Exploring the Data
Data Views
You can set the number of rows and columns, define the number of peak graphics displayed
on the screen, and change chromatographic display options.
• You can define a maximum of 50 cells.
• The x-axis range for each peak is 2.0 times the detected peak width, 50% on either side.
• The peak baseline endpoints are displayed as blue circles.
Note When the endpoints appear as blue circles, you cannot select and drag them to
manually integrate the peak.
Y To set the number of rows and columns
1. Right-click the plot and choose Options from the shortcut menu.
The Multi-Peak Options dialog box opens. For more information, see “Multi-Peak
Options dialog box parameters” on page 180.
2. In the Arrangement box, enter the number of columns and rows.
You can set the normalization level to be the same for all peaks, or you can normalize each
peak individually.
Y To set the normalization levels
1. Right-click the plot and choose Options from the shortcut menu.
The Multi-Peak Options dialog box opens. For more information, see “Multi-Peak
Options dialog box parameters” on page 180.
2. In the Intensity Scale box, select either the Normalize to Largest Peak or Use Fixed
Intensity Range option.
Y To set label and plotting styles
From the window menu, choose Options > Display Options > Chromatogram.
The Chromatogram Display Options dialog box opens. These label and plotting styles
apply to the multi-peak views. For more information, see “Chromatogram Display
Options dialog box parameters” on page 155.
When the view is locked (lock icon is yellow), you cannot select any menu item in the
shortcut menu that changes the .raw file. Menu items that would cause changes are grayed
out. For more information about locks, see “Toolbar for Selectable Views” on page 150.
Table 68 lists the parameters for the Multi Peak Plot view.
178
LCquan User Guide
Thermo Scientific