Chapter 5: Using the LCD Display
Alarm: Flashes red if an instrument alarm is active, and remains solid
green if no alarms are active. If any of the alarm conditions are detected,
the Alarm button flashes on the Main tab, the aRI tab, and the Alarms
tab. Navigate to the Alarms tab to determine the nature of the alarm.
Purge Off: This is a toggle button that turns the purge valve On or Off. In
the ON state, the purge valve purges the reference chamber. If the purge
valve is OFF, then fluid flows from the IN port, through the flow cell
sample chamber, and then through the OUT port. If the purge valve is
ON, then fluid flows from the IN port, through the flow cell sample
chamber, through the flow cell reference chamber, and finally to the OUT
port.
Zero: Press ENTER on this button to cause the current value of dRI to be
considered “zero” value.
5.3 aRI Tab
The aRI tab is used to view the aRI value of the solvent in the flow cell.
Figure 5-2: aRI tab
In order to measure the aRI, the same liquid must be present in both the
sample and reference chambers. The purge valve should be ON and both
chambers should be carefully flushed with the liquid to be measured.
RIU: With the purge valve ON, displays the aRI of the liquid in the flow
cell.
5-4
Optilab rEX User’s Guide (M1500 Rev. D)