Download Operating instructions Analytical balance

Transcript
16.2 Issue functions
16.2.1 Automatic data output / Auto Print function
Data output ensues automatically without pressing PRINT, as soon as the respective
output condition is fulfilled. This is defined by the menu setting.
Tab. 1:
stable /
positive
Mode 1
LD. .
Mode 2
LD.UL.
Mode 3
LD. .Z



stable /
negative

-
Stabilisation /
zero display

Check
weighing
-
Output for stable and
positive weighing value
-
Output for stable and
positive or negative
weighing value.
-
Output for stable and
positive weighing value.
New output only after zero
display and stabilisation
-
Output for stable and
positive or negative
weighing value. New output
only after zero display and
stabilisation.
Mode 4
LD.UL.Z



Mode 5
LD.OK .
-
-
-

If the Auto Print function is
connected to the check
weighing function, data of
stable weighing values are
output with indicator display
..
Enable Auto Print function:
1. Invoke Data Output menu
 Press PRINT for 3 sec. in weighing mode.
2. Select function
 Press the navigation buttons ( ) repeatedly
until „APL.PRN“ is displayed.
 Confirm with PRINT
 Press the navigation buttons ( ) repeatedly until
„AUTO.PRN“ is displayed. The current setting is marked
by the stability display ( ).
76
ABS-N_ABJ-NM-BA -e-1312