Download User Manual - Gill Instruments
Transcript
Operation 3.3.7 Configuring MetPak for MODBUS MetPak Supported Modbus Specification MODBUS RTU or ASCII. Baud Rate 9600 or 19200 Baud. COMMS RS232 point to point only, RS485 2 wire networkable. Modbus Parameters ASCII or RTU. Parameter Character Error Check Frame Start Frame End Gaps in Message Start Bit Data Bits Parity Stop Bits Modbus ASCII ASCII 0 to 9 and A to F (Hex) Longitudinal Redundancy Check (LRC) Character ':' (3A Hex) Characters CR/LF (0D/0A Hex) 1 Second 1 7 Even/Odd None 1 2 Modbus RTU Binary 0 to 255 Cyclic Redundancy Check (CRC) 3.5 Characters Silence 3.5 Characters Silence 1.5 Times Character Length 1 8 Even/Odd None 1 2 Requires:MetPak with version 2346-4.00 firmware and onwards. MetSet Software version 2.02-02 and onwards. Connect a MetPak to a PC COM port (default communication is RS232). Open Gill MetSet Software. Click on Connect and Read to reach the Editing Pages as shown below. Configuration Click on the MODBUS Page Mode: Choose between RTU and ASCII, in this case RTU. Address: Choose a MetPak Address number to be between 1 to 247. Data Bits: Choose 7 for Modbus ASCII and 8 for Modbus RTU. Stop Bits: Choose 1 for Even/Odd Parity or 2 for No Parity. Parity: Choose Even/Odd or None. Termination Char ASCII code: Choose between 0-255. Inter Char Timeout (ms): Choose between 500ms to 10000ms. MetPak User Manual © Gill Instruments Ltd Page 44 Last Change 26 March, 2014