Download 223 Sample Changer User's Guide
Transcript
Maintenance A blown fuse may indicate the existence of another problem in the instrument. If the replacement fuses blow, do not try others. To change a fuse, follow these steps. 1 Disconnect the power cord from the power outlet and from the rear panel receptacle. 2 Locate the fuse drawer on the rear panel. See page 2-3 if necessary. 3 Insert a small screwdriver into the notch located at the left of the fuse drawer. 4 Twist the screwdriver to open and remove the fuse drawer. The fuse drawer contains one 2.0A T Slo-Blo fuse (5 x 20 mm size) for a 100/120 voltage selection. It contains two 2.0A fuses for a 220/240 voltage selection. 5 Remove the old fuse(s) and insert the new fuse(s). 6 Insert the fuse drawer into its receptacle in the sample changer. Replacing Parts Replacing a Fuse 4 4-5