System Configuration
Field
Preview before Analysis
Comment
Determines waveform preview options. Values include:
„
No
Waveforms are never previewed.
„
Always
Waveforms are always previewed.
„
Yellow electrodes
Waveforms are previewed when the Hookup Advisor
indicator shows a yellow or red electrode.
„
Red electrodes
Waveforms are previewed when the Hookup Advisor
indicator shows a red electrode.
For additional information, refer to “Hookup Advisor
Module” on page 5-12.
Reanalysis
Enables/disables the reanalysis feature, which allows you
to adjust the following ECG measurements:
„
P Duration
„
PR Interval
„
QRS Duration
„
QT Interval
Available only if Audit Trail is disabled and one of the
following options is activated: ME12, MEHR, MI12, or
MIHR. For more information on activating options, refer to
“Options Setup” on page 9-29.
For more information on the reanalysis feature, refer to
“ECG Reanalysis” on page 5-8.
QTC Calculation
Determines which formula will be used to correct QT
calculations. Options are:
„
Bazett
QTc = QT HR
-------60
„
Framingham
60-⎞
QTc = QT + 154 ⎛ 1 – ------⎝
HR⎠
„
Fridericia
HRQTc = QT 3 ------60
In all formulas, HR = Heart Rate. Bazett is available only if
the MEHR or MIHR option is activated. Framingham and
Fridericia are available only if the ME12 or MI12 option is
activated.