Download Spectrophotometer User Guide
Transcript
Sipper screen * SIPPER * SIPPER MODE AIR GAP SAMPLE VOL. SAMPLE LOW VOL. : : : : : : OFF SIP 50 cm 1.000 ml WASTE OFF VIEW CALIB CALIBRATE Use this screen to set up the MiniSipper for the required analysis. These parameters are saved with any data produced by the software. To reach this screen, press Accessories from the Home screen, then select Sipper and press Enter. To change a parameter setting, select the parameter and press Enter. See Parameter Entry for more information. Item Function Sipper Turns the MiniSipper on and off. Mode Selects the run mode for the MiniSipper. Sip – Sets the system to fill the flow cell. Sip&Run – Sets the system to fill the flow cell and automatically perform a measurement. The current method is used to produce the result (e.g., if Fixed is current, then the sample will be scanned using the Fixed method as programmed). Continuous – Sets the system to pump continuously to waste. Alternate switch presses will start and stop pumping. The instrument will not process any key presses while the MiniSipper is pumping in continuous mode. AutoSam – Configures the Sipper to operate with an autosampler. Air Gap Enter value between 0 and 500 cm. Sets the gap between the trailing meniscus of the current sample and the leading meniscus of the next sample. The gap is measured to the nearest centimeter. For best results, set the airgap to ≥8 cm from the flow cell. Sample Vol Enter a value between 0.5 and 10.000 ml. Sets the volume of sample to be pumped. 142 Spectrophotometer User Guide