Ocular Calibration Sheet Date: Microscope (type and #): Ocular type: Magnification: Operator: Front lens: Stage micrometer Measurement Position 1 Position 2 Length 1 div. div. mm 2 div. div. mm 3 div. div. mm Mean mm Reticular scale Position 1 Position 2 div. div. div. div. div. div. Length div. div. div. div. Calculations 1) Calculate the length between the two positions on the stage micrometer by subtracting the value of position 1 from the value of position 2. Dualply the length in divisions with the length of each division to get the length in mm. 2) Calculate the length between the two positions on the reticule scale by subtracting the value of position 1 from the value of position 2. 3) Calculate the mean length value of both the stage micrometer and the reticular scale. 4) Calculate the conversion factor: δ= 62 Mean Stage Micrometer Length (mm) Mean Reticule Scale (div) Auto dual wire myograph system Model 510A