59 Placement and calibration of sensors Toxic gas sensors 1.67 lpm is the recommended flow rate to mirror with Percentage of air flowing through the pump Save Figure 4-15: Flow Rate Calibration 6. Manually adjust the flow rate (as needed) by repeatedly pressing the Up/Down key to select the desired level. 7. Press the Save softkey to store this as your flow rate calibration. The flow rate summary will appear. NOTE: you may wish to repeat these steps if the desired level was not achieved. Press the Cancel softkey to stop the calibration. Next, press the Retry softkey and repeat the calibration steps to re-cal the flow rate. 8. Press the Exit softkey to exit this screen. (The pump will turn off.) 9. Press the On/Off/Esc key to return to the start screen. Toxic gas sensors The gas sensor chamber will allow you to measure any one of the single toxic gases listed below. Please refer to Appendix A for detailed specifications. 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. Carbon monoxide (CO) with a range of 0-1000 ppm Chlorine (CL2) with a range of 0-20 ppm Ethylene oxide (EtO) with a range of 0-20 ppm Hydrogen cyanide (HCN) with a range of 0-50 ppm Hydrogen sulfide (HS2) with a range of 0-500 ppm Nitrogen dioxide (NO2) with a range of 0-50 ppm Nitric oxide (NO) with a range of 0-100 ppm Oxygen (02) with a range of 0-30% Sulphur dioxide (SO2) with a range of 0-50 ppm Ozone (O3) with a range of 0 – 1.0 ppm