CALIBRATION
The A/D converter can be calibrated using the 6012 internal
voltage reference or an external voltage reference. For 12–bit
accuracy, you must use a voltmeter with an accuracy of 0.02% or
better.
To calibrate:
1.
Connect the digital voltmeter ground lead to J5, pin 2 (Gnd).
2.
Connect the digital voltmeter “+” lead to the test point TP1 at
the upper right–hand corner. Refer to Figure 2–2 (page 14).
3.
Adjust trim pot R3 for 5.000 VDC.
Analog Inputs – 52