Download User Manual Reference/Professional Temperature
Transcript
Use the NUMERIC keys to set the value of the voltage and/or the temperature. Current(C) and temperature(T) span: (using current) The minimum and the maximum of the current and the corresponding temperature span can be set here. Use the NUMERIC keys (RTC only) / ARROW keys (PTC only) to set the value of the current and/or the temperature. Number of wires: (using RTD) The number of wires used for the sensor under test can be selected here. Choose between 2, 3 or 4 wires. Use stability criteria: (using thermocouple, voltage (RTC only), current and RTD) Beside the stability check on the Reference sensor, it is also possible to ensure that the Sensor Under Test (SENSOR) is stable before the temperature is indicated as stable. “yes” – Stability will be checked on both Reference sensor (TRUE) temperature and Sensor Under Test (SENSOR) temperature. “no” – Stability will be checked on Reference sensor (TRUE) temperature only. Stability tolerance: (using thermocouple, voltage (RTC only), current and RTD) Enter the Stability tolerance (temperature) by pressing the NUMERIC keys (RTC only) / ARROW keys (PTC only). The expected performance of the Sensor Under Test should be considered before setting the tolerance. 128206 00 2011-10-25 61