Download DP User Manual

Transcript
Once the settings for the step fitting are reasonable, the table of values
can be saved from the software using the Save Step Table button in
the operation panel.
The save dialog shown here prompts for the location and name of the
file.
The data is always saved as a plain text (ASCII)
file. The final data column, p-Value, is a measure
of the significance of the step. A smaller value
gives a larger statistical weight to the step.
6.4.10 Minimum value
This function calculates the minimum value of the force curve. If this is performed after the Baseline operation, this gives a measure of the maximum adhesion.
The Minimum function displays the lowest value
read from the curve along with the position information.
This operation is designed to be used after the
Baseline operation, as shown here, where the zero
force level has already been set. In this case, the
Minimum function gives a measure of the maximal
adhesion (in this case 82.02 mV). If this operation
is used on a force curve where the deflection has
been calibrated in Newtons, then the adhesion
value here is also in Newtons.
The data is read from the Retract channel by default, it can also be set to the Extend channel.
DP Data Processing Software
Version 4.2
67