Download MAC 1600 Maintenance
Transcript
System Configuration Field Preview before Analysis Comment Determines waveform preview options. Values include: No Waveforms are never previewed. Always Waveforms are always previewed. Yellow electrodes Waveforms are previewed when the Hookup Advisor indicator shows a yellow or red electrode. Red electrodes Waveforms are previewed when the Hookup Advisor indicator shows a red electrode. For additional information, refer to “Hookup Advisor Module” on page 5-12. Reanalysis Enables/disables the reanalysis feature, which allows you to adjust the following ECG measurements: P Duration PR Interval QRS Duration QT Interval Available only if Audit Trail is disabled and one of the following options is activated: ME12, MEHR, MI12, or MIHR. For more information on activating options, refer to “Options Setup” on page 9-29. For more information on the reanalysis feature, refer to “ECG Reanalysis” on page 5-8. QTC Calculation Determines which formula will be used to correct QT calculations. Options are: Bazett QTc = QT HR -------60 Framingham 60-⎞ QTc = QT + 154 ⎛ 1 – ------⎝ HR⎠ Fridericia HRQTc = QT 3 ------60 In all formulas, HR = Heart Rate. Bazett is available only if the MEHR or MIHR option is activated. Framingham and Fridericia are available only if the ME12 or MI12 option is activated. 2028451-182B MAC™ 1600 9-9