Download User Manual Reference/Professional Temperature
Transcript
4.8.2 Showing switch test results Two types of switch test results are available: Results during a switch test. Results of a finished switch test. Results during a switch test Access the Switch Test Result List by selecting “Result” from the Switch Test menu. This shows the results that are currently available. These results change as the test progresses. Press to return to the switch test. Finished switch test results At the end of a switch test the results are displayed. These show the temperature when the thermostat has closed and the temperature when it has opened – whichever comes first. The difference between these 2 temperatures is calculated as the hysteresis. Select “Save” to save the results storing them in the calibrator’s memory. 50 2011-10-25 128206 00