Download Operators Manual
Transcript
Infusion Pump Analyzer Operating Instructions 3 Flow Test Screen (End of Test Mode) This view of the FLOW TEST screen allows the operator to save, print or delete the results of the test. CHANNEL 1 FLOW 0.00 VOLUME ELAPSED TIME 0.00 Avg ml /h ml Save This saves the results of the flow test, or series of tests, to non-volatile memory for later printing or transfer to a computer via the RS232 port. Save & Print 00: 00: 00 SAVE SAVE & PRINT PRINT GRAPH STATUS DELETE This saves the results as described above, then sends a report to a printer connected to the printer port on the instrument. The report will be in either full or summary format, as previously selected by the operator from the UTILITIES / PRINTER menu. Print This sends a report to a printer connected to the printer port on the instrument. The report will be in either full or summary format, as previously selected by the operator from the UTILITIES / PRINTER menu. Graph This switches to the FLOW GRAPH screen for the current channel to allow a graph of the results to be viewed before they are saved or printed. Status This switches to the STATUS ALL CHANNELS screen to allow the operator to use another channel or change settings via the UTILITIES menu. Delete This deletes the current set of results from the operational memory of the instrument, without saving to the non-volatile memory. CAUTION: Once this option has been invoked there will not be any way to recover, view or print the results of this test. 3-17