White Light screen Use this option to align the grating so the zero order diffraction (white light) passes through the sample compartment. You can see the beam when you place a white card or similar item in the light path. The visible beam can be helpful for aligning optical accessories in the sample compartment. In double beam instruments, the action of the chopper causes the light to alternate between the sample and reference beams. To realign the grating, press Initialize. When alignment is completed, press Stop to return the grating to its normal position. To display the Setup screen, press Setup Page. Setup CVC screen Use this option to access the options available for setting up and running the Calibration Verification Carousel (CVC). See the CVC section of this manual for more information. Lamps screen The Lamp functions can be accessed from the Setup screen or from the Home screen, if User Log-on is not in use. The Lamps screen shows the energy level and status (On, Off or Failed) of the available lamps (deuterium and/or tungsten halogen). The Lamps screen function keys differ depending on the instrument model. TUNGSTEN HOURS ENERGY : : : * LAMPS * ON 239 10% D2 HOURS ENERGY : : : OFF 55 0% SWITC H D2 D2 ENERG Y W ENERGY RESET D2 HRS RESET W HRS For AquaMate Plus, BioMate 6, Omega, Evolution 160 and UV-10 spectrophotometers 130 Spectrophotometer User Guide