IDA-4 Plus Operators Manual User Performance Check The Calibration is set at the factory. For accurate Re-Calibration or resetting of Calibration the unit must be returned via the dealer. Users may wish to check for faults after a period of use or after shock to the unit. We suggest the pressure measurement performance is checked by connecting an external pressure gauge and syringe to the fluid inlet connector of each channel fitted. Initiating an occlusion test on each channel in turn (each Fluid-inlet must be vented to atmosphere at the start of the test), and then raise the pressure to 1 Bar on each channel fitted and checking that the pressure readings are approximately as expected. We suggest that flow measurement performance is checked using a syringe pump which has been calibrated with an accurate burette. We suggest checking each channel at 20 ml/hr. Please note that a new syringe must be fitted before performing this test, and the syringe pump must be allowed time to come up to speed before starting the Flow Test on each channel. 4-2