Download User's Manual
Transcript
Basic Operation When you press the SINGLE key, the oscilloscope will fill the pre-trigger buffer, and continue to flow data through the pre-trigger buffer until the Auto trigger overrides the searching and forces a trigger. At the end of the trace, the oscilloscope will stop and display the results. Normal mode Use Normal trigger mode for low repetitive-rate signals or when Auto trigger is not required. In Normal mode the oscilloscope must fill the pre-trigger buffer with data before it will begin searching for a trigger event. While searching for the trigger, the oscilloscope overflows the pre-trigger buffer; the first data put into the buffer is the first pushed out. When the trigger event is found, the oscilloscope will fill the post-trigger buffer and display the results. If the acquisition was initiated by RUN/STOP, the process repeats. If the acquisition was initiated by SINGLE, then the acquisition stops. In either Auto or Normal mode, the trigger may be missed. This is because the oscilloscope will not recognize a trigger event until the pre-trigger buffer is full. 66