Installation
Minimum Volume Applications
To measure minimum sample volumes, use microcells (offered by PerkinElmer).
The minimum sample volume required is a function of the cell internal width or
volume and is specified below.
Cell Type
Cell
Internal
Width
Pathlength
Minimum
Volume
Required
1 cm
200 µL
B0631071
(pair)
1 cm
400 µL
B0631064
(pair)
Pathlength
Minimum
Volume
Required
0.5 µL
0.01 cm
2 µL
B0631082
2.5 µL
0.5 cm
5 µL
B0631080
5 µL
0.1 cm
10 µL
B0631083
5 µL
1.0 cm
10 µL
B0631081
30 µL
1.0 cm
50 µL
B0631079
2 mm
Height of
liquid slightly
more than
height of
4 mm
beam
Cell
Volume
Cell window
completely
filled with
liquid
Part
Number
Part
Number
NOTE: You should align microcells very carefully in the radiation beam by following the
procedures in Aligning the Single Cell Holder on page 65. When aligning
microcells, fill each cell with the minimum volume of liquid specified in the above
table to make sure that the liquid meniscus is not in the radiation beam.