-trol measurement. WARNING Bad Sensor - pH Continue <ENTER> or <ESC> Air bubbles in cassette or void over sensor. Possible poor sample preparation or corrupt optics giving < 7,000 counts. Check for air bubbles in cassette. WARNING Bad Sensor– PCO2 Remix sample and rerun with new cassette. 8-19 8 Troubleshooting Error Message Possible Causes Remedial Action WARNING Bad Sensor – PO2 Cycle analyzer power. WARNING Bad Sensor – Na Open SMC cover and check operation of Blue LED’s. WARNING Bad Sensor – K Replace the SMC Assembly. WARNING - Unstable pH Check Cassette Possible poor sample preparation or air bubbles in cassette. Check for air bubbles in cassette. Remix sample and rerun with new cassette. WARNING - Unstable PCO2 WARNING - Unstable PO2 WARNING - Unstable Na WARNING - Unstable K 8-20 WARNING – Gas Expires Soon! Press YES to Continue Gas will expire in two weeks. WARNING - No Pat. ID Entry Continue <ENTER> Patient ID is selected as a required entry Press <ENTER> to continue and enter a Patient ID WARNING - No Op ID Entry Continue <ENTER> Operator ID is selected as required entry Press <ENTER> to continue and enter an Operator ID WARNING - No ID Entry Continue <ENTER> Patient and operator ID is selected as required entry. Press <ENTER> to continue and enter a patient and operator ID. WARNING No Sample Detected Analyzer timed out before sample was detected by LG 1. Reposition sample device and rerun sample. 8 Troubleshooting Error Message WARNING - Check Pump Continue <ENTER> or <ESC> Possible Causes LG1 and LG2 detected improper movement of buffer within cassette. Remedial Action Open SMC and reseat cassette. Perform Flow Test in system diagnostics. Replace Pump Cartridge and seals. Check tubing for leaks. WARNING - Check Pump Please remove cassette If above message occurs repeatedly, this message is displayed. Replace cassette. No Display or power to analyzer LCD display does not indicate analyzer operation. Check if EPROM’s are seated properly. Check fuses F1 and F3. Check for proper battery/charger power. Replace main board. Check cable connections. Excessive use of gas Possible leak in gas path. Check for proper sealing of pump cartridge. Check for leak in tubing or tubing not connected. Clean seal in gas module assembly. Replace gas module. 8-21 8 Troubleshooting Error Message Possible Causes Remedial Action New SRC Lot! A new SRC is being run and has not been previously measured. Print and delete previous SRC data prior to measuring new SRC. Printer does not print Possible printer paper jam or malfunction. Check paper supply. Check printer for paper jams. Check printer cables for secure fit. Replace printer. Replace main board. SRC’s fail Possible problem with OPTI measurement system or SRC. If only one SRC is failing, run QC samples. If QC is OK, replace SRC. If all SRC’s are failing, reset unit and repeat measurements. Open the SMC cover. Verify three blue LED’s are visible and steady. If not, replace the SMC cover. Replace the SMC module. 8-22 8 Troubleshooting Error Message QC out of range Possible Causes Possible problem with OPTI measurement system. Remedial Action Repeat QC measurement with new cassette ensuring that the cassette is wiped dry prior to measurement. If PCO2 or PO2 values are out, check the cassette for air bubbles or voids. Verify that the QC vial is not too warm or too cool before performing measurement. Replace calibration gas bottle. 8-23 8 Troubleshooting Troubleshooting Flowcharts Opti Troubleshooting - 1 Instrument not READY or Lockup / Freeze Temperature Error Temp. High / Low No Display or black squares on top line BarCode Reader dead / Instrument locked-up / freeze "System Error Cycle Power" Message Wait 5 minutes for temp. to come into range Connect to Power Supply Cycle Power Frequent Occurrence? YES Check for Room Temp to be 15-32°C Check if EPROM's are seated correctly Initialize Instrument (Feed & Cover) Initialize Instrument (Feed & Cover) Initialize Instrument (Feed & Cover) Return Instrument (RA) Replace Bar Code Reader Return Instrument (RA) or Install Current Software version Return Instrument (RA) 8-24 Return Instrument (RA) NO Cycle Power (Initialize NOT necessary) 8 Troubleshooting OPTI Troubleshooting - 2 Cassette does not calibrate Obtain: - Cal. Report - Error Report 'Bad Calibration' Message Frequency high? 'Bad Cassette' Message NO Check Temp T1 and T2 YES YES New Cassette Lot? YES 'Bad Cassette' & 'Bad Calibration' Message 'Bad Sensor' Message Was Cass. Pouch opened prematurely? Clean Optics Re-run Sample Run DIAG1 Optics Test with LOW SRC NO New Cassette Lot? Obtain Calibration Report Change Gas Bottle Inspect for poke through in cass pouch (hole) Replace Cassettes Replace pump cart, check for leaks Replace bad Cassettes NO Replace Gas Bottle Check DIAG Valve Remove PumpCart Listen for gas flow Gas Flow ok? Values w/in Limit? pH, CO2 >30000 O2 >20000 Na/K > YES NO Return Instrument. Bad LED in Optic NO Return Instrument (Suspect bad Gas Module) Re-run Sample Install current Software Version YES Replace Cassettes 8-25 8 Troubleshooting OPTI Troubleshooting - 3 Errors during Calibration (Meas. & Control) 'Misseat 1' Message 'Misseat 2' Message 'Check Pump' Message Check Home Pos. visually (Pins in 12 & 6 o'clock pos.) Was cass. wiped before insertion? Run Flowtest DIAG-Flow (use new cass.) Check top plate to move freely. Remove any debris from SMC Clean Optics & Upper Plate (LG1, LG2 free) Replace Pump cart, check for leaks. Clean Nipple in SMC YES Flow test fails? NO Check DIAG LEDs LT1 & LT2 toggle Check DIAG LED L1, L2, L3 with new cass: 1200 - 2800 Rerun meas. using new cassette Rerun meas. using new cassette Rerun meas. using new Cassette Return Instrument (RA) 8-26 Rerun Flow test DIAG-Flow (use new cass.) YES NO Flow Test fails? ok 8 Troubleshooting OPTI Troubleshooting - 4 QC Values out of Range Check Time/ Date setting Check baro Obtain Info: - QC Lot # - Assay Ranges - QC Stats - SW Version Check for Corr Factors off ('-' on header) NO pH Value high pH Value low pCO2 Value high pCO2 Value low pO2 Value high pO2 Value low Check sampling technique (NO syringe!) Replace Gas Bottle Replace Gas Bottle Check Technique Check for voids and bubbles Check Sampl. Technique (NO Syringe!) Check Ampoule handling Frequent unstable pH? Install current SW Version Replace Cassettes Check Sampling Technique (NO Syringe!) Controls not at room temp? (refrigerated? ) Controls not at room temp? (hot?) YES Replace Cassettes Long cal. (blinking '*')? Check Sampl. Technique (NO Syringe!) Replace Cassettes YES New cass. fix problem? NO NO Replace Cassettes Return Instr. (Suspect Temp. T1, T2 problem) YES ok Establish custom assay ranges YES High Altitude? NO Check storage temp of Control (18 - 27°C) Replace Cassettes 8-27 8 Troubleshooting OPTI Troubleshooting - 5 Problems during Sample Analysis No Error Message Bubble seen in Cassette (void) 'Warning Bubble' Message NO Bubble > 1mm over Sensor? Results are ok Use Results 'Unbelievable' Results w/o Error Message 'Unstable Sensor' Message Check Time/ Date setting Check Baro Cassettes wiped before insertion? Check for Corr Fact. off ('-' on header) Do SRC's pass? YES Bubble in Reservoir? YES NO Bubble in Sensor? Rerun Meas. w/ new cassette NO YES Rerun Meas. w/ new cassette Rerun Meas. w/ new cassette Replace Cassette s (if frequent) Check Sampl. Technique Rerun sample Cassettes wiped before insertion? Do SRC's pass? New Cassette Lot? YES NO YES Blood Sample? NO Results are ok. Use Results Replace Cassettes (if frequent) NO pH > 7.5 pH < 7.1 or Neonate? YES pH unstable? YES NO Rerun sample Use pH if w/in 0.04 pH pCO2, pO2, Na, K unstable? Check for Bubble in resp. Sensor Rerun Meas. w/ new cassette NO Very high or low values? YES Sample Results are ok Use Results 8-28 8 Troubleshooting OPTI Troubleshooting - 6 SRC Test fails Run Optics Test with EMPTY SMC Return Instr. (Optics PCB humidity effect) NO Values w/in limit pH, CO2 <1200 O2 < 2000 YES Leave Instrument on wait 15 minutes Rerun SRC Test Instrument ok (if Warm-up acceptable) YES SRC passes? NO Return Instrument (Bad Optics PCB) 8-29 9 Electronic Diagrams 9 Electronic Diagrams OPTI 1 System The system electronic diagrams are included to provide service personnel a reference for tracing system signals when troubleshooting the OPTI 1 system. The electronic diagrams have been divided into sections to accommodate standard paper sizes. Each diagram provides signal labels which indicate input and output signals for that section of the circuitry. In addition to the signal labels, numbers are provided at each signal label indicating the appropriate diagram sheet number where that signal is connected. This is done to easily locate signal interconnection throughout each section of the electronic diagram. For the OPTI Main Board, there are 21 electronic diagram sections included within the service manual. Sheet one contains a system block diagram. This sheet is included to provide service personnel a reference for the overall system main signals and assembly interconnections. 9-1 Charge LED Barcode Reader RS232 Printer Heater Peristaltic Pump External Barcode Cassette Seat Detector Front Sample Detect LED's LG1 LG2 Cassette Seat Detector Rear Valve Drive Optics Module Serial Interface Controller Barcode Controller Battery Charger Control Printer Decoder Driver Peristaltic Pump and Valve Drive Controller A/D Preamplifier Photo Detector CPU Voltage Regulator D/A Converter Heating & Temperature Control RAM MUX LED Detection and Driver PhotoDetect Sample LG2 Heater Ram Battery Backup Barometer Real Time Clock Cassette Detector PhotoDetect Sample LG1 OPTI 1 - System Block Diagram 9-2 Sample Measurement Chamber Top Plate A/D Converter Clock EPROM Address Bus Display Decoder RAM Microprocessor MC68372 Buffer EPROM Address Decoder RAM Data Bus EEPROM READY - Cassette Barcode Menu <ENTER> Keyboard Controller Keyboard Decoder 1 2 3 4 5 6 7 8 9 0 ESC Status LED Charger Enter Battery Sheet 1 MAIN BOARD - Layout Sheet 2 9-3 CEO MAIN BOARD - RAM Battery Backup/Clock 9-4 Sheet 3 MAIN BOARD - Microprocessor Circuit Sheet 4 9-5 MAIN BOARD - 16 Bit EPROM Circuit 9-6 Sheet 5 MAIN BOARD - 16 Bit RAM Circuit Sheet 6 9-7 WR MAIN BOARD - Barcode and Serial Interface Circuit 9-8 Sheet 7 Sheet 8 QH COLC COLC QH MAIN BOARD - Keypad and Display Circuit 9-9 QH 9-10 QH QH MAIN BOARD - Cassette Detection Logic Circuit Sheet 9 Sheet 10 VO MAIN BOARD - Battery Charger Circuit 9-11 Sheet 11 VO MAIN BOARD - Voltage Regulator Circuit 9-12 MAIN BOARD - SMC Temperature Control Circuit Sheet 12 9-13 MAIN BOARD - SMC Heat Transistor Circuit 9-14 Sheet 13 MAIN BOARD - SMC Heater Control Circuit Sheet 14 9-15 MAIN BOARD - Digital to Analog Converter Circuit 9-16 Sheet 15 MAIN BOARD - A/D Converter Circuit Sheet 16 9-17 MAIN BOARD - Barometer Circuit 9-18 Sheet 17 Sheet 18 HOME QH MAIN BOARD - Motor Drive Control Circuit 9-19 M2 HOME M3 MAIN BOARD - Printer Controller Circuit 9-20 Sheet 19 MAIN BOARD - Cassette Detection Circuit Sheet 20 9-21 Sheet 21 M2 M3 MAIN BOARD - Interconnections 9-22 OPTIC MODULE - Microprocessor Circuit Sheet 1 9-23 OPTIC MODULE - Photo Detection Circuit 9-24 Sheet 2 OPTIC MODULE - Photo Detector Gain Circuit Sheet 3 9-25 OPTIC MODULE - A/D Converter Circuit 9-26 Sheet 4 OPTIC MODULE - Cassette Detection Circuit Sheet 5 9-27 HEATER BOARD - SMC Top Plate 9-28 DISPLAY BOARD - Interconnect C L O C 9-29 PRINTER BOARD - Interconnect Diagram 9-30 OPTI CCA System The system electronic diagrams are included to provide service personnel a reference for tracing system signals when troubleshooting the OPTI CCA system. The electronic diagrams have been divided into sections to accommodate standard paper sizes. Each diagram provides signal labels which indicate input and output signals for that section of the circuitry. In addition to the signal labels, numbers are provided at each signal label indicating the appropriate diagram sheet number where that signal is connected. This is done to easily locate signal interconnection throughout each section of the electronic diagram. For the OPTI Main Board, there are 21 electronic diagram sections included within the service manual. Sheet one contains a system block diagram. This sheet is included to provide service personnel a reference for the overall system main signals and assembly interconnections. 9-31 Charge LED Barcode Reader RS232 Printer Heater Peristaltic Pump External Barcode Sample Detect LED's LG1 LG2 LG3 Cassette Seat Detector Front tHb / SO2 Module Valve Drive Optics Module Serial Interface Controller Barcode Controller Battery Charger Control Printer Decoder Driver Voltage Regulator Heating & Temperature Control Peristaltic Pump and Valve Drive Controller A/D Preamplifier Photo Detector CPU D/A Converter EEPROM LED Detection and Driver Photo Detect LG3 Heater Ram Battery Backup Real Time Clock tHb/SO2 Controller Cassette Detector Photo Detect LG2 Photo Detect LG1 A/D Converter Clock Barometer EPROM Address Bus Microprocessor MUX OPTI CCA - System Block Diagram 9-32 Sample Measurement Chamber Top Plate MC68372 Buffer Display Decoder EPROM RAM Address Decoder IR Interface RAM Data Bus EEPROM Keyboard Controller READY - Cassette Barcode Menu <ENTER> Keyboard Decoder 1 2 3 4 5 6 7 8 9 0 ESC Status LED Charger Enter Battery Sheet 1 MAIN BOARD - Layout 9-33 9-34 PR5 PR4 INH VCC 8MHz PR3 1 X1 GND OUT VCC VCC 2 3 1 OPTIC-RXD C44 0.1UF 4 C43 0.1UF 1 26 50 48 44 46 49 45 47 43 52 51 18 19 2 3 17 8 7 A14/PB6 A15/PB7 AN6/PE6 AN7/PE7 MC68HC811E2FN GND VCC A12/PB4 A13/PB5 AN4/PE4 AN5/PE5 SS/PD5 SCK/PD4 MOSI/PD3 TXD/PD1 MISO/PD2 RXD/PD0 E R/W/STRB AS/STRA A10/PB2 A11/PB3 AD7/PC7 VRL AN2/PE2 AN3/PE3 AD6/PC6 VRH A9/PB1 AD4/PC4 AD5/PC5 XIRQ A8/PB0 AD3/PC3 IRQ AN1/PE1 AD1/PC1 AD2/PC2 AN0/PE0 AD0/PC0 MODB/VSTBY OC2/OC1/PA6 PAI/OC1/PA7 OC4/OC1/PA4 OC3/OC1/PA5 OC5/OC1/PA3 IC2/PA1 IC1/PA2 IC3/PA0 MODA/LIR RESET XTAL EXTAL U10 25 23 24 22 20 21 5 6 4 35 36 38 37 39 41 40 42 15 16 13 14 12 10 11 9 28 27 29 31 30 32 33 34 VCC VCC PR2 PR6 PR18 1 1.30K R31 PR20 PR19 2 PR15 PR14 PR7 9 PR16 PR13 PR10 PR9 PR21 PR17 PR12 PR11 PR8 MMPQ2222A U6D 7 MISO 1 RESET OPTICST VCC PR1 INTEGRATE_RELAY_LOW MOSI SCLK OPTIC-TXD CHAN_1-4_PD_SELECT CHAN_2-5_PD_SELECT CHAN_3-6_PD_SELECT CHAN_1-4_GAIN_SELECT CHAN_2-5_GAIN_SELECT CHAN_3-6_GAIN_SELECT DAC_LOAD_LOW DAC_CLEAR_LOW DAC_CS_LOW ADC_CS_LOW ADC_CLK OPTICS MODULE - Digital Circuit Sheet 1 8 10 2 2 11 NOTE: Make traces to 1, 2, and 3 cuttable. +5VA 4 1 2 1 0.1UF C4 LMC6084AIM U17A 1 C5 0.1UF 2 DPDT Relay 1N4148S D1 1 10 3 VCC SHIELDS CONNECTED TO AGND 2 3 2 -5VA PR32 4 2 5 6 2 Socket C51 0.1UF R1 5.11K 2 1 6 1 2 1 0.1UF C27 0.1UF C32 LIN COM +5VA -5VA 7 INTEGRATE_RELAY_LOW 7 8 1 2 10K R38 10M LMC6084AIM 7 U17B V- K1 R37 PR27 5 330pF PR31 V+ 7 9 1 4 3 ION 6 220pF 1 C46 1 GND ZS1 PR26 S2833 B6 S2833 4 3 2 C45 1 CHAN_3-6_GAIN_SELECT PR23 1 2 1 2 GAS 2 1 OFFSET V+ -5VA DG419 8 2 B3 PR25 4 0.1UF 2 NO NC U14 PR24 5 + - PR22 C7 LIN COM V- 1 6 1 1 +5VA 4 GND CHAN_3-6_PD_SELECT 0.1UF 2 C6 5 + - Channels 3 & 6 NO NC U1 DG419 8 2 PR29 2 PR28 OPEN R29 1 PR30 !_CHAN_3-6_OUT OPTICS MODULE - Amplifier Circuit Sheet 2 VL 3 2 VL 3 9-35 9-36 DG419 8 2 1 PR49 PR48 V1 2 1 2 S2833 B4 S2833 B1 4 3 ION 4 3 GAS VCC PR47 OFFSET PR37 PR46 R39 10M 1 C52 0.1UF 10 9 8 LMC6084AIM U17C 12 13 1 1 U17D 330pF 220pF LMC6084AIM C50 R40 10M C49 2 2 C53 0.1UF 14 SHIELDS CONNECTED TO AGND 330pF 1 C47 220pF R3 5.11K PR36 1 PR42 2 1 1 1 7 2 V- V+ 7 -5VA 2 1 C26 1 +5VA NO NC LIN COM 2 0.1UF 6 1 C31 GND 0.1UF C25 LIN COM 1 +5VA 4 6 1 0.1UF C30 R2 5.11K 0.1UF DG419 8 2 U3 -5VA VL CHAN_1-4_GAIN_SELECT V+ -5VA NO NC U16 10K R42 1N4148S D2 1 10 3 8 4 3 ION 4 2 C48 5 0.1UF 2 GND C11 4 LIN COM +5VA 5 1 1 3 6 1 C10 VL 0.1UF 2 2 DPDT Relay 6 5 4 2 7 9 S2833 B5 S2833 3 2 2 CHAN_1-4_PD_SELECT Channels 1 & 4 V+ 2 PR50 1 K2 PR38 1 2 1 B2 PR40 1 10K PR33 DG419 8 2 2 GAS 1 7 Socket ZS2 2 R41 7 -5VA NO NC U15 PR39 1 0.1UF 2 LIN COM +5VA VV+ INTEGRATE_RELAY_LOW OFFSET 1 C9 6 1 1 4 V- CHAN_2-5_GAIN_SELECT CHAN_2-5_PD_SELECT 0.1UF C8 4 GND GND + 2 5 - 5 + - Channels 2 & 5 PR44 1 NO NC U2 OPEN R27 2 PR43 DG419 8 2 PR41 2 PR45 OPEN R28 PR34 PR35 1 !_CHAN_1-4_OUT !_CHAN_2-5_OUT OPTICS MODULE - Amplifier Circuit Sheet 3 3 VL 3 2 2 2 2 VL 3 2 2 2 10K R25 10K R30 10K 1 PR53 1 PR52 1 U4A 9 5 6 2 3 2 2 10 1 -5VA 2 0.1UF C28 1 1 8 C24 1 7 0.1UF 1 C29 2 LMC6084AIM 1 0.1UF LMC6084AIM U4B 10K R24 1 +5VA 2 2 0.1UF C33 10K R5 1 LMC6084AIM U4C 4 !_CHAN_1-4_OUT !_CHAN_2-5_OUT !_CHAN_3-6_OUT R23 10K R22 11 + - + 2 - PR51 2 0.1UF C23 + - 1 PR54 PR55 PR56 CHAN_1-4_OUT CHAN_2-5_OUT CHAN_3-6_OUT OPTICS MODULE - Inverter Circuit Sheet 4 9-37 9-38 CHAN_1-4_OUT SCLK ADC_CLK PR59 R33 10K 10K 2 C38 0.1UF +5VA +5VA 1 1 2 + C39 22uF_6.3V PR60 U11 LM4040 PR58 PR57 5 3 4 2 1 13 9 8 7 6 22 21 20 18 10 11 2 MOSI R35 1 1 PD A/DIN2 VA+ U13 ADC12034 DOR CONV A/DIN1 MXOUT2 AGND DGND DOUT 2 1 2 2 10K R26 15 17 23 16 12 14 19 C41 22uF_6.3V EOC VCC MXOUT1 SCLK CCLK DIN CS REF- REF+ COM CH3 CH1 CH2 CH0 + 2 ADC_CS_LOW 1 1 CHAN_3-6_OUT C40 0.1UF 24 CHAN_2-5_OUT R34 1.30K 1 2 PR61 12 13 1 2 LMC6084AIM U4D 15 R4 + - +5VA 14 OFFSET MISO OPTICS MODULE - A-D Converter Circuit Sheet 5 1 2 PR65 11 1 2 LM324 -5VA 11 5 7 4 6 - + LM324 1 LM324 1 2 2 2 1 2 0.1UF C37 0.1UF C35 14 1 1 1 PR70 PR71 2 1 1.30K R17 C18 0.1UF R19 1.30K C19 0.1UF PR74 11 PR72 13 PR76 PR77 U6B MMPQ2222A PR78 PR79 PR80 MMPQ2222A U6C 12 1 121 R16 121 R18 121 1 10K 1 2 3 +12 - + 2 2 U8B U8A 13 12 U8D 1 U12 LM4040 2 R7 GND 1 14 2 5 121 PR64 PR68 PR67 1 2 R8 VSS -5VA 18 17 4 511 R6 VOUTF VOUTG VOUTH PR73 14 PR63 DAC8800 LD CLR SDI CLK CLK R20 R21 1.30K U6A MMPQ2222A 3 LEDGND PR62 VCC 9 VDD 13 12 10 6 15 16 5 2 DAC_LOAD_LOW DAC_CLEAR_LOW MOSI SCLK VOUTE VOUTC VOUTD VOUTA VOUTB 1 DAC_CS_LOW 8 7 VREFH1 VREFH2 C20 0.1UF 15 2 + VREFL1 VREFL2 LM324 U8C 8 2 2 - + 16 19 3 4 9 10 1 1 20 2 0.1UF C22 1 U5 1 PR69 PR81 +12 1 - VCC PR66 PR75 LED1_A LED2_A LED3_A C21 0.1UF 4 6 8 10 12 14 3 5 7 9 11 13 HEADER 7X2 2 1 J2 LED1_K LED2_K LED3_K LED_6_REF LED_5_REF LED_4_REF RT1_2 RT1_1 LED4_A LED5_A LED6_A LED6_K LED5_K LED4_K OPTICS MODULE - LED’s Gasses & DAC Circuit Sheet 6 2 6 2 2 9-39 9-40 LM324 NOTE: Make these traces cuttable. - + 14 U9D PR82 1 1 121 R10 2 13 12 2 1 R11 1.30K C15 0.1UF 11 PR91 6 LEDGND LM324 2 8 PR90 MMPQ2222A U7B MMPQ2222A U7C 5 - + 1 PR88 R13 1.30K C16 0.1UF 13 14 9 1 2 121 R12 1 10 U9C 2 PR87 4 LED_4_REF LM324 1 2 - 121 R14 1 U9B 2 R15 1.30K PR89 C17 0.1UF MMPQ2222A U7A 3 7 1 15 16 PR85 11 0.1UF C34 2 + 2 LM324 1 PR86 2 6 4 -5VA - + 2 0.1UF C36 1 5 2 3 1 1 PR84 U9A +12 1 LED_5_REF LED_6_REF PR83 +12 C3 0.1UF PR97 PR96 PR95 PR94 PR93 PR92 LED4_K LED4_A LED5_K LED5_A LED6_K LED6_A OPTICS MODULE - LED’s & Lytes Circuit Sheet 7 2 12 2 4 OPTIC-TXD VCC 1 2 PR105 TUBING PR100 TUBING FRONT Q1 SFH310-2 ZWS1 Q3 SFH310-2 +5VA ZWS3 THIRD +5VA R9 10K 1 PR99 9 PR102 10 12 9 11 7 PR103 TUBING ISO1 GP2S22 +12 +5VA PR104 PR98 TUBING C12 0.1UF REAR Q2 SFH310-2 ZWS2 +5VA VCC PR106 1 1 PR107 1 C42 22uF_6.3V + PR108 PR109 C1 22uF_6.3V 1 C14 0.1UF RT1_2 RT1_1 LEDGND OPTIC-RXD RESET OPTICST + -5VA 2 + C2 0.1UF +5VA ZPWB1 ZWB1 1 MMPQ2222A U7D ZWS4 2 100 1 1 R36 VCC 4 2 HEADER 6X2 2 4 6 8 1 3 5 7 J3 CONNECTOR ZJ1 3 2 PR101 18 20 22 24 17 19 21 23 26 10 12 14 16 9 11 13 15 25 2 4 6 8 1 3 5 7 J1 HEADER_13X2 CABLE ZC1 1 1.30K R32 -5VA VCC OPTICST PR110 C13 22uF_6.3V 4 OPTICS MODULE - Connector Circuit Sheet 8 2 2 2 2 2 8 10 9-41 9-42 11 VCC PR CL 74HC08 U58D VCC 13 13 10 CLK D Q Q U42B 11 74HC74 8 9 13 11 74HC14 U9F 74HC14 U9E 12 10 5 74HC14 U61C 6 VCC R102 1.0K R101 1.0K R100 1.0K PR26 VCC 2 12 12 VCC VCC VCC 1 THE INPUT TRACES FOR THESE PARTS SHOULD HAVE A SMALL PAD AND BE CUTTABLE BETWEEN THE PAD AND POWER 0.1uF 2 8 4 7 CS7 1 PR228 C77 1 VCC GND BC1 5 2 1 PCS3 PR227 VOUT BQ2201 I+ BC2 3 2 1 PCS0 PCS1 PCS2 PR225 VCC 1 THS CEO PR240 Y1 1 3 6 U31 32.768kHz PR27 2 PR226 PR239 VL3032 BT1 4 8 X2 0.1uF 1 C84 R99 1.0K VCC C56 6 PCS3 2 3 VCC 5 4 1 2 6 3 2 1 PCS0 PCS1 PCS2 VCC 1 5 4 5 2 6 7 PCS3 PCS0 PCS1 PCS2 0.1uF 1 DS1202 NC CLK RST GND X1 DATA VCC U5 Y2 2 Y1 Y6 Y7 G2A G2B Y1 Y2 B C Y6 Y7 G2A G2B 0.1uF 1 C71 74HC138 Y5 G1 Y4 Y3 Y0 A U27 74HC138 Y5 G1 Y3 Y4 Y0 C 2 B 6 A U36 PR52 7 9 10 11 12 13 15 14 9 7 10 11 12 13 14 15 5 3 1 VCC GND SPICS9 SPICS10 SPICS11 SPICS12 SPICS13 PR200 SPICS0 SPICS1 SPICS2 SPICS3 SPICS4 SPICS5 SPICS6 SPICS7 U49A 74HC125 74HC14 U20C 1 CS BCHRG PR208 PR28 74HC14 U20A MISO MOSI SCK 2 VDD PR147 11 3 PR258 PR254 PR146 PR253 PR252 SPICS10 74HC14 U20E 74HC14 U20B PR256 10 4 PR292 PR255 PR30 PR29 PR148 PR165 PR164 PR257 PR362 PR201 PR202 PR166 SEEPROMCS AUXCS1 AUXCS2 PRSEL+ PWROUTSEL+ DIGGENOUT KEYPADCS LCDEGENINSEL ANAGENOUTA/DSELECTSAMPSEN2SAMPSEN1- RAMVDD RAM-CS MISO MOSI SCK MAIN BOARD - RAM Battery Backup / Clock Sheet 1 2 2 2 PEB7 PR345 9-43 C106 2 1 19 13 15 17 PE6 PE7 PR317 R168 5.1K PR347 R169 5.1K PR295 PR237 PR346 1 1 2 0.1uF 1 74HC244 1G 2G 2A2 2A3 2A4 2Y2 2Y3 2Y4 2 VCC 7 5 3 D15 PEB6 91 LCDRS OPTICST D11 D12 D13 D14 2 97 94 93 92 R48 5.1K 1 R45 5.1K D3 D15 D6 D7 D8 D9 D10 1A2 1A3 1A4 2A1 D11 D12 D13 D14 1 103 102 100 99 98 1A1 VDD VDD VDD VDD D0 D1 D2 D3 D4 D5 1Y2 1Y3 1Y4 2Y1 D7 D8 D9 D10 PE0 111 110 109 108 105 104 1Y1 PE0 PE1 PE2 PE3 PE4 VDD VDD VDD VDD R170 5.1K PR343 1 1 2 2 HEADER_2 2 1 JP4 5.1K R155 5.1K R167 RESET HALT BERR TP12 TP13 TP14 TP15 T2CLK DEBUG MODE VCC VCC 1 MOSI MISO PCS3 PCS2 PCS1 PCS0/SS SCK PR152 RXD TXD PCS3 PCS2 PCS1 PCS0 SCKMOSI MISO 5.1K R171 1 2 1G VCC IRQ5 R116 2 PR230 PR206 49 48 47 46 45 44 43 MC68332 1 1G R115 VCC PR151 2 IRQ[1..6] VCC R154 5.1K VCC R114 2 PR171 PR169 VCC 11 PR204 PR205 R153 5.1K VCC 74HC14 10 PR130 8 5.1K 2 PR172 2 U51E R139 1 PR170 PR168 PR167 R166 74HC08 U58B PR369 5.1K 1 5 4 MMBTA06LT1 Q8 VCC PR342 R165 1.0K PR203 5.1K 1 5.1K 1 22pf 2 C104 R138 330K PR150 PR207 R112 2 2 VCC PR530 332RESET TP12 TP13 56 55 54 53 52 58 57 70 68 69 132 131 130 129 128 TP8 TP9 TP10 TP3 13 12 11 10 9 6 5 4 3 TP0 TP1 TP2 16 15 14 IRQ4 FREEZE/QOUT TSTIME/TS BKPT/DSCLK IFETCH/DSI IPIPE/DSO RXD TXD PR235 PE7 PE6 PE3 PE2 PE1 PEB1 PEB2 PEB3 PEB4 2 4 6 8 11 D3 D4 D5 D6 80 81 86 87 88 TXDSEL BC_DDY BC_CDY 18 16 14 12 9 VDD VDD 89 D8 D0 D1 D2 IRQ5 IRQ6 IRQ7 RESET RXDSEL PEB0 U43 MODCK 1 5.1K DSACK0 1 VSS VSS VSS 1 PR344 SIZ0 RMC AVEC DSACK1 73 72 71 D9 SIZ1 IRQ1 IRQ2 IRQ3 IRQ4 PR316 VSS VSS VSS VSS R46 IRQ6 IRQ7 PR238 VSS VSS VSS VSS 77 76 75 74 D[0..7] 2 IRQ2 IRQ3 IRQ4 IRQ5 5.1K R140 IRQ1 D[0..7] 1 VSS VSS VSS VSS D[8..15] VCC VSTBY R/W DS CLKOUT AS A1 A0 VDDSYN D[8..15] PR236 20 90 A1 A9 A8 A7 A6 A5 A4 A3 A2 XFC 66 82 79 85 25 24 23 22 21 A6 A5 A4 A3 A2 A13 A12 A11 A10 A18 A17 A16 A15 A14 TP9 TP10 TP11 TP4 TP5 TP6 TP7 TP8 TP0 TP1 TP2 TP3 PR232 10M R137 1 2 16MHz BC_DWR PR233 31 30 27 26 VDD VDD A10 A9 A8 A7 7 1 42 41 38 37 36 35 33 32 50 39 28 18 A18 A17 A16 A15 A14 A13 A12 A11 116 107 96 84 63 CSBOOT 19 78 114 113 112 126 CSBOOT 17 8 2 A[1..19] PR176 59 51 40 34 29 CSBOOT 101 95 83 67 FC1/CS4 FC0/CS3 BGACK/CS2 BG/CS1 BR/CS0 61 CS3 127 117 106 BCWR 2 XTAL EXTAL A21/CS8 A20/CS7 A19/CS6 FC2/CS5 64 120 119 118 115 60 62 A19 CS5 1 BCRD 1 PR294 2 2 A23/CS10 A22/CS9 1 32.768kHz R113 2 VCC 74HC14 U51D 9 5.1K 1 IRQ4 IRQ5 IRQ6 IRQ1 IRQ2 IRQ3 VCC C122 0.1uF PR427 PR229 PR132 SOT-23 1N4148S D36 6 VCC PR371 R183 1.0M 32KHZ BC_CRD BC_RTS HOMELG PRHOME SMCLG KEYPRESS MOSI MISO SCK PCS0 RXD TXD PCS3 PCS2 PCS1 PRINT-CLK PRHEAD-EN STATLEDBLNK SCANCLK BEEP PUMP-CLOCK TP2 TP3 TP0 TP1 C123 0.1uF 2 124 123 122 121 U24 PR234 2 PR293 1 CS9 CS8 CS7 125 0.1uF C86 22pf Y2 2 RAM-OE RAM-CS CS10 2 C85 PR231 1 C105 1 RAM-WE2 RAM-WE1 PR153 0.1uF 1 0.1uF 2 R136 10 1 PR173 PR154 0.1uF 2 0.1uF C66 1 PIN 61 C96 CLOSE TO VCC 2 1 PR174 1 2 0.1uF C97 2 0.1uF C65 2 C49 1 1 1 1 1 2 PR175 VCC RESET MAIN BOARD - Microprocessor Sheet 2 2 2 2 2 2 1 9-44 A2 A1 MOSI SCK SEEPROMCS PR91 PR95 VCC PR111 D2 PR115 D3 A11 PR99 D7 D8 VCC A14 A15 A16 A17 A18 A19 A1 A2 A3 A4 A5 A6 A7 A8 A9 A10 A11 A12 A13 D9 A16 PR74 1 28 29 3 2 30 31 12 11 10 9 8 7 6 5 27 26 23 25 4 512Kx8_ROM VPP A2 A3 A4 A5 A6 A7 A8 A9 A10 A11 A12 A13 A14 A15 A16 A17 A18 A0 A1 U66 HIGH BYTE A15 PR67 OE CE D3 D4 D5 D6 D7 D0 D1 D2 A17 PR69 D5 PR114 PR113 PR97 PR100 PR86 PR107 PR106 D10 PR110 VCC D11 ZSCT@U66 D7 D0 D1 D2 D3 D4 D5 D6 A14 PR73 SOCKET_32Y D6 24 22 21 13 14 15 17 18 19 20 A13 PR70 SOCKET_32DG OE CE D0 D1 D2 D3 D4 D5 D6 D7 A12 PR98 ZSCT@U67 512Kx8_ROM VPP A11 A12 A13 A14 A15 A16 A17 A18 A3 A4 A5 A6 A7 A8 A9 A10 A0 A1 A2 U67 D4 1 12 11 10 9 8 7 6 5 27 26 23 25 4 28 29 3 2 30 31 LOW BYTE A10 PR96 24 22 13 14 15 17 18 19 20 21 PR108 D12 0.1uF VCC 0.1uF 2 0.1uF 2 D8 D9 D10 D11 D12 D13 D14 D15 A18 PR72 C6 PR112 D1 VCC A1 A2 A3 A4 A5 A6 A7 A8 A9 A10 A11 A12 A13 A14 A15 A16 A17 A18 A19 A9 PR75 1 PR92 D0 A8 PR68 C5 5 6 7 8 A7 A6 VCC PR71 PR89 1 2 VSS ORG NC VCC SOCKET_8 ZSCT@U6 93C56 DOUT DIN CLK CS U6 A5 PR93 C4 4 3 2 1 A4 PR90 1 A3 PR94 PR109 D13 A19 PR66 PR87 D14 PR88 D15 MISO CSBOOT D[8..15] A[1..19] D[0..7] MAIN BOARD - 16 Bit EPROM Sheet 3 A1 A2 A3 A4 A5 A6 A7 A8 A9 A10 A11 A12 A13 A14 A15 A16 A17 A18 A19 24 29 22 1 31 2 30 4 28 3 26 23 25 7 6 5 27 GND D4 D5 D6 D7 D1 D2 D3 D0 16 19 20 21 15 17 18 13 14 D0 D1 D2 D3 D4 D5 D6 D7 A1 A2 A3 A4 A5 A6 A7 A8 A9 A10 A11 A12 A13 A14 A15 A16 A17 A18 A19 24 29 22 1 31 2 30 4 28 3 26 23 25 7 6 5 27 10 9 8 12 11 VCC 1 GND D4 D5 D6 D7 D1 D2 D3 D0 2 C27 0.1uF U12 RAM_512KX8 OE WE CS1 A17 A18 A14 A15 A16 A11 A12 A13 A8 A9 A10 A4 A5 A6 A7 A1 A2 A3 A0 VCC U13 RAM_512KX8 OE WE CS1 A17 A18 A14 A15 A16 A11 A12 A13 A8 A9 A10 A4 A5 A6 A7 A1 A2 A3 A0 32 10 9 8 12 11 2 C28 0.1uF 1 32 CS RAMVDD 16 19 20 21 15 17 18 13 14 D8 D9 D10 D11 D12 D13 D14 D15 RAM-OE RAM-WE1 RAM-WE2 D[8..15] D[0..7] A[1..19] MAIN BOARD - 16 Bit RAM Sheet 4 9-45 9-46 2 PP0 Q5 Q4 Q3 Q2 Q1 D5 D4 D3 D2 D1 74HC574 Q7 Q6 D7 D6 D8 D10 D9 D12 D11 PP[0..7] 19 17 18 15 16 D15 D14 D13 5.1K R184 2 1 PR393 5.1K R156 74HC574 D1 D3 D2 D5 D4 D7 D6 D8 OC CLK U28 2 Q1 Q3 Q2 Q5 Q4 Q7 Q6 Q8 Q9 MMBTA06LT1 PR319 19 17 18 15 16 12 13 14 BUZZER BZ1 PP0 PP2 PP1 PP4 PP3 PP7 PP6 PP5 16 25 37 35 36 33 34 30 31 32 12 13 7 5 11 29 38 UE1 FRQ CSS GRB BPR WR RD VS HBCR-1612 GND IOM PP0 PP2 PP1 PP4 PP3 PP6 PP5 PP7 PR182 PR181 PP5 PP1 PP4 PR178 PP0 15 14 10 8 43 44 41 42 40 4 1 2 3 9 18 19 20 21 22 23 24 PR177 XTAL 1 XTAL 2 LED SDI DDY RTS CDY DWR CRD RST U5S U2S CST UEE EA PDR PGB C3C I2C EX3 VCC VCC VCC 2 C32 0.1uF 1 PP6 PR183 PP2 PR179 13 12 PR297 PP7 PR184 PP3 PR180 74HC32 U10D 2 C58 1.0K 0.1uF 1 1 R254 74HC32 PR82 VCC 11 PR296 6 U10B PR532 2 5 4 1N4148S D8 R42 20.0K PR62 PR81 C37 22uF/16V Q19 MMBTA06LT1 PR488 + 2 1 VS 2 4 3 6 5 9 8 7 1 11 VCC 2 1 BEEP 4 3 PP2 PP1 Q8 D8 12 13 14 PR262 PR318 0.1uF 1 VCC 3 74HC32 U10A 1 1 5.1K R255 PR60 PR61 2 1 5.1K 2 2 1 1 2 2 3 4 1 2 J14 RJ45 8 6 7 4 5 2 3 1 J1 SHEADER_4X1 1.0M R41 1.0M R30 SHIELD OUTPUT +5V GND C31 0.1UF PR78 C8 0.1UF PR10 VCC DIG V+ SW 2 16MHz 6 5 PP4 PP3 1 U37 VCC VCC R8 VCC SHIELD GND BC_RTS BC_CRD BC_DWR 2 D[8..15] 9 8 7 OC CLK U21 74HC14 U51A PR531 C87 VCC 2 PP7 PP6 PP5 1 11 2 74HC14 4 PR259 1 BCRD PR368 3 U51B PR263 1 BCWR RESET BC_DDY BC_CDY PR261 PR260 MAIN BOARD - Barcode Interface Sheet 5 1 VCC 4 13 12 11 10 0.1uF 74HC595 G RCLK SRCLR SRCLK SER U63 C153 1 2 QH' QG QH QE QF QD QC QB QA 9 6 7 5 4 2 3 15 1 9 PR470 74HC14 U20D 6 5 Q Y7 PR485 PR468 C57 1nF PR149 PR486 PR505 2 14 VCC 1 STATLEDBLNK PR502 2 13 MOSI 9 LCDRS R70 1.0K 12 LCDRS 1 11 LCDE- 5 Q Y6 G2B 74HC138 Y5 G2A Y4 G1 PR487 ROWA ROWB ROWC ROWD 1 LCDD0 LCDD1 LCDD2 LCDD3 LCDD4 LCDD5 LCDD6 LCDD7 5.1K 13 2 12 PR314 PR365 PR504 VCC LCDBL ROWA ROWB ROWC ROWD PR509 74HC14 U20F PR364 LCDE+ R253 74HC74 3 2 U42A PR341 D 7 9 10 11 13 12 15 14 CLK PR469 8 VCC 1 KEYPADCS 6 U49B 74HC125 5 4 6 Y2 C Y3 Y0 Y1 A B CL MISO 8 PR 74HC08 U62D 74HC74 8 3 1 2 KEYROWD 9-47 0.1uF 2 16 18 20 22 24 15 17 19 21 23 VCC 0.1uF 1 VCC COLA COLB COLC COLD R243 5.1K 2 VCC PR501 C127 2 1 12 14 11 13 2 C119 Q18 MMBTA06LT1 SOT-23 8 10 7 9 26 6 25 4 5 2 3 1 J13 PR500 PR145 SHEADER_13X2 PR484 PR366 PR409 2 RESET PR315 Q U52B 9 0.1uF 2 2 SCK 1 8 PR506 COLA COLB COLC COLD PR498 9 74HC597 DIP RCLK SRCLR SRCLK SRLOAD H G F E D C A B SER U53 74HC08 74HC08 U62B U62A STATLEDG STATLEDR 74HC14 U61D 10 12 11 13 7 6 5 4 2 3 15 1 14 PR497 VCC 5 4 2 1 PR499 QH 6 3 9 C149 1 PR454 VCC PR455 PR456 0.1uF PR503 VCC 9 PR391 0.1uF 1 2 74HC08 U62C C128 10 2 8 PR457 1 2 KEYPRESS CL VCC CLK Q U50 C120 1 4 74HC125 74HC74 6 D 13 10 Q 11 12 CL VCC CLK PR U49C 3 U52A 5 VCC VCC VCC R244 5.1K 2 SCANCLK 2 Q PR PR363 5.1K R182 4 1 D VCC 10 VCC 2 VCC PR390 1 PR389 1 KEYCOLB 1 PR467 1 5.1K R245 R246 5.1K C148 0.1uF R230 5.1K MAIN BOARD - Keypad and Display Sheet 6 1 9-48 Q6 Q7 Q4 Q5 Q2 Q3 Q1 6 12 13 11 10 14 4 3 6 5 9 11 12 5 PR392 9 7 5 6 4 2 3 15 1 10 U49D 74HC125 11 13 74HC14 U61E 11 PR478 74HC08 1 12 1 VCC 74HC14 U61A 2 VCC 2 2 4 1 2 C138 VCC CLK INH H G E F D C A B SER PR494 5.1K R251 PR483 R219 20.0K SH/LD 74HC165 QH QH U64 0.1uF 1 7 9 PR480 5.1K R252 C144 0.1UF PR477 PR482 1 74HC14 U51C QH' QH QG 74HC595 QF G QE QC QD QB QA RCLK SRCLK SRCLR SER U65 256 HZ 1024 HZ 512 HZ 4096 HZ 2048 HZ 8192 HZ 16384 HZ 3 1 HOMELG 74HC4024 RST CLK U58A 1 2 15 6 5 4 14 3 13 11 12 10 2 VCC C157 12 2 74HC14 13 PR491 PR474 U51F 2 1 VCC PR476 5.1K R256 VCC PR496 MMBTA56LT1 Q17 0.1uF 1 PR481 VCC 5.1K 1 0.1uF 2 5.1K VCC 12 PR452 2 PR508 R242 2 C156 1 R257 1 PR507 MMBTA56LT1 Q16 74HC14 U61F VCC PR475 VCC 4 13 1 0.1uF 1 C121 5.1K R241 74HC14 U61B VCC Q20 MMBTA56LT1 2 3 2 VCC 5.1K 2 VCC VCC PR465 1 150R R223 R218 49.9K PR492 1 PR426 PR495 PR493 PR490 R229 C143 0.1UF VCC 1 MMBTA56LT1 Q21 1 PR370 PR367 2 1 2 1 1 SMCLG GENINSEL MISO SCK DIGGENOUT RESET MOSI 32KHZ U57 PR408 2 PR479 PR453 PR451 PR463 150R R250 2 PR460 HOMELGT SMCLGT KEYROWD HRC HRA GASPRESS1 CASS_F_COL HOMELED CASS_F_AN STATLEDG STATLEDR CASSLGDRVR MAIN BOARD - Cassette Detection Logic Sheet 7 2 2 2 2 1 3 SWITCHCRAFT RAPC-712 POWER_JACK J2 SLEEVE_NEG PIN_POS PR5 PR1 1 C 1 2 RTC1 2.87K 8 7 5 6 3 4 MCV TCO SNS TM2 TS BAT VSS BQ2003 CHG TEMP TM1 DIS MOD DCMD DVEN 9 10 12 11 14 13 15 16 1 2 VCC 2 2 CCMD U2 PC_LED D1 1 1 1 1 2 C7 0.1UF PR9 R3 1.21K PR8 R2 30.1K 1 2 PR209 5.1K R7 1.0K R51 1 C19 0.1UF 2 R4 68.1K 1 MMBTA56LT1 Q3 2 TO-220 C3 0.1UF PR7 PR35 PR241 TIP42 Q2 R92 1.21K R93 392R PR118 Q1 MMBTA06LT1 SOT-23 SOT-23 1N4148S D14 R24 470_1/2W PR34 PR136 PR533 PR55 1 100K 2 1 PR6 2 2 PR3 TP2 R1 5.1K 1.0_1W R37 1 1.0K R5 TP 1 TP15 TP PR33 R22 100K C20 0.1UF 1 R21 1 PR2 2 2 RELAY_SPDT K1 2 1 TP 1.0_1W R36 10 1 PR32 2 RTC2 3.74K C2 0.1UF 1 TP1 C40 0.1UF 1 1 CUT FOR NIMH CUTTABLE TRACE 2 1 PR54 C1 0.1UF VO D2 1N4148S 1N5821 D7 TP14 TP R50 NUT_6_32 ZN@Q2 WASHER_STAR_6 ZW@Q2 SCREW_6_32X0.25 ZS@Q2 100 R20 2 VI U1 78L05 3AT F1 FUSECLIP ZCLIP2 PR117 1 1 FUSECLIP ZCLIP1 PR31 RB1 100K RB2 909K R23 10_PTC PR56 PR53 TP TP4 BATTERY_NEG 1 1 1 GND TP38 TP THERMISTOR PR101 PR378 SW_DPST S1 BATTERY_POS D4 1N5821 D5 1N4148S SOT-23 LHEADER_4 3 4 2 1 J3 BCHRG SW_POWER MAIN BOARD - Battery Charger Sheet 8 2 2 1 1 1 2 2 2 2 2 2 2 2 2 2 2 9-49 9-50 1 8 7 6 5 22uF/16V C73 C80 0.1uF PR244 C + VI U32 79L05 C 2 2 LT1054 V+ OSC VREF OUT 1 2 1 + VO VOUT TP25 TP C51 100uF/25V TP31 TP 2 FB/SD C+ GND C- C67 22uF/16V 1 + C59 0.1uF VIN U22 LM7805 PR189 1 22uF/16V 1 2 3 4 +12V SVA D30-125-08 TP16 TP TO-220 ZF@U22 PR186 Q6 TIP42 D30-125-08 ZF@Q6 2 U25 VOUT PR213 R85 1.0K 2 PR134 VIN 4 5 2 SINGLE POINT GROUNDING ON/OFF OUT FB PR187 C81 0.1uF PR267 C68 0.1uF PR212 -5V +5V +12V C108 0.1uF D30-125-08 ZF@U44 PR138 PR135 NOTE: HEATSINK LM2575 TO COPPER PLANE ON PC BOARD 1 5 PIN TO-263(S) U14 LM2576S-5 TP24 TP 1 C72 PR210 C41 0.1uF VIN 1.0K SWITCHRGND 1 R84 TIP105 TO-220 TP23 TP VIN U44 LM7806 D9 1N5821 100uH L1 PR102 VOUT VSUP + TP37 TP 1 1 2 3 K32 JP2 C107 0.1uF PR350 5.1K 2 D30-125-08 ZF@Q5 R86 C33 1000uF/16V (1-2 default) 1 PR211 SVA D30-125-08 PR188 1 U17 LM7812 TP26 TP 2 3 ZF@U17 2 1.0K R49 Q4 D30-125-08 ZF@Q4 1 +6V PR214 C52 0.1uF 5V, 3A 2 NOTE: HEATSINK LM7812, LM7805 AND LM7806 TO COPPER PLANE ON PC BOARD PR119 PRINTGND 1.0K R83 1 PR116 GND POWERGND 1.6AT F3 3AT 1 FUSECLIP PR534 ZCLIP6 FUSECLIP 1 PR336 PR137 PR185 R71 68.1K PR133 2 1 OVERTEMP SWITCH_VS 1 BATSENSE SW_POWER FUSECLIP ZCLIP5 F2 ZCLIP4 FUSECLIP + 1 + ZCLIP3 TO-220 Q5 IRFZ34 D6 ICTE.5 C50 2200uF/16V TP7 TP PR546 PR155 PRINTGND POWERGND PVCC SWITCH_5V S5V VCC VSUP VS SVA VHTR MAIN BOARD - Voltage Regulator Sheet 9 2 2 2 C 2 2 1 C 2 2 1 2 2 1 1 2 VREF 1 2 PR270 500POT R126 PR245 PR271 PR246 PR272 2 1 2 1 1 2 1 2 1 9-51 2 -5V -V6 6 2 1 R146 10 +5V 2 1 PR305 +V6 REF02 GND +V PR247 1 C101 0.1uF 2 PR327 4 2 U33 24.90K 0.1% R147 R106 9.09K CW TO RAISE TOPPLATE TEMPERATURE PR308 R107 200R R108 9.09K R109 9.09K R110 9.09K 1 2 13 12 TLC2274 SO-14 - + U41D 10 9 2 3 2 2 14 R122 5.1K R157 5.1K PR303 1 C100 0.1uF R145 10 1 -V6 8 U40C SO-14 PR269 1 1 SO-14 TLC2274 0.22uF 1 C93 +V6 1 U40A TLC2274 0.22uF 1 C109 0.1% 24.90K R125 24.90K 0.1% R123 1 TP34 TP 1 2 2 24.90K 0.1% R121 24.90K 0.1% R174 R124 500POT PR307 CCW TO RAISE OPTIC TEMPERATURE 2 2 PR354 PR266 PR329 2 PR302 12 13 5 6 2 0.22uF 1 C92 1.0M R119 1 1 0.22uF 1 C110 1.0M R159 R120 20.0K 2 2 R158 20.0K 1 +5V C82 0.1uF R111 10 PR268 24.90K 0.1% 2 PR328 2 PR273 SVA TOPPLTNTC2 PR304 1 R175 1 TOPPLTNTC1 OPTICNTC2 OPTICNTC1 4 2 + 11 - PR379 + - PR353 1 + - + - TP36 TP TLC2274 SO-14 14 U40D TP33 TP TLC2274 SO-14 7 U40B TP39 TP PR301 PR330 VREF PLATE" "TOP TP32 TP MODULE" "OPTIC TOPPLTCTRL VREF OPTICCTRL MAIN BOARD - SMC Temperature Control Sheet 10 2 2 2 9-52 1 2 1 2 1 2 1 2 -5V R197 10 -V8 C132 0.1uF 1 PR399 1 +V8 2 R208 10.0K 1 30.1K R209 2 C131 0.1uF 2 PR411 R196 10 1 PR413 1 D49 1N4148S SOT-23 2 PR443 PR412 6 5 - + 2 -V8 11 +5V 10.0K R236 2.4_1W 1 - + 4 R205 2 3 +V8 20.0K R210 1 TLC2274 SO-14 7 U55B TLC2274 SO-14 1 U55A "TOP PLATE" TLC2274 SO-14 8 U55C PR400 -3.32V PR414 PR438 1 D48 1N4148S SOT-23 2 1 PR398 - + TLC2274 SO-14 C140 0.22uF C125 0.22uF C124 0.22uF 1 1 1 392R R221 392R R198 392R R185 2 2 2 + + + 1 PR441 R222 10 1 PR459 R237 2.4_1W 2 10.0K R206 1 9 10 PR395 - PR394 2 VREF 1 2 6.2_1W R200 10.0K R190 1 1 13 U55D 14 2 TOPPLTEMIT2 2 2 10.0K R189 2.4_1W R191 10.0K R186 1 + 1 5.1K 2 PR439 1 R207 PR440 2 2 PR396 6.2_1W R192 R199 2 10.0K 12 2 D30 1N4148S SOT-23 2.4_1W 2 R187 1 TOPPLTEMIT1 1 1 D29 1N4148S SOT-23 PR374 1 PR352 R173 10 PR375 2 TOPPLTCTRL PR377 PR397 PR351 2 1 OBLKEMIT2 PR376 5.1K R188 1 OBLKEMIT1 OTOPEMIT2 OTOPEMIT1 OPTICCTRL 1 "OPTIC MODULE" C145 22uF/16V PR458 C151 22uF/16V PR415 C154 22uF/16V PR373 -LTGATEREF TOPPLTBASE OBLKBASE OTOPBASE MAIN BOARD - SMC Heater Transistor Sheet 11 2 2 2 2 2 1 2 1 2 1 2 1 9-53 2 -5V R127 10 -V9 1 C94 0.1uF +V9 2 PR309 C111 0.1uF 1 PR333 R177 21.5K 150K R130 2 PR357 PR312 PR355 PR358 5 6 10M R178 SO-14 TLC2274 1 34.0K 2 2 7 PR359 PR380 PR310 PR334 R163 20.0K U41B TOP TEMP IND CW TO LOWER R129 5KPOT 1 R151 OPTIC TEMP IND CW TO LOWER 1 2 R164 10 R176 10.0K 1 2 PR332 1 +5V VREF TOPPLTNTC1 150K R160 2 2 + 1 R161 5KPOT 34.0K R148 1 - OPTICNTC1 1 392R R180 R128 20.0K 10 9 3 2 -V9 2 1 +V9 4 90.9K 1 2 2 SOT-23 1N4148S D31 10M R179 TLC2274 SO-14 8 U41C 0.22uF 1 C103 R150 2 TLC2274 SO-14 1 U41A 90.9K R149 0.22uF PR381 2 1 1 C102 + 11 - 2 + - PR306 PR311 PR331 1 1 1.0K R152 1.0K R162 2 2 PR335 PR356 BCHRG TOPPLTTEMP OPTICTEMP MAIN BOARD - SMC Heater Control Sheet 12 2 9-54 -3.32V 2 -LTGATEREF 1 MOSI TP22 TP 2 0.1uF 1 2 R59 150R C63 R79 150R 2 SAMPSEN2- SENSE_F_K SCK SAMPSEN1- 1 RESET SENSE_R_K TP19 TP 0.1uF 1 C46 VCC VCC 11 10 8 4 7 9 5 13 15 14 11 10 8 4 7 9 5 13 15 14 AD7543 SOL-16 STB4 STB3 STB2 STB1 SRI LD2 LD1 CLR VREF VDD U23 AD7543 SOL-16 STB4 STB3 STB2 STB1 SRI LD2 LD1 CLR VREF VDD U15 DGND AGND IO2 IO1 RFB DGND AGND IO2 IO1 RFB 12 3 2 1 16 12 3 2 1 16 PR161 PR124 PR197 PR143 0.1uF 1 C76 6 5 10 9 13 12 2 3 - + - + - + SVA 4 + - 2 LM324 SO-14 7 U30B LM324 SO-14 8 U30C LM324 SO-14 14 U30D LM324 1 U30A PR221 PR222 PR223 PR196 1N4148S D12 1N4148S D16 1N4148S D17 1N4148S D13 PR220 SENSE_R_AN "REAR" SENSE_F_AN "FRONT" PR195 MAIN BOARD - Digital to Analog Converter Sheet 13 11 R33 5.1K R34 5.1K TP 1 1 1 1 1 2 C163 0.1UF R29 150K TP TP11 R43 5.1K TP TP12 1 2 C162 0.1UF R47 1.0M TP TP10 R44 10.0K 1 TP TP9 2 C161 2.2uF/10V PR538 PR537 REF+ 2 + TEST VA+2 CH8 CH9 CH10 CH11 CH12 CH13 CH14 CH15 10 11 12 13 15 22 24 25 26 17 16 1 2 3 4 5 6 7 8 9 U11 ADC12038 MXOUT1 A/DIN1 MXOUT2 A/DIN2 VA+ CS DIN SCLK CCLK REF+ REF- CH0 CH1 CH2 CH3 CH4 CH5 CH6 CH7 COM 28 TP TP6 2 PR84 10 EOC DOR CONV PD AGND DGND DOUT R35 2 20 27 21 19 18 14 23 2 TP13 2 PR85 1 PR548 PR64 1 + C26 22uF/16V DOUT PR105 VA+1 1 AUXCS2 PR547 2 + PR536 PR535 C38 0.1UF 10 11 12 13 15 22 24 25 26 17 16 + MISO C39 22uF/16V VA+2 U29 ADC12038 MXOUT1 A/DIN1 MXOUT2 A/DIN2 VA+ CS DIN SCLK CCLK REF+ REF- 1 THBSIG THBREF PR539 2 PR65 R28 10.0K +5V STANDOFF ZST1 PR248 1 2.2uF/10V C160 VCC SENSE_IF_EM 1 PR63 1 512KHz PR521 R97 10.0K t 2 1 74HC4024 1 RST 12 11 9 6 5 4 3 1 Q1 Q2 Q3 Q4 Q5 Q6 Q7 1 CLK R94 30.1K 1 U69 C165 0.1UF R95 150K TP27 TP RT1 10K4A1 EOC DOR CONV PD AGND DGND DOUT 20 27 21 19 18 14 23 C61 0.1UF CH0 CH1 CH2 CH3 CH4 CH5 CH6 CH7 CH8 CH9 CH10 CH11 CH12 CH13 CH14 CH15 BAROMETER TOP PLATE TEMPERATURE OPTIC BLOCK TEMPERATURE FRONT LIGHTGATE REAR LIGHTGATE BATTERY/SUPPLY VOLTAGE CASSETTE PRESENT DETECTOR INSIDE CASE TEMPERATURE ION SIDE LIGHTGATE SPARE SPARE SPARE TEST VOLTAGE (~ 2048) SPARE tHb SIGNAL tHb REFERENCE A/D CHANNEL ASSIGNMENT DOUT 2 2 1 C164 0.1UF R96 150K PR215 PR218 CH0 CH1 CH2 CH3 CH4 CH5 CH6 CH7 COM PR156 + 2 16MHz D15 1N4148S SOT-23 PR216 TP TP28 1 2 3 4 5 6 7 8 9 28 TP TP29 CH0 CH1 CH2 CH3 CH4 CH5 CH6 CH7 10 2 2 TP TP30 PR217 R72 C60 22uF/16V 1 A/DSELECTMOSI SCK REF+ 1 + 1 BARO-OUT TOPPLTTEMP OPTICTEMP SENSE_F_EM SENSE_R_EM BATSENSE CASSLG VREF D11 1N4148S SOT-23 1 D10 1N4148S SOT-23 +5V MISO C62 22uF/16V VA+1 MAIN BOARD - Analog to Digital Converter Sheet 14 2 2 VCC 2 2 2 2 2 2 1 1 9-55 2 2 VREF 1 2 1 2 1 2 1 2 -5V 2 -V15 30.1K R52 2 PR139 1 R56 10 1 +V15 LMC6084 1 U18A PR120 R55 10.0K PR123 1 13 12 - + 10.0K -OUT 1 R58 TP21 TP 2 2 1 2 0.22uF 1 C43 21.5K R53 2 LMC6084 SO-14 14 3 2 1 VREF -EX -3.57V PR219 U18D PR141 0.1% 24.90K R87 0.22uF 1 C69 392R R75 2 PR140 3 2 1 SCX15A 5 BARO1 TP TP18 5 6 24.90K 0.1% R89 PR121 1 2 1 PR193 150R R77 1 R54 5.1K 1 2 JP3 24.90K 0.1% R76 R67 100POT HEADER_2 2 SPAN CCW TO INCREASE LMC6084 7 +OUT U18B 2 PR194 1 C44 0.1uF 1 +V15 1 PR158 R88 10.0K PR192 -V15 20.0K 2 PR159 2 C53 0.1uF R65 10 R66 34.0K 5KPOT PR540 R74 2 PR191 PR122 R57 10.0K PR142 10 9 2 1 0.22uF 1 C45 0.1% 24.90K R78 LMC6084 8 U18C 2 PR160 1 TP 1.0K R64 TP20 2 + 1 +5V PR157 PR190 4 + - 11 + - 9-56 + - R73 C42 22UF/16V BARO-OUT MAIN BOARD - Barometer Sheet 15 2 1 2 4 2 2 C74 1 2 PR313 1 2 L297 20DIP VS OSC Q7 MMBTA06LT1 SOT-23 PR290 C75 1nF 12 16 S5V 1 5.1K R135 C INH2 D HOME SENSE1 SENSE2 VREF GND 7 8 9 3 14 13 15 2 6 2 R134 5.1K PR291 S5V 1 0.1uF 2 CONTROL ENABLE SYNC RESET A INH1 B PR279 SVA C113 0.1uF PR280 C117 0.1uF SVA PR281 PR337 1 5.1K R98 100K 1 11 10 1 CW/CCW STEP HALF/FULL 4 5 PR277 C115 0.1uF 2 19 20 17 18 U34 PR276 S5V 1 PR249 PR251 9 VALVMTR-EN PUMP-DIR PUMP-ENA ILIMIT PRMTR-EN PRINT-DIR 2 R133 S5V PR287 QH' 15 1 2 3 4 5 6 7 1 1 PR289 74HC595 16SOIC RCLK G SRCLK SRCLR QA QB QC QD QE QF QG QH 2 0.1uF + 1 PUMP-CLOCK TP3 TP2 TP1 12 13 11 10 SER U35 1 C83 2 TP0 ANAGENOUT- RESET 14 VCC 1 POWERGND PWROUTSEL+ SCK MOSI VCC L293 VS VSS C114 0.1uF 8 16 10 9 15 7 2 1 C95 0.1uF L293 VS VSS IN3 ENB IN4 IN1 ENA IN2 U46 PR282 RAD TH TP35 TP GND GND GND GND OUT4 OUT3 OUT2 OUT1 GND GND GND GND OUT4 OUT3 OUT2 ENA IN2 IN3 ENB IN4 OUT1 IN1 U47 C118 100uF/25V 8 16 10 9 15 2 1 7 C147 0.1uF 4 5 12 13 14 11 6 3 4 5 12 13 14 11 6 3 3 6 7 8 4 1 9 5 2 VS PR383 PR386 VS MIC5841 SOL-18 VDD SCK DIN DOUT STB OE/RST VSS VEE VEE U60 17 16 15 14 13 12 11 10 18 D39 1N4148S SOT-23 PR385 D32 1N4148S SOT-23 D41 1N4148S SOT-23 PR388 D34 1N4148S SOT-23 K OUT1 OUT2 OUT3 OUT4 OUT5 OUT6 OUT7 OUT8 D40 1N4148S SOT-23 PR283 D33 1N4148S SOT-23 D42 1N4148S SOT-23 PR288 D35 1N4148S SOT-23 VSUP PR250 PR448 D20 1N4148S SOT-23 PR275 D24 1N4148S SOT-23 D22 1N4148S SOT-23 PR284 D26 1N4148S SOT-23 PR285 PR449 D19 1N4148S SOT-23 D23 1N4148S SOT-23 D25 1N4148S SOT-23 D21 1N4148S SOT-23 PR450 1 1 1 1 2.2 R131 2.2 R194 3.9 R132 3.9 R195 PR520 2 2 2 2 PR278 PR384 PR274 PR286 PR387 VLIMIT NOISE PUMP-D PUMP-C PUMP-B PUMP-A SYNC VALVE-D VALVE-C VALVE-B VALVE-A PRMTR-EN PRINT-DIR AUX GV_1 LCDBL SWITCH_VS SWITCH_5V FAN TEMPCTRL MAIN BOARD - Motor Drive Sheet 16 2 2 1 2 2 1 2 2 9-57 9-58 SCK MOSI 1 0.1uF C88 12 16 11 10 1 17 18 19 20 L297 20DIP VS OSC CONTROL ENABLE SYNC CW/CCW STEP HALF/FULL RESET U39 PR320 VCC C INH2 D HOME SENSE1 SENSE2 VREF GND A INH1 B 7 8 9 3 14 13 15 2 4 5 6 PR321 5 2 3 6 7 8 4 1 9 PR322 MIC5841 SOL-18 VDD SCK DIN DOUT STB OE/RST VSS VEE VEE U59 K OUT1 OUT2 OUT3 OUT4 OUT5 OUT6 OUT7 OUT8 C129 0.1uF PR323 10 18 17 16 15 14 13 12 11 +6V PVCC 1 PRINTGND S5V 0.1uF 1 C139 1 VLIMIT PRMTR-EN SYNC PRINT-DIR PRINT-CLK PRSEL+ PRHEAD-EN 2 C130 0.1uF PR324 8 16 10 9 15 2 1 7 L293 VS VSS IN3 ENB IN4 IN1 ENA IN2 U54 PR325 PR429 PR428 GND GND GND GND OUT4 OUT3 OUT2 OUT1 4 5 12 13 14 11 6 3 PR431 PR410 PR430 PR432 D46 1N4148S SOT-23 PR437 D44 1N4148S SOT-23 PR433 D47 1N4148S SOT-23 PR349 D45 1N4148S SOT-23 PR435 PR434 D28 1N4148S SOT-23 PR372 D38 1N4148S SOT-23 D37 1N4148S SOT-23 D27 1N4148S SOT-23 1 1 0 R172 0 R220 2 2 PR348 PR436 M4 M2 M3 M1 H1 H2 H3 H4 H5 H6 H7 H8 MAIN BOARD - Printer Control Sheet 17 2 2 2 3 1nF 8 2 +V69 LMC6082 1 U38A 1 49.9K R117 2 PR264 1 1 1 2 1 -V69 2 2 2 C99 4 R143 49.9K 1 2 2 -5V 2 -V69 5 1nF R104 1.0M 6 2 2 C79 R103 49.9K 1 C78 1nF PR242 U38B LMC6082 7 1 150R R105 2 PR243 1N4148S SOT-23 D18 1 C90 0.1uF 49.9K R142 R144 1.0M PR265 R118 316K 1 C89 0.1uF +V69 1 +5V 1 1 PR326 C98 1nF PR300 1 CASSLGCOL R141 392R PR299 1 + - VCC 1 + - PR298 C91 0.1uF CASSLG MAIN BOARD - Cassette Detect Sheet 18 2 2 2 2 2 9-59 9-60 12 14 11 13 S5V SHEADER_7X2 2 4 6 8 10 1 3 5 7 9 J5 IRDA AMP 103670-1 SHEADER_2X1 2 1 J8 13 4 16 1 11 9 6 8 RELAY_DPDT 16DIP K2 PR425 PR424 PUMP-D PUMP-A CASSLGCOL SENSE_R_EM OPTICNTC2 OBLKEMIT2 PR406 PR407 SHEADER_6X1 AMP 103670-5 1 4 3 2 6 5 J7 "PUMP MOTOR" PRHOME M2 M3 H1 H3 H5 H7 SW2 PVCC 1 PUMP-C 26 25 +12V OTOPEMIT2 OPTIC-RCV RESET A K C E R231 392R 2 D43 1N4148S SOT-23 18 20 22 24 17 19 21 23 +5V PR466 S5V 7 5 3 1 4 2 10 12 14 16 18 20 22 24 26 9 11 13 15 17 19 21 23 25 C150 0.1uF R235 5.1K 1 392R R234 SHEADER_13X2L 2 4 6 8 1 3 5 7 J12 PVCC 2 SW1 PR471 SLHEADER_7X2 13 11 9 8 6 "PRINTER" 2 "GAS VALVE 1" 90.9K R228 14 12 10 J10 "SMC" SHEADER_5X1 AMP 103670-4 4 5 1 2 3 1 A K C E 1 PUMP-B VCC 16 15 "FAN CONTROL" 14 13 SHEADER_13X2 8 10 12 4 6 2 7 9 11 3 5 1 SMCLGT VALVE-B VALVE-A AUX GV_1 J9 R233 5.1K R227 49.9K 1 NOISE RESET +12V -5V J11 VHTR SENSE_R_K SENSE_F_K TOPPLTEMIT2 TOPPLTNTC1 1 IRDACTRL RS232RXD DATA/CTRL RS232TXD FAN CASSLGDRVR SENSE_F_EM OPTICNTC1 SENSE_IF_EM OBLKEMIT1 OBLKBASE OTOPEMIT1 OTOPBASE OPTIC-TXD VCC "OPTIC MODULE" SHEADER_10X2 20 16 18 15 17 19 12 14 11 13 6 8 10 5 7 9 PR464 1 OPTICST CASS_F_COL SENSE_R_AN SENSE_F_AN 2 4 1 3 VS 1 CASS_F_AN TOPPLTEMIT1 TOPPLTBASE TOPPLTNTC2 J6 "SMC TOP PLATE" VHTR R232 5.1K PRINTGND KEYROWD KEYCOLB HRA HRC M4 M1 H8 H4 H6 H2 HOMELGT HOMELED VALVE-D VALVE-C GASPRESS1 MAIN BOARD - Interconnect Sheet 19 2 2 2 2 2 9-61 RXDSEL OPTIC-TXD VCC C29 2 PR44 PR46 PR83 PR76 0.1uF 1 12 U7D 74HC125 74HC14 U9A PR45 VCC PR43 3 74HC14 U9B OPTIC XMIT 1 11 0.1uF C11 2 1 2 9 8 15 9 14 7 3 4 C12 2 2 18 11 12 10 13 6 5 4 2 9 5 6 8 PR37 PR42 U7B 74HC125 74HC08 U58C OPTIC RECEIVE 10 0.1uF 1 PR541 PR77 U7C 74HC125 8 U4A 74HC4066 VCC 1 TR2IN TR2OUT LT1180 RC2OUT TR1IN ON/OFF RC1OUT RC2IN TR1OUT C2- C2+ C1- C1+ RC1IN V- V+ 13 IRDA/RS232 5 1 0.1uF 2 + C21 2 1 C22 0.1uF PR14 0.1uF 1 2 10 9 74HC32 U10C PR38 8 PR39 PR40 PR542 9 74HC14 U9D 8 5 10 U4D 74HC4066 (2-3 default) PR58 8 U4B 74HC4066 K32 3 2 1 JP1 74HC14 U9C PR11 PR17 9 6 U4C 74HC4066 PR47 VCC 3 VCC 0.1uF 1 C23 U7A 74HC125 PR543 2 4 PR15 2 1 TXDSEL TXD DE9S 9 4 8 2 U3 C9 1 11 5 3 PR13 0.1uF C10 PR36 VCC + 2 7 6 1 DCD DSR RXD RTS TXD CTS DTR RI GND PR41 BRACKET PR12 JACKPOST SCREW_M3_5MM PR4 ZJP2 ZS2 + P1 JACKPOST SCREW_M3_5MM ZB1 ZJP1 12 + ZS1 R25 5.1K 3 PR16 IRDACTRL OPTIC-RCV RXD RS232RXD RS232TXD MAIN BOARD - Serial Interface Sheet 20 2 1 6 4 10 13 4 TP44 -5V R203 10.0K 4 2 5 1 1 2 -5V SIG +5V +5V 2 14 13 SHEADER_7X2 12 8 10 11 7 9 780_A 780_M REF 2 -5V 1 850_A 850_M 670_A 12 PR444 PR422 1 3.01K R204 2 1 1 10 R271 2 R274 10 2 PR545 1 PR527 R268 10.0K 2 1G R276 2 2 1 5 6 1 12 13 9 10 1nF 2 C137 R275 10 PR528 PR525 2 1nF R213 5.11K 1 850 780 1nF 2 1 C142 670 LMC6084 8 U56C 2 LMC6084 7 R270 1G 1G R269 U56B 1 LMC6084 14 U56D 2 1 4 6 1 2 PR405 2 PR416 R217 20KPOT 1 PR423 3.92K R216 1 PR401 2 R272 10 2 R266 10.0K +2.5V R265 10.0K 1 PR421 +2.5V 9.09K PR526 1 R212 2 PR529 C112 0.1UF 10 20KPOT R215 PR544 1 1 2 VCC U45B 74HC4066 PR462 2 PR473 R273 10 R277 5KPOT PR550 C134 2 3 5 11 13 1 R240 5.1K R239 5.1K 1 1 PR472 R249 5.1K R248 5.1K 1 1 PR489 R226 5.1K R225 5.1K 1 1 1.0K R238 1nF 2 C152 1.0K R247 1nF 2 C155 1.0K R224 1nF 2 C146 2 2 2 SOT-23 MMBTA06LT1 Q14 SOT-23 MMBTA06LT1 Q15 SOT-23 MMBTA06LT1 Q13 + PR445 PR403 + PR418 + 1 1 J4 C136 0.1UF U45A 74HC4066 -2.5V 9 PR549 1 2 780_K 850_K +5V 1 670_K C135 0.1UF +5V 1 8 PR447 1 2 PR420 11 LMC6084 U56A 1 C116 0.1UF U45D 74HC4066 6 3 2 -5V VCC PR446 PR461 1 R181 5.1K 2 PR360 6 7 9 PR361 1 3 PR404 1 2 PR419 QH' QG QH 4 5 2 3 2 20.0K R202 74HC595 G RCLK QC QD SRCLK SRCLR QE QF QA QB SER PR338 2 U45C 74HC4066 1 13 12 11 10 PR340 2 VREF AUXCS1 SCK RESET PR339 1 1 + 15 1 2 2 - U48 1 + 14 1 + - 1 - 9-62 + - MOSI C133 2.2uF/10V PR402 C141 2.2uF/10V PR417 C126 2.2uF/10V PR382 850_M 850_K TP43 SOT-23 1 1 31R6 MMBTA06LT1 Q11 VCC TP46 SOT-23 1 R214 MMBTA06LT1 Q12 VCC 2 TP41 SOT-23 MMBTA06LT1 Q10 VCC 31R6 R201 31R6 R211 1 10R0 R193 2 2 2 TP42 850 780 780_A 670_A SIG REF 850_A 780_M 780_K TP45 670_K TP40 IRDA/RS232 DATA/CTRL MAIN BOARD - tHb / SO2 Laser / LED Driver Sheet 21 2 2 2 2 2 1 10 2 1 R69 2 2 PR163 +V33 -V33 C55 0.1UF C54 0.1UF 13 10 1 LMC6082 2 7 1 5 C47 1nF 1 R61 121K PR127 C48 0.1UF 2 U19B 9 VCC PR128 1 1 1 PR144 6 6 8 13 U16D 74HC4066 3 R90 57.6K PR224 R62 7.50K R91 8.66K PR129 1 -5V +5V R68 2 12 U16C 74HC4066 10 LMC6082 1 2 10.0K 1 1 PR125 U16B 74HC4066 +V33 8 4 11 4 R60 R82 5.1K 3 2 12 2 9 VCC 1 U16A 74HC4066 2 C64 1nF 1 U26D 74HC4066 1.0K U19A 2 2 1 8 1 -V33 1 C70 0.1UF 1 REF SIG R80 PR162 PR126 U26A 74HC4066 VCC PR199 PR198 2 + 780 10 3 1 - U26B 74HC4066 1 + 11 U26C 74HC4066 5 - 850 4 R63 24.3K R81 11.0K TP50 TP TP TP49 THBREF THBSIG MAIN BOARD - tHb / SO2 Amplifier Sheet 22 2 2 2 2 6 2 5 2 9-63 9-64 1 2 +VTCO 1 2 1 -VTCO C159 0.1uF PR514 2 2 -5V 2 1 C158 0.1uF PR515 1 PR519 R264 5.1K PR518 R261 5.1K PR513 13 12 9 10 6 5 2 3 - + - + - + -VTCO - + +VTCO 11 +5V TP47 TP 4 V(t)OPTIC 1 R260 1.30K R258 5.11K 2 2 2 5.1K R263 LM324 14 U68D 5.1K R262 LM324 8 U68C LM324 7 U68B LM324 1 U68A 1 1 PR517 PR516 PR511 PR510 1N4148S D51 1N4148S D50 VCC 1 OPTICNTC1 TOPPLTNTC1 V(t)TOP -5V TP TP48 R259 5.1K PR512 Q22 MMBTA06LT1 TEMPCTRL OVERTEMP MAIN BOARD - Overtemperature Protection Sheet 23 2 Gas Board Gas Board ZC2 BOT_BASE LED5_A LED5_C LED4_A LED4_C LED5_A LED5_C LED4_A LED4_C Ion Board LED6_C LED6_C CABLE LED6_A LED6_A BOT_EMIT_2 COLLECTOR BOT_EMIT_2 BOT_BASE TOP_EMIT_1I COLLECTOR TOP_EMIT_2I TOP_BASE TOP_EMIT_1I CABLE TOP_EMIT_2I TOP_BASE ZC1 Ion Board OPTICS MODULE - Interconnect Board Sheet 1 9-65 9-66 + + + + + + + + + + + + + + LED6_A LED5_A LED4_A + + + + + + + + + + + + + + J6 t 7X2 IDC/PCB 13 11 9 5 7 3 1 BOT_EMIT_2 COLLECTOR BOT_BASE TOP_EMIT_1 COLLECTOR TOP_BASE 14 12 10 6 8 4 2 THERMISTOR RT1 LED6_C LED5_C LED4_C LED1_C LED2_C LED3_C 1 ohm R2 1 ohm R1 TOP_EMIT_2G TOP_EMIT_1G 1 ohm R4 2 1 2 1 2 1 1 ohm R3 LED D3 LED D2 LED D1 Parts added later 3 3 Gas Board Q2 MJE800 Q1 MJE800 LED 3 LED 2 LED 1 1 LED1_A LED2_A LED3_A 14 10 12 8 6 4 2 Parts added later 2 ZJ2 13 9 11 7 5 3 1 7X2 IDC/PCB 1 7X2 IDC BOT_EMIT_1 COLLECTOR BOT_BASE TOP_EMIT_2 COLLECTOR TOP_BASE J2 ZJ1 6X2 IDC Bottom Heater 2 2 LED CABLE ZC4 ZC3 HEATER CABLE CON3 2 3 1 J1 LED6_A LED6_C LED5_A LED5_C LED4_A LED4_C TOP_BASE TOP_EMIT_2I TOP_EMIT_1I COLLECTOR BOT_BASE BOT_EMIT_1 LED6_A LED6_C LED5_A LED5_C LED4_A LED4_C TOP_BASE TOP_EMIT_2I TOP_EMIT_1I COLLECTOR BOT_BASE BOT_EMIT_1 OPTICS MODULE - LED’s Gasses and Heaters Sheet 2 LED6_A LED6_C LED5_A LED5_C LED4_A LED4_C 2 1 2 1 2 1 LED D4 LED D5 LED D6 Parts added later Q4 MJE800 Q3 MJE800 LED 4 LED 5 LED 6 1 LED6_A LED6_C LED5_A LED5_C LED4_A LED4_C CON3 Bottom Heater 1 2 1 2 3 J10 3 1 TOP_BASE TOP_EMIT_2I TOP_EMIT_1I COLLECTOR BOT_BASE BOT_EMIT_1 3 Parts added later 2 TOP_BASE TOP_EMIT_2I TOP_EMIT_1I COLLECTOR BOT_BASE BOT_EMIT_2 Ion Board OPTICS MODULE - LED’s ION and Heaters Sheet 3 9-67 PR10 1 D4 S1133-01 2 2 1 +5V 2 1 2 LMC6082 1 2 LMC6082 REF PR7 PR3 PR27 PR11 +5V -5V OPB806 -5V SIG +5V 670_K LASER_K 14 13 IDCPCB_7X2 8 10 12 6 5 9 11 4 3 7 2 1 REF 670_A 850_A 850_M 780_A 780_M SOCKET14 CABLE14 P1 ZJ2 ZC2 "FRONT" 18 20 17 19 PR15 PR6 PR18 LD1 RLD-78MC 1 PR22 R5 82.5K "REAR" 2 10K4A1 RT1 HLMA-KH00 D5 PR14 1 7 R4 10.0K PR32 D2 HLMA-KH00 16 15 2 U1B SIG R3 10M 2 14 13 IDCPCB_10X2 12 11 3 1 220pF C4 R1 10 0.1UF C2 1 U1A 0.1UF 1 3 1 PR13 8 10 9 4 6 3 5 7 2 1 P2 Q2 MJE800 t 1 1 2 ZPWB1 Z1 8 C5 0.1UF PR19 PR17 2 LD2 RLD-85PC "POST ION" HLMA-KH00 D3 PR26 1 PR24 R6 82.5K PR16 C6 0.1UF ITEMS INSIDE OF DASHED LINES ARE NOT PART OF THE BB7010 ASSEMBLY, BUT ARE ADDED AT A LATER STEP AS PART OF THE BP7058 ASSEMBLY NOTE: 2 1 5 6 4 PR1 PR5 2 3 1 3 1 PR9 D1 S1087-01 2 2 C1 4 ISO1 "FRONT SEAT" PR21 PR20 SOCKET20 ZJ1 1 + PR4 R2 10 PR2 220pF 1 2 PR30 1 - -5V C3 PR31 2 + 2 PR33 PR28 PR23 1 - 9-68 Q1 MJE800 CABLE20 ZC1 PR29 D6 MTSP-295 OPTICS MODULE - SMC HEATER/tHb Board Sheet 1 2 2 2 2 2 1 1 VCC 1 2 1 VCC R10 4.7 R9 1G STANDOFF STANDOFF Z4 ZPWB1 ZB1 1 1G R4 150 R2 J1 1 2 2 150 R3 2 3 1 LHEADER_8 8 7 6 5 4 3 2 1 Z3 IDCCONN26 4 2 STANDOFF 3 1 CABLE26 Z7 10 8 6 STANDOFF Z2 14 12 9 7 5 Z6 13 11 Z1 J5 HEADER_7X2 1 GL5ED60 D1 RED 2 GREEN STATLEDBLNK 4.7 R1 ROWA ROWB ROWC ROWD LCDBL 1 VCC 2 2 4 6 8 10 12 14 16 18 20 22 24 26 IDCPCB_5X4+2X3 1 3 5 7 9 11 13 15 17 19 21 23 25 J2 STATLEDR STATLEDG CON1 1 J4 CON1 1 J3 COLA COLB COLC COLD VCC LCD BACKLIGHT ANODE LCD BACKLIGHT CATHODE DISPLAY IR Board - Interconnect Sheet 1 2 9-69 VCC 9-70 14 13 1 2 C1 22pF 2 C2 22pF GND + 1 10M VCC TOIM3232 13 C4 22uF/16V 1 R5 S2 C5 0.1UF 100 JP1 HEADER_2 2 1 2 1 X2 VCC 1 R8 301 R11 2 1 3.6864MHz Y1 7 12 10 9 15 1 C6 0.22uF 1 4 6 CX1 RXD TXD CX2 LEDC LEDA 2 7 8 U2 + 1 IDCCONN14 S1 RD_LED TD_LED RD_IR 14 HSDL-1000 GND Z9 X1 TD_232 RD_232 TD_IR VCC_SD 5 VCC VCC 6 4 3 BR/D RESET 16 2 1 U1 3 CABLE14 Z8 6 8 10 12 C7 2.2uF/10V 1 IDCPCB_7X2 2 4 2 0.1uF C3 1 5 7 9 11 1 2 1 3 J6 VCC R6 7.5 VCC 1 DATA/CONTROL RXD TXD DEVICE_RESET VCC D2 HSDL-4220 R7 7.5 DISPLAY IR Board - IR Circuit Sheet 2 2 2 2 5 2 2 2 8 1 1 2 2 10 Illustrated Parts List 10 Illustrated Parts List OPTI 1 System The Illustrated Parts List for the AVL OPTI 1 and OPTI CCA analyzers are included to provide service personnel with part location diagrams. Please note that parts, which are specific to one type of analyzer only, are identified in the part description as to the appropriate analyzer type. The following diagrams are provided with a corresponding reference to the current part number: Main Upper Housing (Outer View) Main Upper Housing (Inner View) Lower Housing Assembly Printer Assembly SMC Assembly Gas Module Assembly Display Assembly Battery Receptacle SMC Top Cover Assembly Top Heater Plate 10-1 10 Illustrated Parts List Main Upper Housing (Outer View) Item Part Number Description 1 BP7022 ASSEMBLY: SMC Cover OPTI 1 1a BP7047 ASSEMBLY: SMC Cover OPTI CCA * Not available as a spare part 2 RE7009 PLATE: Retainer Plate Hinge Shaft 3 DF7001 SPRING: Torsion 4 DZ7006 RETAINER: Shaft 5 RE7035 GEAR: SMC Cover 6 DS7016 SCREW: M# X 8 PPH 7 DF7001 SPRING: Torsion 8 RE7034 RETAINER: Printer Cover 9 RE7028 COVER: Printer Access 10 BP7012 ASSEMBLY: Peri Pump Cartridge 11 RE7026 RECEPTACLE: Pump Cartridge 12 BP7011 ASSEMBLY: Printer 13 BK7001 CABLE: Printer 14 BP7017 ASSEMBLY: Battery Receptacle 15 ZB7002 LABEL: Name Plate OPTI 10-2 10 Illustrated Parts List Main Housing Assembly (Outer View) 8 1 9 2 3 10 4 5 11 6 7 12 13 14 15 10-3 10 Illustrated Parts List Main Upper Housing (Inner View) Item Part Number Description 1 BP7017 ASSEMBLY: Battery Receptacle 2 BP7027 ASSEMBLY: Pump Motor 3 BP7016 ASSEMBLY: Display OPTI 1 3a BB7013 ASSEMBLY: Display IR OPTI CCA 4 BP7018 ASSEMBLY: SMC OPTI 1 4a BP7053 ASSEMBLY: SMC OPTI CCA 5 RE7024 DOOR: Storage Compartment 6 RE7023 COMPARTMENT: Storage 7 BP7013 ASSEMBLY: Exhaust Fan 8 YB7008 RETAINER: Fan 9 MB7005 FOOT: Adhesive 10 BP7015 ASSEMBLY: Gas Module 11 RE7036 FRAME: Bar Code 12 RE7027 LENS: Bar Code 13 RE7018 GUIDE: Bar Code 14 EL7001 READER: Bar Code 15 YB7001 PANEL: Control 16 RE7025 LATCH: SMC Cover 17 RE7032 PIVOT: Handle 18 BP7030 HANDLE: Complete OPTI 10-4 10 Illustrated Parts List Main Housing (Inner View) 1 2 3 4 8 5 7 6 9 14 11 16 12 13 18 10 17 15 10-5 10 Illustrated Parts List Lower Housing Assembly Item Part Number Description 1 RE7010 HOUSING: Lower Main 2 RE7029 COVER: EPROM 3 HB7000 TAPE: Foam Adhesive 4 BB7008 ASSEMBLY: Main PCB OPTI 1 4a BB7011 ASSEMBLY: Main PCB OPTI CCA * Not available as a spare part 5 DS0168 SCREW: Phillips Pan Head DIN 7985 M4 X 8 6 MB7005 FOOT: Adhesive 7 BP7028 PROGRAM SET: OPTI 1 7a BP7060 PROGRAM SET: OPTI CCA 10-6 10 Illustrated Parts List Lower Housing Assembly 7 5 4 3 2 1 6 10-7 10 Illustrated Parts List Printer Assembly Item Part Number Description 0 BP7011 ASSEMBLY: Printer 1 DS7005 SCREW: Thumb Slotted 2 RE7031 COVER: Printer Inner 3 DS7009 SCREW: M2 X 4 PPH 4 MB7000 PRINTER: OPTI 5 BB7003 ASSEMBLY: PCB Printer 6 BK7001 CABLE: Printer Assembly 10-8 10 Illustrated Parts List Printer Assembly 1 2 3 4 5 6 10-9 10 Illustrated Parts List SMC Assembly Item Part Number Description 1 BP7014 ASSEMBLY: SMC OPTI 1 1a BP7018 ASSEMBLY: SMC OPTI CCA 2 ED5000 LIGHT GATE: 3 RE7030 SEAL: SMC Gas I/O Port 10-10 10 Illustrated Parts List SMC Assembly 2 3 1 10-11 10 Illustrated Parts List Gas Module Assembly Item Part Number Description 0 BP7015 ASSEMBLY: Gas Module 1 RE7014 HOUSING: Gas Module 2 DS7000 SCREW: CSF Phillips 6-32 X .375 Self Locking 3 YB7015 SPACER: Gas Module 4 MB7001 REGULATOR: Gas 5 EU7002 CONNECTOR: 5 Pin Locking 6 EU5005 PIN: Connector Amp # 530151-6 7 SS0063 TUBING: Tygon 8 MB7002 VALVE: Gas 9 DS7001 SCREW: M2 X 14 10-12 10 Illustrated Parts List Gas Module Assembly 5 2 6 1 3 4 7 9 8 10-13 10 Illustrated Parts List Display Assembly Item Part Number Description 1 DS7012 Screw: M3 X 6 2 BP7016 Assembly: Display OPTI 1 2a BB7013 Assembly: Display IR OPTI CCA 10-14 10 Illustrated Parts List Display Assembly 1 2 10-15 10 Illustrated Parts List Battery Receptacle Assembly Item Part Number Description 0 BP7017 ASSEMBLY: Battery Receptacle 1 RE7013 HOUSING: Battery Receptacle 2 BK7003 CABLE: Battery Receptacle 10-16 10 Illustrated Parts List Battery Receptacle 1 2 Negative Thermistor Positive 10-17 10 Illustrated Parts List SMC Top Cover Assembly Item Part Number Part Description 0 BP7022 ASSEMBLY: SMC Cover, Complete OPTI 1 0a BP7047 ASSEMBLY: SMC Cover, Complete OPTI CCA * Not available as a spare part 1 RE7008 COVER: SMC 2 DF7002 SPRING: SMC Cover Compression 3 BP7023 ASSEMBLY: Upper Heater Plate Carrier 3a BP7058 ASSEMBLY: Heater Plate/tHb Block OPTI CCA * Not available as a spare part 4 DZ7004 WASHER: For AB No. 8 Screw 5 DS7013 SCREW: AB No. 8 X .5 inch Phillips 6 DZ7005 CAP: Black AB No. 8 10-18 10 Illustrated Parts List SMC Top Cover Assembly 1 2 3 4 5 6 10-19 10 Illustrated Parts List SMC Top Plate Assembly (OPTI 1 Only) Item Part Number Part Description 1 DS7014 SCREW: M2 X 12 PPH 2 DS7012 SCREW: M3 X 6 3 BB7004 ASSEMBLY: PCB Heater 4 RE7007 CARRIER: Upper Heater Plate 5 YA5022 INSULATOR: Transistor for Heater Module 6 YB7011 PLATE: Upper Heater 10-20 10 Illustrated Parts List SMC Top Plate Assembly (OPTI 1 Only) 1 2 3 4 5 6 10-21 11 Service Spare Parts 11 Service Spare Parts OPTI System The AVL OPTI 1 and OPTI CCA instruments are field repairable. Several assemblies require factory adjustment or alignment and field repair of these assemblies should be limited to assembly replacement. These parts are indicated by an “*”. Parts that are used only within either OPTI 1 or OPTI CCA are indicated. All other parts can are used in either instrument. A list of recommended spare parts is provided below. Recommended Spare Parts List * BP7007 * BP7012 RE7034 RE7028 RE7029 BP7028 BP7060 DS7008 DS7007 BB7008 MB7005 RE7026 DS7006 BP7027 BP7013 DS7003 * BP7015 DS7001 MB7002 DS7000 RE7024 RE7023 DS7004 BP7017 BP7011 BP7016 BB7013 YB7001 DS7016 OPTI 1 OPTI CCA OPTI 1 OPTI 1 OPTI CCA ASSEMBLY, BATTERY ASSEMBLY, CARTRIDGE, PERI-PUMP RETAINER, COVER, PRINTER ACCESS COVER, PRINTER ACCESS COVER, EPROM PROGRAM SET, OPTI 1 PROGRAM SET, OPTI CCA SCREW, M4 X 76 SCREW, M4 X 12 ASSEMBLY, PCB, MAIN, OPTI 1 FOOT, ADHESIVE RECEPTACLE, PUMP CARTRIDGE SCREW, M3 X8, PFH ASSEMBLY, MOTOR, PUMP ASSEMBLY, FAN, EXHAUST SCREW, M4 X 50 ASSEMBLY, GAS MODULE SCREW, M2 X 14, PPH VALVE, GAS SCREW, CSFH, PHILLIPS, 6-32 X .375, SELF LOCK COMPARTMENT, STORAGE DOOR, COMPARTMENT, STORAGE SCREW,M4 X 20, PPH ASSEMBLY, RECEPTACLE, BATTERY ASSEMBLY, PRINTER ASSEMBLY, DISPLAY OPTI 1 ASSEMBLY, DISPLAY IR OPTI CCA PANEL, CONTROL SCREW, M3 X 8, PPH 11-1 11 Service Spare Parts * * * 11-2 BP7022 DS7013 DZ7004 DF7002 DZ7005 DF7001 RE7009 BP7014 BP7018 RE7030 BP7020 EL7001 RE7018 RE7027 RE7025 BP7030 YB7019 YB5012 OPTI 1 OPTI 1 OPTI CCA ASSEMBLY, COVER, SMC OPTI 1 SCREW, AB NO.8 X .5 INCH CUSTOM PHILLIPS WASHER, FOR AB NO. 8 SCREW SPRING, COMPRESSION, SMC COVER CAP, BLACK, AB#8 CUSTOM SPRING, TORSION PLATE, HINGE, RETAINER, SHAFT ASSEMBLY, SMC OPTI 1 ASSEMBLY, SMC OPTI CCA SEAL, GAS 1/0 PORT, SMC ASSEMBLY, MOTOR, VALVE DRIVE READER, BAR CODE GUIDE, BAR CODE LENS, BAR CODE LATCH, SMC COVER HANDLE, COMPLETE, OPTI TUBE, PORT, CLEAN-OUT, OPTI FITTING: TUBING REDUCER VALUE PLASTIC