Calibration Choose the phase you want to calibrate, the display shows CALIBRATION CHOICE PAGE. CALIBRATION CHOICE = 1_ 1. V OUT AND MEAS. 2. I MEAS. 3. EXT Vref. V OUT AND MEAS.: output voltage and voltage measurement accuracy calibration. I MEAS.: current measurement accuracy calibration. EXT Vref.: external Vref calibration. 4.2.1 Output Voltage and Voltage Measurement Calibration On CALIBRATION CHOICE page, press 1 , ENTER to do the output voltage and voltage measurement calibration. CALIBRATION CHOICE = 1. V OUT AND MEAS. 3. EXT V. 1 2. I MEAS. Fig 3.8.1.1 Please Remove Load Before Calibrating Press <ENTER> to start 1. V OUT AND MEAS. ACCURACY CALI A. KEYIN THE MEASURED Vdc Vdc offset = _ mV 150V RNG T In this step A of V OUT AND MEAS. ACCURACY CALI, the user should keyin the AC source's DC output voltage measured by digital voltage meter ( DVM ) in mV. Then, 4-3