Download Operating instructions Analytical balance
Transcript
KERN ABS 80-4N ABS 120-4N 0.1 mg 0.1 mg 80 g 120 g 0.2 mg 0.2 mg Linearity ± 0.3 mg ± 0.3 mg Recommended adjusting weight not supplied (class) 80 g (E2) 100 g (E2) 4h 8h Readability (d) Weighing range (max) Reproducibility Warm-up time Stabilization time Weighing Units Smallest part weight for piece counting Reference quantities at piece counting Weighing plate, stainless steel Dimensions of the housing (B x D x H) [mm] Dimensions Glass wind screen [mm] Net weight (kg) Permissible ambient condition Humidity of air AC adapter (Primary) Rated electric power supply Pollution Degree Overvoltage Category Altitude Installation Site ABS-N_ABJ-NM-BA-e-1312 3 sec. mg, g mg, g, ct 1 mg 5, 10, 20, 50,100 ø 91mm 210 x 340 x 325 174 x 162 x 227 (weighing space) 6 +5° C to +40° C 20 ~ 85 % relative (not condensing) AC 100 -240 V, 400 mA 50/60Hz DC 12 V, 1.25 A 2 Category II Up to 2000 m device may only be used indoors 5