4.8.2
Showing switch test results
Two types of switch test results are available:
 Results during a switch test.
 Results of a finished switch test.
Results during a switch test

Access the Switch Test Result List by selecting “Result” from
the Switch Test menu.
This shows the results that are currently available. These
results change as the test progresses.

Press
to return to the switch test.
Finished switch test results
At the end of a switch test the results are displayed. These show the
temperature when the thermostat has closed and the temperature
when it has opened – whichever comes first. The difference between
these 2 temperatures is calculated as the hysteresis.

Select “Save” to save the results storing them in the
calibrator’s memory.
50
2011-10-25
128206
00